BD BBL™ Sensi-Disc™ Ciprofloxacin 5 µg 231658Recalls & reported problems
Becton, Dickinson And Company · BD BBL™ Sensi-Disc™ Ciprofloxacin 5 µg · Class 2
All models of BD BBL Sensi-Disc Ciprofloxacin 5 µg (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30382902316580), received between May 5, 2022 and May 21, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 8, 2024Class IIOpen, Classified
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Z-1027-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING BD BBL¿ SENSI-DISC¿ CIPROFLOXACIN - 5 ¿G THE ZONE OF INHABITATION WAS LARGER THAN EXPECTED. THIS A REPORT OF 1 OCCURRENCE FOR TWO BATCH NUMBERS (1127037, 1335457). NO FALSE RESULTS WERE REPORTED OUT TO THE CLINICIAN OR THE PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER PROBLEM: CUSTOMER IS REPORTING 231658 HAS LARGER ZONE OF INHIBITION THAN EXPECTED STEPS TAKEN WITH CUSTOMER/TROUBLESHOOTING: CUSTOMER IS REPORTING THEIR 231658 DISCS FROM LOT 1127037 AND 1335457 HAVE A ZONE OF INHIBITION OF >38MM WHEN THE EXPECTED ZONE SIZE SHOULD BE 29MM TO 38MM. NO ORGANISM TYPE WAS SPECIFIED. THE CUSTOMER STATES THEY REPEATED WITH A FRESH ISOLATE AND RESULTS WERE STILL OVER >38MM. THEY RECEIVED A NEW LOT NUMBER OF DISCS, 1335458, AND THE QC PASSED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14464492 ↗MalfunctionMay 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT WHILE USING THE CUSTOMER BD BBL¿ SENSI-DISC¿ CIPROFLOXACIN - 5 G. THIS EVENT OCCURRED 1 TIME(S). AT THIS TIME IT IS UNCLEAR WHETHER OR NOT PATIENT SAMPLES WERE INVOLVED. THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING A ZONE OF INHIBITION FOR THEIR 231658 DISC THAT IS LARGER THAN EXPECTED", "STEPS TAKEN WITH CUSTOMER/TROUBLESHOOTING: CUSTOMER IS REPORTING THEIR 231658 DISCS FROM LOT 1127037 AND 1335457 HAVE A ZONE OF INHIBITION OF >38MM WHEN THE EXPECTED ZONE SIZE SHOULD BE 29MM TO 38MM. NO ORGANISM TYPE WAS SPECIFIED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14289083 ↗
Other SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices
- BD BBL Sensi-Disc Ampicillin 2 ug 231263Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL2 recalls0 reports
- BD BBL™ Sensi-Disc™ Piperacillin/Tazobactam 100/10 µg 231691Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall1 report
- BD BBL™ Sensi-Disc™ Ceftazidime 30 µg 231633Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall1 report
- BD BBL™ Sensi-Disc™ Levofloxacin 5 µg 231706Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall1 report
- BD BBL™ Sensi-Disc™ Cefepime 30 µg 231696Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall1 report
- BD BBL™ Sensi-Disc™ Chloramphenicol 30 µg 231274Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall0 reports
- BD BBL™ Sensi-Disc™ Cefotaxime 30 µg 231607Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall0 reports
- BD BBL™ Sensi-Disc™ Chloramphenicol - 30 µg 230733Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page