CLANCY 72203163Recalls & reported problems
Smith & Nephew, Inc. · CLANCY 7.5 MM FLEXIBLE DRILL · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010654564), received between Dec 11, 2018 and Feb 23, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation2
- Break1
- Crack1
Patient problems
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryFeb 23, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE PROCEDURE, WHEN STARTING TO DRILL INTO THE PATIENTS FEMORAL CONDYLE THE FLEXIBLE CLANCY DRILL BROKE. ALL THE BROKEN PIECES WERE REMOVED WITH A HEMOSTAT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT DELAY USING A BACK-UP DEVICE TO COMPLETE THE BONE HOLE THAT WAS ALREADY STARTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16432734 ↗InjuryJun 2, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING SURGERY, WHILE MAKING THE FEMUR TUNNEL, WHEN THE FLEXIBLE DRILL TOUCHED THE BONE IT BROKE. EVERYTHING WAS REMOVED FROM THE PATIENT USING A HEMOSTAT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. IN THE SAME BONE HOLE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10112658 ↗MalfunctionJan 13, 2020Material Separation; Material Separation
IT WAS REPORTED DURING AN ACL BTB SURGERY THE DRILL BIT BROKE WHILE ATTEMPTING TO DRILL, SURGERY WAS COMPLETED WITH A 7.0 MM FLEXIBLE BIT. BROKEN PIECES WERE REMOVED FROM THE PATIENT WITH A HEMOSTAT. NO PATIENT INJURIES OR DELAY REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9580581 ↗MalfunctionDec 11, 2018Crack
IT WAS REPORTED THAT DURING A LIGAMENTOPLASTY THE 7.5MM CLANCY FLEXIBLE DRILL CRACKED. IT TOOK ABOUT 50 MINUTES TO REMOVE THE PIECE FROM THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8153144 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.