BD BBL Sensi-Disc Ciprofloxacin 5 µgRecalls & reported problems across all models
Becton, Dickinson And Company · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names at least one BD BBL Sensi-Disc Ciprofloxacin 5 µg model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of BD BBL Sensi-Disc Ciprofloxacin 5 µg by Becton, Dickinson And Company in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between May 5, 2022 and May 21, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
BD BBL Sensi-Disc Ciprofloxacin 5 µg recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jan 8, 2024Class IIOpen, Classified
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Z-1027-2024 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
Most reported BD BBL Sensi-Disc Ciprofloxacin 5 µg problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
BD BBL Sensi-Disc Ciprofloxacin 5 µg models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- BD BBL™ Sensi-Disc™ Ciprofloxacin 5 µg 231658SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall2 reports
- BD BBL™ Sensi-Disc™ Ciprofloxacin 5 µg 231657SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall0 reports
Recent BD BBL Sensi-Disc Ciprofloxacin 5 µg adverse event reports
The 2 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD BBL¿ SENSI-DISC¿ CIPROFLOXACIN - 5 ¿G THE ZONE OF INHABITATION WAS LARGER THAN EXPECTED. THIS A REPORT OF 1 OCCURRENCE FOR TWO BATCH NUMBERS (1127037, 1335457). NO FALSE RESULTS WERE REPORTED OUT TO THE CLINICIAN OR THE PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER PROBLEM: CUSTOMER IS REPORTING 231658 HAS LARGER ZONE OF INHIBITION THAN EXPECTED STEPS TAKEN WITH CUSTOMER/TROUBLESHOOTING: CUSTOMER IS REPORTING THEIR 231658 DISCS FROM LOT 1127037 AND 1335457 HAVE A ZONE OF INHIBITION OF >38MM WHEN THE EXPECTED ZONE SIZE SHOULD BE 29MM TO 38MM. NO ORGANISM TYPE WAS SPECIFIED. THE CUSTOMER STATES THEY REPEATED WITH A FRESH ISOLATE AND RESULTS WERE STILL OVER >38MM. THEY RECEIVED A NEW LOT NUMBER OF DISCS, 1335458, AND THE QC PASSED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14464492 ↗MalfunctionMay 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING THE CUSTOMER BD BBL¿ SENSI-DISC¿ CIPROFLOXACIN - 5 G. THIS EVENT OCCURRED 1 TIME(S). AT THIS TIME IT IS UNCLEAR WHETHER OR NOT PATIENT SAMPLES WERE INVOLVED. THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING A ZONE OF INHIBITION FOR THEIR 231658 DISC THAT IS LARGER THAN EXPECTED", "STEPS TAKEN WITH CUSTOMER/TROUBLESHOOTING: CUSTOMER IS REPORTING THEIR 231658 DISCS FROM LOT 1127037 AND 1335457 HAVE A ZONE OF INHIBITION OF >38MM WHEN THE EXPECTED ZONE SIZE SHOULD BE 29MM TO 38MM. NO ORGANISM TYPE WAS SPECIFIED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14289083 ↗
Other Becton, Dickinson And Company products
- BD Vacutainer SST10 models · 4,985 reports · 3 recalls
- BD Luer-Lok48 models · 5,117 reports
- BD Insyte Autoguard33 models · 4,465 reports · 8 recalls
- BD SafetyGlide17 models · 2,185 reports
- BD Vacutainer Eclipse Blood Collection Needle4 models · 1,880 reports
- BD Vacutainer Push Button Blood Collection Set9 models · 1,831 reports
- BD Nexiva25 models · 1,614 reports · 1 recall
- BD BACTEC MGIT Mycobacteria Growth Indicator Tubes2 models · 1,554 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.