CentraxRecalls & reported problems across all models
Howmedica Osteonics Corp. · 12 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 54
- Deaths reported
- 1
- Injuries reported
- 14
- Recalls
- 0
We found 12 models of Centrax by Howmedica Osteonics Corp. in the FDA's device registry (GUDID). Together they are named in 54 adverse event reports in the FDA's MAUDE database, received between Feb 27, 2017 and Feb 27, 2023 — classified as 1 death, 14 injuries, 39 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported Centrax problems
Device problems
- Device Operates Differently Than Expected14
- Insufficient Device Problem Information13
- Adverse Event Without Identified Device or Use Problem6
- Device Difficult to Setup or Prepare6
- Device Dislodged or Dislocated4
- Fitting Problem3
- Mechanical Problem3
- Device Damaged Prior to Use2
Patient problems
- No Consequences Or Impact To Patient19
- No Known Impact Or Consequence To Patient19
- Injury6
- No Clinical Signs, Symptoms or Conditions5
- Unspecified Infection4
- Joint Dislocation3
- Pain3
- Insufficient Information2
Centrax models
12 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Centrax 6226-2-645Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented11 reports
- Centrax 6226-2-641Prosthesis, hip, semi-constrained, metal/polymer, cemented9 reports
- Centrax 6226-2-652Prosthesis, hip, semi-constrained, metal/polymer, cemented8 reports
- Centrax 6226-2-644Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented7 reports
- Centrax 6226-2-650Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented6 reports
- Centrax 6226-2-642Prosthesis, hip, semi-constrained, metal/polymer, cemented3 reports
- Centrax 6226-2-643Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death2 reports
- Centrax 6226-2-648Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- Centrax 6226-2-651Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- Centrax 6226-2-653Prosthesis, hip, semi-constrained, metal/polymer, cemented2 reports
- Centrax 6226-2-246Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
- Centrax 6226-2-646Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 report
Recent Centrax adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryFeb 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT, 0002249697-2023-00192 WAS FOUND TO BE A DUPLICATE OF MFR REPORT # 0002249697-2019-01386. ALL DATA FOR THIS PATIENT'S EXPERIENCE WILL BE CAPTURED UNDER MFR REPORT # 0002249697-2019-01386. THIS PRODUCT INQUIRY IS BEING CLOSED AS A DUPLICATE. THE FOLLOWING HAS BEEN REPORTED: ADVERSE EVENT THAT OCCURRED IN JOIN STUDY CASE ((B)(4)). (B)(6), 2018: BHA PERFORMED. (B)(6), 2018: EARLY POSTOPERATIVE INFECTION WAS CONFIRMED. CLEANING AND REVISION OF INNER HEAD AND BIPOLAR CUP WAS PERFORMED. STEM PRESERVED. (B)(6), 2019: PATIENT'S OUTCOME WAS CONFIRMED AS "RECOVERY".
Patient problems: Unspecified Infection
MAUDE report 16442177 ↗InjuryFeb 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT BECAME INFECTED POSTOPERATIVELY AND UNDERWENT LESION SCRAPING.
Patient problems: Unspecified Infection
MAUDE report 16402213 ↗MalfunctionMar 29, 2022Difficult to Insert; Failure to Advance; Difficult to Insert; Failure to Advance
DURING BHA, THE HEAD MOVEMENT WITHIN THE CENTRAX IS STIFF AND DOES NOT TURN PROPERLY, UNLIKE USUAL. HE MADE SEVERAL ATTEMPTS TO LOCK THE LOCKING RING. THE RING MIGHT BE DEFORMED DUE TO IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13937997 ↗MalfunctionDec 23, 2021Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information
ONE OF THE CENTRAX LOCKING RINGS LIFTED UP A LITTLE AFTER LOCKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13079125 ↗InjuryOct 28, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THREE MONTHS AFTER THE BHA FOR FEMORAL NECK FRACTURE, THE IMAGE SHOWED THAT THE METAL HEAD CAME OFF FROM THE CENTRAX BIPOLAR CUP DURING REDUCTION. AFTER REVISION, NO DAMAGE WAS FOUND ON THE BIPOLAR CUP.
Patient problems: Joint Dislocation
MAUDE report 12715959 ↗InjuryMay 6, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
NEUROVASCULAR DISORDER ON (B)(6) 2018. TREATED WITH ORTOP. UPDATE APRIL 29, 2021: THIS EVENT WAS REPORTED AS MULTICENTER RESULTS SURVEY AT BHA AND PROVIDED BY CLINICAL. ACCORDING TO SURGEON, NEUROVASCULAR DISORDER IS RELATED TO PROCEDURE. BUT IT IS UNKNOWN IF IT IS RELATED TO DEVICE OR ONLY PROCEDURE. IT CAN BE ¿CAUSAL RELATIONSHIP¿ IS UNKNOWN, TOO.
Patient problems: Insufficient Information
MAUDE report 11784817 ↗InjuryApr 7, 2021Naturally Worn; Naturally Worn
ON (B)(6) 2015 THE FIRST THA PROCEDURE WAS PERFORMED. THE PATIENT COMPLAINED OF PAIN FROM AROUND (B)(6) 2021. ON (B)(6) 2021 REVISION WAS PERFORMED FOR SUSPICION OF METALLOSIS DUE TO WEAR OF POLYETHYLENE FROM CENTRAX. PLEASE INVESTIGATE THE DEGREE OF POLYETHYLENE WEAR.
Patient problems: Inflammation; Pain
MAUDE report 11632148 ↗InjuryFeb 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS EVENT IS ABOUT A PATIENT REGISTERED FOR JOIN RESEARCH. THE STUDY WAS DISCONTINUED DUE TO WORSENING GENERAL CONDITION TRIGGERED BY JAUNDICE AND ACUTE CHOLECYSTITIS.
Patient problems: Insufficient Information; Jaundice
MAUDE report 11348322 ↗MalfunctionDec 3, 2020Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
DURING THE INITIAL SURGERY, WHEN ASSEMBLING THE METAL HEAD WITH CENTRAX, THE LOCKING RING PARTIALLY FLOATED ABOUT 1 TO 1.5 MM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10940514 ↗MalfunctionJul 21, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING A BHA, THERE WAS A GAP BETWEEN A 52MM SHELL AND A LOCKING RING. THEREFORE, A SURGEON DECIDED TO USED 51MM BUT IT WAS ALSO SAME CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10300928 ↗MalfunctionJul 21, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING A BHA, THERE WAS A GAP BETWEEN A 52MM SHELL AND A LOCKING RING. THEREFORE, A SURGEON DECIDED TO USED 51MM BUT IT WAS ALSO SAME CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10300925 ↗MalfunctionApr 13, 2020Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
IT WAS REPORTED THAT, DURING A BHA THE LOCKING RING WAS NOT PROPERLY LOCK WITH A SHELL, A SURGEON CHECKED AND NOTICED A DEFORMED PE OF SHELL. A SPARE CENTRAX WAS USED ON THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9954271 ↗InjuryMar 27, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
JOIN FOR BHA. RIGHT BHA WAS PERFORMED ON (B)(6) 2017. DUE TO INFECTION, HEAD WAS REMOVED ON (B)(6) 2018. CUP WAS REVISED ON (B)(6) 2018.
Patient problems: Injury; Unspecified Infection
MAUDE report 9888488 ↗MalfunctionMar 25, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT, DURING A BHA, AFTER THE LOCKING, A SURGEON NOTICED A GAP BETWEEN THE SHELL AND LOCKING RING. HE USED A SPARE CENTRAX INSTEAD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9879132 ↗InjuryMar 16, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON AN UNKNOWN DATE, THE PATIENT UNDERWENT BHA PROCEDURE. ON UNKNOWN DATE AFTER BHA THE BIPOLAR CUP WAS DISLOCATED, AND ON (B)(6) 2020, "TTHE" REVISION WAS PERFORMED (THE BIPOLAR CAP AND THE INNER HEAD WERE REVISED). THE REPORTING SURGEON¿S OPINION: IT WAS FIRST CASE THAT ONLY THE BIPOLAR CAP WAS DISLOCATED. THERE WAS A POSSIBILITY THAT LOCKING HAD NOT BEEN DONE PROPERLY AT THE INITIAL BHA. BUT THE CAUSE WAS UNKNOWN. THE REPORTING SR¿S OPINION: THE INNER POLYETHYLENE OF THE BIPOLAR CAP WAS FOUND TO BE DEFORMED AFTER THE REMOVAL. IT WAS UNKNOWN IF THIS WAS THE CAUSE OF THE EVENT. BUT PLEASE INVESTIGATE WHY IT WAS DEFORMED AND THE CAUSE OF DISLOCATION.
Patient problems: Injury
MAUDE report 9834670 ↗
Other Howmedica Osteonics Corp. products
- Triathlon608 models · 3,524 reports · 15 recalls
- Scorpio107 models · 113 reports · 28 recalls
- Trident160 models · 1,280 reports · 2 recalls
- Accolade57 models · 1,228 reports
- Duracon58 models · 65 reports · 16 recalls
- LFIT23 models · 852 reports
- TRIDENT II42 models · 188 reports · 12 recalls
- INSTRUMENT131 models · 290 reports · 7 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.