Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNKRecalls & reported problems
Dexcom, Inc. · G5 Mobile Receiver Kit (Retail) Adult Pink · Class 3
All models of Dexcom G5 (21) · Dexcom G5 Mobile Receiver Kit (Retail) (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Dexcom G5 mobile receiver shows glucose readings and alerts from a sensor worn on the body. The 311 reports mostly describe the receiver displaying wrong messages, resetting unexpectedly or stopping working. There are no deaths and 2 injuries reported. A closed recall from July 2019 covers a software problem that could leave the receiver stuck on its startup screen, so patients could not get glucose values or alerts.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 311
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 311 adverse event reports that list this exact device (DI 10386270000269), received between Aug 17, 2016 and Aug 23, 2019 — classified as 2 injuries, 309 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (190 reports).
1 FDA recall record name this device's identifier.
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jul 12, 2019Class IITerminated
It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not to be able to receive glucose values or alerts; 2. Reinitialization of the receiver without user interaction. If this occurs, the receiver will either reboot and operate normally or require the user to press the Select button on the receiver in order to resume normal operation. In this situation, the user is provided an audio and vibratory alert every 5 minutes. Until the select button is pressed, patients will not receive glucose values or alerts.
Z-0929-2020 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Displays Incorrect Message190
- Inappropriate or Unexpected Reset33
- Wireless Communication Problem20
- Defective Alarm19
- Unable to Obtain Readings17
- Device Inoperable15
- No Audible Alarm5
- Unexpected Shutdown4
Patient problems
- No Consequences Or Impact To Patient309
- Fracture, Arm1
- Hypoglycemia1
- Malaise1
Recent adverse event reports
The 25 most recent of 311 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8926396 ↗MalfunctionJun 24, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE TEST WAS PERFORMED AND PASSED. A RECEIVER BOOT TEST WAS PERFORMED AND PASSED. THE LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8727216 ↗MalfunctionJun 6, 2019Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT A "ERR 121" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8674782 ↗MalfunctionMar 5, 2019Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN 'ERR 121' ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8392459 ↗MalfunctionFeb 19, 2019Unexpected Shutdown; Unexpected Shutdown
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8352088 ↗MalfunctionFeb 14, 2019Unexpected Shutdown; Unexpected Shutdown
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8341410 ↗MalfunctionFeb 6, 2019Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8313413 ↗MalfunctionJan 28, 2019Display or Visual Feedback Problem; Display or Visual Feedback Problem
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019 THE RECEIVER DISPLAY WAS FROZEN. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8283628 ↗MalfunctionJan 22, 2019Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, AN "ERR 121" AND "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8269501 ↗MalfunctionNov 2, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 69". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8033323 ↗MalfunctionOct 15, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7966699 ↗MalfunctionOct 9, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7950055 ↗MalfunctionOct 4, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON 09/15/2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7934496 ↗MalfunctionSep 25, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON 09/05/2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7908314 ↗MalfunctionSep 14, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7877143 ↗MalfunctionSep 10, 2018Unexpected Shutdown; Unexpected Shutdown
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7861950 ↗MalfunctionSep 6, 2018Defective Alarm; Defective Alarm
DEXCOM WAS MADE AWARE ON 08/15/2018 THAT ON (B)(6) 2018, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7851701 ↗MalfunctionSep 6, 2018Defective Alarm; Defective Alarm
DEXCOM WAS MADE AWARE ON 08/15/2018 THAT ON (B)(6) 2018, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7851695 ↗MalfunctionAug 27, 2018Unexpected Shutdown; Unexpected Shutdown
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7819305 ↗MalfunctionAug 14, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN ERR121 WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7783104 ↗MalfunctionAug 13, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE PATIENT EXPERIENCED INITIALIZING SCREEN WITHOUT MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7776404 ↗MalfunctionJul 30, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7732047 ↗MalfunctionJul 24, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7716963 ↗MalfunctionJul 12, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7681608 ↗MalfunctionJul 11, 2018Unable to Obtain Readings; Unable to Obtain Readings
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7679682 ↗
FDA clearance and classification
Its record lists premarket approval application P120005, decided on Oct 5, 2012 with FDA decision code APWD (not a standard approval code). Check the FDA record for its current status. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: The device is a glucose-monitoring device indicated for detecting trends and tracking patterns in adults (age 18 and older) with diabetes. The device is indicated for use as an adjunctive device to complement, not replace, information obtained from standard home glucose monitoring devices. The system aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
Common questions about the Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNK
Has the Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNK been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0929-2020 (Class II), began on Jul 12, 2019: It has been reported that use of the mobile receiver with software version SW10617 rev 4.0.1.048 have reported: 1. the receiver becoming stuck on initialization screen when powering on. This will cause patients not… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNK?
The FDA's MAUDE database lists 311 adverse event reports for this device, received between Aug 17, 2016 and Aug 23, 2019. In those reports, the most common specific device problems are “Device Displays Incorrect Message” (190), “Inappropriate or Unexpected Reset” (33) and “Wireless Communication Problem” (20); the most common patient problems are “Fracture, Arm” (1), “Hypoglycemia” (1) and “Malaise” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNK?
None of the reports are classified as deaths, 2 as injuries and 309 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Dexcom G5 Mobile Receiver Kit (Retail) STK-RF-PNK, and is it still sold?
The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G5 Mobile Receiver Kit (Retail), model STK-RF-PNK). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page