Capsurefix DXAC/DSPRecalls & reported problems across all models
Medtronic, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 372
- Deaths reported
- 5
- Injuries reported
- 155
- Recalls
- 0
We found 2 models of Capsurefix DXAC/DSP by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 372 adverse event reports in the FDA's MAUDE database, received between Apr 7, 2016 and Aug 26, 2026 — classified as 5 deaths, 155 injuries, 212 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (124 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Capsurefix DXAC/DSP problems
Device problems
- Under-Sensing124
- Over-Sensing85
- High Capture Threshold70
- Adverse Event Without Identified Device or Use Problem59
- Device Dislodged or Dislocated27
- Signal Artifact/Noise26
- Failure to Capture23
- Pacing Problem22
Patient problems
- No Clinical Signs, Symptoms or Conditions167
- No Known Impact Or Consequence To Patient110
- Unspecified Infection42
- No Consequences Or Impact To Patient12
- Chest Pain6
- Dyspnea6
- Bacterial Infection5
- Discomfort3
Capsurefix DXAC/DSP models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Capsurefix® DXAC/DSP 5568-45permanent pacemaker Electrode4 deaths192 reports
- Capsurefix® DXAC/DSP 5568-53Permanent defibrillator electrodes1 death180 reports
Recent Capsurefix DXAC/DSP adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Failure to Capture; High impedance; High Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50457146 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Pulse Generator (IPG) system was explanted due to an infection. No further patient complications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50451274 ↗MalfunctionAug 25, 2026Pacing Problem; Under-Sensing
It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events which at times resulted in unnecessary pacing. The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50432875 ↗InjuryAug 20, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to an infection. No further patient complications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50350048 ↗MalfunctionAug 18, 2026Electrical /Electronic Property Problem; Under-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50297100 ↗MalfunctionJun 26, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING ON STORED EPISODES, RESULTING IN NON-PHYSIOLOGIC INTERVALS. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT DUE TO HIGH THRESHOLDS. THE MEASUREMENT WAS NOTED TO BE CONSISTENT WITH HISTORICAL MEASUREMENTS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636157 ↗InjuryJun 22, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25573608 ↗InjuryJun 1, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT APPROXIMATELY TEN WEEKS AFTER A DEVICE EXCHANGE PROCEDURE, THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO A DEVICE POCKET INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection
MAUDE report 25355660 ↗MalfunctionMay 19, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING AND BLANKING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS WHICH RESULTED IN UNNECESSARY PACING, MISCLASSIFICATION OF NON-SUSTAINED VENTRICULAR TACHYCARDIA EPISODES, AND THE TERMINATION OF THE AT/AF DETECTED EVENTS. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING AND/ OR BLANKING WHICH RESULTED IN OVER PACED BEATS. LOW P WAVES WERE ALSO OBSERVED. BOTH LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25215466 ↗MalfunctionMay 13, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25158433 ↗MalfunctionApr 27, 2026Intermittent Capture; Electrical /Electronic Property Problem; High Capture Threshold
IT WAS REPORTED THAT THERE WERE HIGH THRESHOLDS AND AN INABILITY TO CONFIRM CONSISTENT CAPTURE ON THE RIGHT ATRIAL (RA) LEAD. IT WAS NOTED THERE WAS AN ATRIAL LEAD POSITION CHECK FAILURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25006134 ↗MalfunctionApr 24, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS AND CAUSING UNNECESSARY ATRIAL PACING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24982317 ↗MalfunctionApr 17, 2026Electrical /Electronic Property Problem; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. A FAILED ATRIAL LEAD POSITION CHECK WAS ALSO NOTED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24907710 ↗InjuryMar 19, 2026Signal Artifact/Noise; Intermittent Capture; Fracture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A FRACTURE, INTERMITTENT CAPTURE AND NOISE SENSING. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A FRACTURE, RISING PACING IMPEDANCE AND HIGH THRESHOLDS. THE RA AND RV LEADS WERE EXPLANTED AND WERE REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). THE IPG WAS EXPLANTED AND WAS REPLACED FOR UNKNOWN REASONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637297 ↗MalfunctionFeb 25, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA (AT) /ATRIAL FIBRILLATION (AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24444528 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page