Dimension EXL 200 Integrated Chemistry SystemRecalls & reported problems across all models
Siemens Healthcare Diagnostics Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 516
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls
- 1
We found 2 models of Dimension EXL 200 Integrated Chemistry System by Siemens Healthcare Diagnostics Inc. in the FDA's device registry (GUDID). Together they are named in 516 adverse event reports in the FDA's MAUDE database, received between Jul 18, 2017 and Sep 29, 2026 — classified as 1 injury, 515 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (199 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Dimension EXL 200 Integrated Chemistry System recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 28, 2018Terminated
Thermal chamber motors have electrical cables with single insulation rather than the required double insulation. There is no exposed current due to the single insulation that could result in a potential electrical shock
Z-0409-2019 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
Most reported Dimension EXL 200 Integrated Chemistry System problems
Device problems
- Low Test Results199
- High Test Results131
- Incorrect, Inadequate or Imprecise Result or Readings51
- Mechanical Problem38
- Adverse Event Without Identified Device or Use Problem23
- No Apparent Adverse Event12
- Smoking12
- Leak/Splash9
Patient problems
- No Clinical Signs, Symptoms or Conditions312
- No Known Impact Or Consequence To Patient150
- No Consequences Or Impact To Patient45
- Patient Problem/Medical Problem3
- Needle Stick/Puncture2
- Abrasion1
- Conjunctivitis1
- Dehydration1
Dimension EXL 200 Integrated Chemistry System models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Dimension® EXL™ 200 Integrated Chemistry System 10636929ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall373 reports
- Dimension® EXL™ 200 Integrated Chemistry System 10636928ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall143 reports
Recent Dimension EXL 200 Integrated Chemistry System adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Low Test Results
The customer reported that they obtained two erroneously depressed Sodium (Na) results on a patient sample on a Dimension EXL 200 Integrated Chemistry System.The initial result was reported to the physician(s) and was questioned. The sample was repeated on the same system and the result obtained was higher than the initial result but was in the critical range. The repeat result was reported to the physician(s) and was questioned. The physician(s) asked to redraw the sample. A new sample was drawn from the same patient and processed on the same system. The result obtained was higher than the previous results, considered correct and reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed Na results.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928237 ↗MalfunctionSep 25, 2026Low Test Results
The customer reported that they obtained an erroneously depressed Sodium (Na) result on a patient sample on a Dimension EXL 200 Integrated Chemistry System.The initial result was not reported to the physician(s). The sample was repeated on the same Dimension EXL 200 Integrated Chemistry System and higher result was obtained. The repeat result was reported to the physician(s) as correct result.There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed Na result.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895615 ↗MalfunctionSep 11, 2026Low Test Results
The customer reported that an erroneously depressed Sodium (Na) result was obtained on a Dimension EXL 200 Integrated Chemistry System. The sample was repeated on the original system. Additionally, a new sample was drawn from the same patient and processed on the original system. The redrawn sample result was higher than the initial result, considered correct, and reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed Sodium result.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50728740 ↗MalfunctionMay 22, 2026High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED SODIUM (NA) RESULT WAS OBTAINED FOR A PATIENT ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S). THE PHYSICIAN QUESTIONED THE RESULT. THE RESULT WAS REPEATED ON THE SAME INSTRUMENT. THE REPEAT RESULT WAS LOWER THAN THE ERRONEOUS RESULT. THE REPEAT RESULT WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257106 ↗MalfunctionMay 22, 2026Low Test Results
THE CUSTOMER REPORTED THAT AN ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT WAS OBTAINED FOR A PATIENT SAMPLE ON THE DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS QUESTIONED. THE SAMPLE WAS REPEATED ON THE SAME DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE REPEAT RESULT WAS HIGHER THAN THE INITIAL RESULT AND WAS CONSIDERED CORRECT. THE PATIENT WAS SENT TO THE EMERGENCY ROOM (ER) DUE TO THE DEPRESSED RESULT. A NEW SAMPLE WAS DRAWN AGAIN IN THE ER AND TESTED; THE RESULT OBTAINED WAS SIMILAR TO THE PREVIOUS REPEAT RESULT. BOTH REPEAT RESULTS WERE CONSISTENT AND CONSIDERED CORRECT. THE REDRAWN SAMPLE RESULT WAS REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25252674 ↗MalfunctionMay 15, 2026Low Test Results
THE CUSTOMER REPORTED A FALSELY DEPRESSED SODIUM (NA) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON THE SAME INSTRUMENT. THE REPEAT RESULT WAS HIGHER THAN THE ERRONEOUS RESULT. THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25184842 ↗MalfunctionMay 5, 2026Low Test Results
THE CUSTOMER REPORTED THAT A FALSELY DEPRESSED POTASSIUM (K) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULT. THE SAMPLE WAS REPEATED ON THE ORIGINAL INSTRUMENT FOR TROUBLESHOOTING PURPOSES. THEN THE SAME SAMPLE WAS REPEATED ON NON-SIEMENS INSTRUMENT. THE REPEAT RESULT FROM NON-SIEMENS INSTRUMENT RECOVERED HIGHER THAN THE INITIAL RESULT AND WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO A FALSELY DEPRESSED K RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25078915 ↗MalfunctionApr 1, 2026Low Test Results
A FALSELY LOW SODIUM TEST RESULT WAS OBTAINED FOR ONE PATIENT SAMPLE FROM A DIMENSION EXL 200 ANALYZER. THE INITIAL TEST RESULT WAS NOT RELEASED TO A PHYSICIAN. THE SAME SAMPLE WAS REPEATED ON THE SAME AND AN ALTERNATE DIMENSION EXL 200 ANALYZER. THE TEST RESULT FROM THE ALTERNATE ANALYZER WAS HIGHER, CONSIDERED CORRECT AND RELEASED TO A PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755352 ↗MalfunctionMar 9, 2026Low Readings
THE CUSTOMER REPORTED A FALSELY DEPRESSED N-TERMINAL PRO-BRAIN NATRIURETIC PEPTIDE (NTP) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON THE SAME INSTRUMENT THE REPEAT RESULT WAS HIGHER THAN THE ERRONEOUS RESULT. THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24543617 ↗MalfunctionFeb 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED A DISCORDANT RESULT WAS OBTAINED FOR A GLUCOSE (GLUC) RESULT ON 1 PATIENT SAMPLE ON A DIMENSION EXL 200 ANALYZER. THE RESULT WAS FLAGGED WITH AN ABNORMAL REACTION AND WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON THE SAME INSTRUMENT. THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ABNORMAL REACTION ERRORS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24349635 ↗MalfunctionFeb 11, 2026Low Readings
THE CUSTOMER REPORTED MULTIPLE FALSELY DEPRESSED TOTAL BILIRUBIN (TBI) RESULTS WERE OBTAINED ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON THE SAME DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE FOURTH RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315151 ↗MalfunctionFeb 11, 2026Low Readings
THE CUSTOMER REPORTED A FALSELY DEPRESSED TOTAL BILIRUBIN (TBI) RESULT WAS OBTAINED ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON THE SAME DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE REPEAT RESULT WAS REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315071 ↗MalfunctionFeb 3, 2026High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED TACROLIMUS (TAC) RESULT WAS OBTAINED ON 1 PATIENT SAMPLE ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE SAMPLE WAS INITIALLY TESTED ON AN ALTERNATE DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE SAMPLE WAS REPEATED ON THE DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE REPEAT RESULT WAS HIGHER THAN THE INITIAL RESULT AND CONSIDERED ERRONEOUS. THE REPEAT, ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE INITIAL RESULT WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED TAC RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24247596 ↗MalfunctionFeb 3, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED TACROLIMUS (TAC) RESULTS WERE OBTAINED ON TWO PATIENT SAMPLES ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ALTERNATE DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE REPEAT RESULTS WERE LOWER THAN THE ERRONEOUS RESULTS. THE REPEAT RESULTS WERE REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24246993 ↗MalfunctionFeb 3, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED TACROLIMUS (TAC) RESULTS WERE OBTAINED ON FIVE PATIENT SAMPLES ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUS RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ALTERNATE INSTRUMENT. THE REPEAT RESULTS WERE LOWER THAN THE ERRONEOUS RESULTS. THE REPEAT RESULTS WERE REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24246766 ↗
Other Siemens Healthcare Diagnostics Inc. products
- Atellica IM TNIH2 models · 629 reports
- Atellica IM CA 19-93 models · 121 reports · 3 recalls
- ADVIA Centaur XP Analyzer2 models · 269 reports
- ADVIA Centaur CA 19-92 models · 124 reports · 2 recalls
- Atellica IM eE23 models · 65 reports · 3 recalls
- RAPIDPoint 500 Measurement Cartridge4 models · 3 reports · 4 recalls
- ADVIA Centaur eE23 models · 22 reports · 3 recalls
- Atellica IM Fol3 models · 10 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026