- Adverse event reports
- 206
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 3 models of Dexcom G4 Platinum Replacement Receiver Kit by Dexcom, Inc. in the FDA's device registry (GUDID). Together they are named in 206 adverse event reports in the FDA's MAUDE database, received between Jan 27, 2017 and Aug 21, 2020 — classified as 3 injuries, 203 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (100 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Most reported Dexcom G4 Platinum Replacement Receiver Kit problems
Device problems
- Device Displays Incorrect Message100
- Wireless Communication Problem35
- Unable to Obtain Readings25
- No Audible Alarm11
- Defective Alarm9
- Inappropriate or Unexpected Reset8
- Device Inoperable4
- Device Operates Differently Than Expected3
Patient problems
- No Consequences Or Impact To Patient202
- Hypoglycemia3
- Loss of consciousness2
- Memory Loss/Impairment1
- No Known Impact Or Consequence To Patient1
Dexcom G4 Platinum Replacement Receiver Kit models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001Sensor, Glucose, Invasive126 reports
- Dexcom G4 Platinum Replacement Receiver Kit STR-GL-BLUSensor, Glucose, Invasive54 reports
- Dexcom G4 Platinum Replacement Receiver Kit STR-GL-PNKSensor, Glucose, Invasive26 reports
Recent Dexcom G4 Platinum Replacement Receiver Kit adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2020Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10439775 ↗MalfunctionJan 3, 2020Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9545514 ↗MalfunctionJan 2, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9538038 ↗MalfunctionNov 8, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9296248 ↗MalfunctionOct 15, 2019Defective Alarm; Defective Alarm
THE PRODUCT WAS EVALUATED. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. CHARGED AND BOOT WERE PERFORMED AND PASSED. RECEIVER FUNCTIONAL TESTING WAS PERFORMED AND PASSED. "TRY IT" TESTING WAS PERFORMED AND PASSED. DROP TEST FOR INTERMITTENT AUDIO WAS PERFORMED AND PASSED. CASE OPENED FOR INTERIOR INSPECTION WAS PERFORMED AND PASSED. AUDIO SPEAKER RESISTANCE WAS MEASURED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERRORS RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9193667 ↗MalfunctionOct 10, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9177516 ↗MalfunctionAug 6, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE TEST WAS PERFORMED AND PASSED. A RECEIVER BOOT TEST WAS PERFORMED AND FAILED DUE TO IT WOULD NOT BOOT. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. THE LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8864655 ↗MalfunctionJul 30, 2019Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8842370 ↗MalfunctionJun 21, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT AN ERRPR CALL TECH SUPPORT ERROR ICON WAS DISPLAYED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8723330 ↗MalfunctionJun 6, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT A "HWBB1" ERROR AND A "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8674154 ↗MalfunctionMay 29, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT A "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8650003 ↗MalfunctionMay 21, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8627302 ↗MalfunctionMay 20, 2019Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8626222 ↗MalfunctionMay 16, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT AN "ERR 224" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8615850 ↗MalfunctionApr 26, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT A "HWRF" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8554723 ↗
Other Dexcom, Inc. products
- Dexcom G613 models · 637,302 reports · 7 recalls
- Dexcom G6 Transmitter2 models · 469,968 reports
- Dexcom G732 models · 325,909 reports · 8 recalls
- Dexcom G521 models · 206,314 reports · 1 recall
- Dexcom G7 GSS (161)4 models · 141,244 reports · 1 recall
- Dexcom G4 Platinum41 models · 119,308 reports · 2 recalls
- Dexcom G7 Receiver4 models · 33,732 reports · 1 recall
- Dexcom G7 GSS (160)4 models · 33,558 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.