Skip to content

Cobalt XT HF CRT-D MRI SureScanRecalls & reported problems across all models

Medtronic, Inc. · 2 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

6 open recalls name at least one Cobalt XT HF CRT-D MRI SureScan model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
701
Deaths reported
20
Injuries reported
469
Recalls
6

We found 2 models of Cobalt XT HF CRT-D MRI SureScan by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 701 adverse event reports in the FDA's MAUDE database, received between Mar 3, 2021 and Sep 29, 2026 — classified as 20 deaths, 469 injuries, 212 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (100 reports).

6 FDA recall records name at least one of these models by its device identifier, 6 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 10 reportsDec 2023: 9 reportsJan 2024: 12 reportsFeb 2024: 12 reportsMar 2024: 15 reportsApr 2024: 12 reportsMay 2024: 17 reportsJun 2024: 7 reportsJul 2024: 14 reportsAug 2024: 21 reportsSep 2024: 10 reportsOct 2024: 11 reportsNov 2024: 14 reportsDec 2024: 6 reportsJan 2025: 24 reportsFeb 2025: 15 reportsMar 2025: 19 reportsApr 2025: 16 reportsMay 2025: 16 reportsJun 2025: 14 reportsJul 2025: 29 reportsAug 2025: 15 reportsSep 2025: 13 reportsOct 2025: 19 reportsNov 2025: 21 reportsDec 2025: 22 reportsJan 2026: 25 reportsFeb 2026: 25 reportsMar 2026: 15 reportsApr 2026: 18 reportsMay 2026: 22 reportsJun 2026: 8 reportsJul 2026: 6 reportsAug 2026: 11 reportsSep 2026: 21 reportsOct 2026: 0 reports20242025202629
Last 36 months. The most recent month may be incomplete.

Cobalt XT HF CRT-D MRI SureScan recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 16, 2024Class IIOpen, Classified

    Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

    Z-2192-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Apr 16, 2024Class IIOpen, Classified

    Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

    Z-2191-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  3. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1777-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. May 10, 2023Class IOpen, Classified

    There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

    Z-1774-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Jun 22, 2022Class IOpen, Classified

    There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1486-2022 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  6. Apr 12, 2022Class IIOpen, Classified

    Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

    Z-1267-2022 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported Cobalt XT HF CRT-D MRI SureScan problems

Device problems

  • Adverse Event Without Identified Device or Use Problem281
  • Device Sensing Problem100
  • Pacing Problem80
  • Inappropriate/Inadequate Shock/Stimulation68
  • Mechanical Problem55
  • Appropriate Device Problem Term/Code Not Available48
  • Premature Discharge of Battery37
  • Over-Sensing31

Patient problems

  • No Clinical Signs, Symptoms or Conditions284
  • Unspecified Infection181
  • Shock from Patient Lead(s)63
  • Erythema54
  • Purulent Discharge46
  • Discomfort36
  • Erosion32
  • Swelling/ Edema31

Cobalt XT HF CRT-D MRI SureScan models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Cobalt XT HF CRT-D MRI SureScan adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic, Inc. products

All Medtronic, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page