Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001Recalls & reported problems

Dexcom, Inc. · G4 Platinum Replacement Receiver Kit Adult Black · Class 3

All models of Dexcom G4 Platinum (41) · Dexcom G4 Platinum Replacement Receiver Kit (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G4 Platinum receiver is the handheld receiver used with a glucose sensor. Many of the 126 reports describe incorrect on-screen messages, wireless communication problems, and unexpected shutdowns, and most say the patient was not affected. The reports include no deaths and 2 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
126
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 50386270000069), received between Feb 2, 2017 and Aug 21, 2020 — classified as 2 injuries, 124 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (58 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message58
  • Wireless Communication Problem22
  • Unable to Obtain Readings17
  • Inappropriate or Unexpected Reset7
  • Defective Alarm6
  • Device Operates Differently Than Expected3
  • Display or Visual Feedback Problem3
  • No Audible Alarm3

Patient problems

  • No Consequences Or Impact To Patient123
  • Hypoglycemia2
  • Loss of consciousness1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10439775 ↗
  • MalfunctionOct 15, 2019Defective Alarm; Defective Alarm

    THE PRODUCT WAS EVALUATED. AN EXTERIOR VISUAL INSPECTION WAS PERFORMED AND PASSED. CHARGED AND BOOT WERE PERFORMED AND PASSED. RECEIVER FUNCTIONAL TESTING WAS PERFORMED AND PASSED. "TRY IT" TESTING WAS PERFORMED AND PASSED. DROP TEST FOR INTERMITTENT AUDIO WAS PERFORMED AND PASSED. CASE OPENED FOR INTERIOR INSPECTION WAS PERFORMED AND PASSED. AUDIO SPEAKER RESISTANCE WAS MEASURED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERRORS RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9193667 ↗
  • MalfunctionOct 10, 2019Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9177516 ↗
  • MalfunctionAug 6, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE TEST WAS PERFORMED AND PASSED. A RECEIVER BOOT TEST WAS PERFORMED AND FAILED DUE TO IT WOULD NOT BOOT. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. THE LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8864655 ↗
  • MalfunctionJul 30, 2019Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8842370 ↗
  • MalfunctionJun 21, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT AN ERRPR CALL TECH SUPPORT ERROR ICON WAS DISPLAYED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8723330 ↗
  • MalfunctionJun 6, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWBB1" ERROR AND A "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8674154 ↗
  • MalfunctionMay 29, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8650003 ↗
  • MalfunctionMay 21, 2019Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8627302 ↗
  • MalfunctionMay 20, 2019Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8626222 ↗
  • MalfunctionApr 26, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWRF" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8554723 ↗
  • MalfunctionApr 12, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWRF" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8509464 ↗
  • MalfunctionMar 19, 2019Unexpected Shutdown; Unexpected Shutdown

    THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND FAILED DUE TO WINDOW DAMAGE. A RECEIVER CHARGE AND BOOT WAS PERFORMED AND FAILED DUE TO A CALL TECH SUPPORT ERROR MESSAGE. THE DEVICE WAS OPENED, AND AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE LOG COULD NOT BE DOWNLOADED DUE TO THE CALL TECH SUPPORT ERROR MESSAGE. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8433868 ↗
  • MalfunctionMar 5, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN 'HWBBT' ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8392446 ↗
  • MalfunctionDec 31, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN "ERR HWRF" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8206947 ↗
  • MalfunctionDec 19, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8179745 ↗
  • MalfunctionDec 18, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERR HWRF WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR VALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8173371 ↗
  • MalfunctionNov 15, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8075657 ↗
  • MalfunctionOct 12, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERR HW REF WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7962332 ↗
  • MalfunctionSep 14, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7876804 ↗
  • MalfunctionAug 21, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ISSUE WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7803031 ↗
  • MalfunctionAug 10, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERRHWBBT WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7772485 ↗
  • MalfunctionAug 6, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWRF". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7752848 ↗
  • MalfunctionJul 31, 2018Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7737085 ↗
  • MalfunctionJul 27, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7727008 ↗

FDA clearance and classification

Its record lists premarket approval application P120005, decided on Oct 5, 2012 with FDA decision code APWD (not a standard approval code). Check the FDA record for its current status. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

FDA definition of this device type: The device is a glucose-monitoring device indicated for detecting trends and tracking patterns in adults (age 18 and older) with diabetes. The device is indicated for use as an adjunctive device to complement, not replace, information obtained from standard home glucose monitoring devices. The system aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Common questions about the Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001

Has the Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 50386270000069), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001?

The FDA's MAUDE database lists 126 adverse event reports for this device, received between Feb 2, 2017 and Aug 21, 2020. In those reports, the most common specific device problems are “Device Displays Incorrect Message” (58), “Wireless Communication Problem” (22) and “Unable to Obtain Readings” (17); the most common patient problems are “Hypoglycemia” (2) and “Loss of consciousness” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001?

None of the reports are classified as deaths, 2 as injuries and 124 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dexcom G4 Platinum Replacement Receiver Kit STR-GL-001, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G4 Platinum Replacement Receiver Kit, model STR-GL-001). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page