Skip to content

Dexcom G5 Mobile Receiver Kit (Retail)Recalls & reported problems across all models

Dexcom, Inc. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1,659
Deaths reported
0
Injuries reported
5
Recalls
1

We found 3 models of Dexcom G5 Mobile Receiver Kit (Retail) by Dexcom, Inc. in the FDA's device registry (GUDID). Together they are named in 1,659 adverse event reports in the FDA's MAUDE database, received between Aug 17, 2016 and Aug 3, 2020 — classified as 5 injuries, 1,654 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (917 reports).

1 FDA recall record name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Dexcom G5 Mobile Receiver Kit (Retail) recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 12, 2019Terminated

    It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.

    Z-0929-2020 · Root cause: Software design · names 3 of these models · Recall details · FDA record ↗

Most reported Dexcom G5 Mobile Receiver Kit (Retail) problems

Device problems

  • Device Displays Incorrect Message917
  • Inappropriate or Unexpected Reset241
  • Wireless Communication Problem129
  • Unable to Obtain Readings122
  • Device Inoperable89
  • Defective Alarm81
  • No Audible Alarm30
  • Unexpected Shutdown27

Patient problems

  • No Consequences Or Impact To Patient1,654
  • Hypoglycemia2
  • Loss of consciousness2
  • Fracture, Arm1
  • Hyperglycemia1
  • Malaise1
  • Vomiting1

Dexcom G5 Mobile Receiver Kit (Retail) models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Dexcom G5 Mobile Receiver Kit (Retail) adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Dexcom, Inc. products

All Dexcom, Inc. devices →

Sources and updates