CLANCYRecalls & reported problems across all models
Smith & Nephew, Inc. · 12 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 82
- Deaths reported
- 0
- Injuries reported
- 25
- Recalls
- 0
We found 12 models of CLANCY by Smith & Nephew, Inc. in the FDA's device registry (GUDID). Together they are named in 82 adverse event reports in the FDA's MAUDE database, received between Jul 18, 2017 and Mar 10, 2025 — classified as 25 injuries, 57 malfunctions.
The most frequently reported specific device problem is “Break” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CLANCY problems
Device problems
- Break55
- Material Separation11
- Device Markings/Labelling Problem8
- Flaked2
- Material Fragmentation2
- Appropriate Device Problem Term/Code Not Available1
- Crack1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- No Known Impact Or Consequence To Patient34
- No Code Available2
- No Consequences Or Impact To Patient2
- No Information2
- Foreign Body In Patient1
- Insufficient Information1
- Perforation1
CLANCY models
12 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CLANCY 7209739BIT, DRILL20 reports
- CLANCY 7209738BIT, DRILL16 reports
- CLANCY 72202972ACCESSORIES,ARTHROSCOPIC12 reports
- CLANCY 7209737INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT9 reports
- CLANCY 7209008BIT, DRILL8 reports
- CLANCY 72203164ACCESSORIES,ARTHROSCOPIC5 reports
- CLANCY 72203163ACCESSORIES,ARTHROSCOPIC4 reports
- CLANCY 72203165ACCESSORIES,ARTHROSCOPIC4 reports
- CLANCY 72203170ACCESSORIES,ARTHROSCOPIC1 report
- CLANCY 72202594ACCESSORIES,ARTHROSCOPIC1 report
- CLANCY 72203161ACCESSORIES,ARTHROSCOPIC1 report
- CLANCY 72203090ACCESSORIES,ARTHROSCOPIC1 report
Recent CLANCY adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2025Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION, THE CLANCY REAMER BROKE MID-BODY WHILE DRILLING A FEMORAL TUNNEL. TWO PIECES WERE REMOVED FROM THE PATIENT WITHOUT INCIDENT, AS THEY WERE CONNECTED TO A GUIDE PIN THROUGH THE REAMER¿S CANNULATION, GRASPER, AND PICKUPS. THE PROCEDURE WAS COMPLETED USING AN S+N BACKUP DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21566350 ↗MalfunctionSep 9, 2024Break; Break
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE HEAD OF A 10 MM DRILL FLEXIBLE ENDOSCOPIC CANNULA BROKE OFF WHILE DRILLING AN ACL TUNNEL. THE BROKEN DEVICE WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS FINISHED WITH A 9.5 MM SMITH AND NEPHEW DRILL BACKUP DEVICE. THERE WAS A 4 MINUTES SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20169457 ↗MalfunctionJul 5, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE FLEXIBLE ENDOSCOPIC DRILL BROKE IN TWO PARTS DURING DRILLING THE FEMUR. ALL THE PIECES WERE REMOVED FROM THE PATIENT WITH A GRASPER. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A BACK-UP DEVICE IN THE ORIGINAL DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19679916 ↗MalfunctionMay 8, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, AS THE PHYSICIAN BEGAN TO DRILL THE FEMUR TUNNEL, THE FLEXIBLE DRILL BROKE. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19274734 ↗MalfunctionMay 7, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, AS THE PHYSICIAN BEGAN TO DRILL THE FEMUR TUNNEL, THE FLEXIBLE DRILL BROKE. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19264724 ↗MalfunctionApr 18, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE CANNULATED FLEXIBLE DRILL FELT USED AND DID NOT HAVE THE LENGTH MARK NUMBERS ANY LONGER; THE DRILL BROKE INSIDE THE PATIENT, ALL THE PIECES WERE REMOVED WITH A PASSING PIN AND PITUITARY. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19131510 ↗MalfunctionJan 30, 2024Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, TWO (2) FLEXIBLE FEMORAL DRILLS FELT SOFT, LIKE THEY WERE GOING TO BREAK, AT THE MOMENT OF USE, BOTH DRILLS BROKE. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT USING A PITUITARY GRASPER. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS A DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18607767 ↗MalfunctionDec 26, 2023Break; Break
IT WAS REPORTED THAT DURING AN ACL, THE SURGEON FELT LIKE A DRILL FLEXIBLE ENDOSCOPIC CANNULA WAS GOING TO BREAK; AFTER HE DRILLED, IN FACT THE DEVICE BROKE IN HALF. A BACK-UP DEVICE WAS OPENED AND THE SAME ISSUE HAPPENED WITH A SECOND DRILL FLEXIBLE. ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT WITH A PITUITARY. SURGERY RESUMED AFTER A NON-SIGNIFICANT DELAY OF LESS THAN 30 MINUTES WITH THE THIRD OPENED DEVICE, NO VOIDS WERE LEFT IN THE PATIENT WHOSE BONE QUALITY WAS HEALTHY AND HIS HEALTH WAS NOT AFFECTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18395370 ↗MalfunctionOct 17, 2023Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY WHEN THE DRILL BIT DRILLED THE BONE, IT WAS VERY HARD AND THE DRILL BIT BROKE. ALL THE DRILL BIT PIECES WERE REMOVED WITH AN HEMOSTAT. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17952540 ↗MalfunctionOct 5, 2023Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE CLANCY FLEXIBLE DRILL BROKE WHEN USING IT; HOWEVER THE PIECES WERE TRAPPED IN THE DEVICE GUIDE, A HEMOSTAT WAS USED TO TRAP THEM AND REMOVE THEM FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17877367 ↗InjuryFeb 23, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE PROCEDURE, WHEN STARTING TO DRILL INTO THE PATIENTS FEMORAL CONDYLE THE FLEXIBLE CLANCY DRILL BROKE. ALL THE BROKEN PIECES WERE REMOVED WITH A HEMOSTAT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT DELAY USING A BACK-UP DEVICE TO COMPLETE THE BONE HOLE THAT WAS ALREADY STARTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16432734 ↗InjuryJan 11, 2023Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION, WHEN DRILLING INTO THE FEMUR THE BIT OF THE FELXIBLE DRILL BROKE. ALL PIECES WERE REMOVED WITH SUCTION AND WITH HEMOSTATIC FORCEPS. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16137993 ↗MalfunctionDec 6, 2022Break; Break
IT WAS REPORTED THAT, DURING SET UP, THE AUGER OF THE CLANCY FLEXIBLE DRILL, 7.0 MM WAS FOUND FRACTURED/SPLIT INSIDE THE TRAY. NO DELAY WAS REPORTED AND THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK-UP DEVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15925588 ↗InjurySep 22, 2022Break; Break
IT WAS REPORTED THAT DURING USE IN AN UNSPECIFIED PROCEDURE, WHEN THE DRILLING BEGAN IN THE FEMORAL TUNNEL, THE FLEXIBLE DRILL BROKE. IT IS UNKNOWN IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY.
Patient problems: Insufficient Information
MAUDE report 15468248 ↗InjurySep 22, 2022Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY A TUNNEL WAS MADE WITH THE DRILL FLEXIBLE ENDOSCOPIC CANN, HOWEVER THE SPECIALIST DID NOT LIKE THE POSITION VERY MUCH, AND DID A SECOND TUNNEL, WHERE THE DRILL BROKE. AT THE END, THE DOCTOR STUCK WITH THE FIRST TUNNEL OPTION, THE BROKEN PARTS OF THE DRILL WERE COMPLETELY REMOVED. NO DELAY WAS REPORTED, AND THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Perforation; Rupture
MAUDE report 15462925 ↗
Other Smith & Nephew, Inc. products
- FAST-FIX18 models · 9,554 reports
- JOURNEY566 models · 6,250 reports · 4 recalls
- GENESIS II462 models · 4,599 reports · 2 recalls
- DYONICS, POWERMAX, ELITE2 models · 4,308 reports
- TRIGEN381 models · 3,599 reports · 5 recalls
- LEGION446 models · 2,039 reports · 4 recalls
- REAL INTELLIGENCE32 models · 2,017 reports · 3 recalls
- Coblation62 models · 2,126 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page