Atellica IM eE2Recalls & reported problems across all models
Siemens Healthcare Diagnostics Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Atellica IM eE2 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 65
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 3
We found 3 models of Atellica IM eE2 by Siemens Healthcare Diagnostics Inc. in the FDA's device registry (GUDID). Together they are named in 65 adverse event reports in the FDA's MAUDE database, received between Jun 19, 2020 and Apr 11, 2026 — classified as 9 injuries, 56 malfunctions.
The most frequently reported specific device problem is “High Test Results” (65 reports).
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Atellica IM eE2 recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 3, 2022Open, Classified
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Z-1022-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Mar 3, 2022Open, Classified
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Z-1021-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Mar 3, 2022Open, Classified
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Z-1020-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Atellica IM eE2 problems
Device problems
- High Test Results65
Patient problems
- No Clinical Signs, Symptoms or Conditions40
- Insufficient Information25
- No Known Impact Or Consequence To Patient2
Atellica IM eE2 models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Atellica IM eE2 10995561RADIOIMMUNOASSAY, ESTRADIOL1 recall61 reports
- Atellica IM eE2 10995562RADIOIMMUNOASSAY, ESTRADIOL1 recall4 reports
- Atellica IM eE2 11206250RADIOIMMUNOASSAY, ESTRADIOL1 recall0 reports
Recent Atellica IM eE2 adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 11, 2026High Test Results
THE CUSTOMER REPORTED OBSERVATION OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING WHEN USING LOT 106. THE CUSTOMER PERFORMED A REVIEW OF EE2 RESULTS AS A PART OF ROUTINE RESET. DURING THIS REVIEW, INITIAL ELEVATED RESULTS WERE OBTAINED FOR TWO (2) PATIENT SAMPLES IN QUESTION. THE ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S) BUT WEREN'T QUESTIONED. SUBSEQUENTLY, THE SAME SAMPLES WERE RETESTED ON ALTERNATE ANALYZER(S) WHICH PRODUCED LOWER RESULTS. THE LOWER RESULTS WERE CONSIDERED CORRECT BASED ON THE CLINICAL ASSESSMENT. HOWEVER, NO CORRECTED REPORTS WERE ISSUED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24851056 ↗MalfunctionApr 11, 2026High Test Results
THE CUSTOMER REPORTED OBSERVATION OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING WHEN USING LOT 106. THE CUSTOMER PERFORMED A REVIEW OF EE2 RESULTS AS A PART OF ROUTINE RESET. DURING THIS REVIEW, INITIAL ELEVATED RESULTS WERE OBTAINED FOR FIVE (5) PATIENT SAMPLES IN QUESTION. THE ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S) BUT WEREN'T QUESTIONED. SUBSEQUENTLY, THE SAME SAMPLES WERE RETESTED ON ALTERNATE ANALYZER(S) WHICH PRODUCED LOWER RESULTS. THE LOWER RESULTS WERE CONSIDERED CORRECT BASED ON THE CLINICAL ASSESSMENT. HOWEVER, A CORRECTED REPORT WAS ISSUED TO THE PHYSICIAN FOR ONLY ONE PATIENT SAMPLE. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24850206 ↗MalfunctionMar 30, 2026High Test Results
THE CUSTOMER OBTAINED ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) LOT 106 RESULTS ON(B)(6) 2025 ON SERUM SAMPLES FROM A POSTMENOPAUSAL 63-YEAR-OLD FEMALE. THE ELEVATED ATELLICA IM EE2 RESULTS WERE NOT QUESTIONED AT THE TIME AS THE PATIENT WAS REFERRED TO THIS CUSTOMER SITE FOR CONFIRMATION TESTING ONLY. THE PATIENT WAS LATER TESTED AT A HOSPITAL LABORATORY ((B)(6) 2026) WITH AN ALTERNATE ESTRADIOL METHOD AND A POSITRON EMISSION TOMOGRAPHY-COMPUTED TOMOGRAPHY (PET-CT) WAS PERFORMED WITH NO FINDINGS. THE PATIENT ULTIMATELY UNDERWENT OOPHORECTOMY ((B)(6) 2026) AND THE PHYSICIAN WHO PERFORMED THE POST-OPERATIVE CHECK-UP ON THE PATIENT ((B)(6) 2026) BELIEVES THAT THE OOPHORECTOMY WAS AN UNNECESSARY MEDICAL PROCEDURE. SUBSEQUENTLY, SIEMENS HAS BEEN NOTIFIED THAT THE RESULTS OBTAINED BY THIS CUSTOMER ON (B)(6) 2025 ARE NOW IN QUESTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24711772 ↗MalfunctionMar 25, 2026High Test Results
SIEMENS WAS NOTIFIED OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) LOT 107 RESULTS OBTAINED ON SERUM SAMPLES FROM A POSTMENOPAUSAL 63-YEAR-OLD FEMALE. THESE RESULTS WERE OBTAINED PRE AND POST OOPHORECTOMY AND WERE DISCORDANT RELATIVE TO LOWER ALTERNATE METHOD TEST RESULTS AND OTHER DIAGNOSTIC TESTING (POSITRON EMISSION TOMOGRAPHY-COMPUTED TOMOGRAPHY (PET-CT) WITH NO FINDINGS IN (B)(6) 2025). THE ATELLICA IM EE2 RESULT OBTAINED PRIOR TO THE OOPHORECTOMY WAS NOT QUESTIONED AT THE TIME. THE PHYSICIAN WHO PERFORMED THE POST-OPERATIVE CHECK-UP ON THE PATIENT ((B)(6) 2026) QUESTIONED THE RESULTS AND BELIEVES THAT THE OOPHORECTOMY WAS AN UNNECESSARY MEDICAL PROCEDURE (OOPHORECTOMY, (B)(6) 2026).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24684147 ↗MalfunctionAug 1, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A DIFFERENCE IN SERUM PATIENT SAMPLE RESULTS ON A PATIENT WITH ATELLICA IM ENHANCED ESTRADIOL (EE2) LOT 105 DEPENDING ON WHICH ANCILLARY REAGENT PACK IS USED. INITIAL RESULT USING ONE ANCILLARY PACK WAS HIGHER THAN A RETEST RESULT OBTAINED WITH A DIFFERENT ANCILLARY PACK. THE LOWER RETEST RESULT WAS REPORTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT OR USER INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22689153 ↗MalfunctionFeb 7, 2025High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING WHEN USING EE2 LOT 103. THE CUSTOMER IDENTIFIED QUALITY CONTROL (QC) RESULTS ABOVE EXPECTED RANGES, AND REPEAT-TESTED 172 SAMPLES WHICH HAD BEEN TESTED FOR EE2 SINCE LAST PASSING QC. OBSERVED DIFFERENCES BETWEEN INITIAL AND REPEAT RESULTS RANGED FROM -35% TO +107%; FOR 44 SAMPLES, THE DIFFERENCE WAS 30% OR GREATER. ONLY ONE SAMPLE DEMONSTRATED INITIAL RESULT DEPRESSION. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIATION WITH THE OBSERVED DISCORDANCE. RESULT REPORTS WERE CORRECTED FOR ONLY THREE OF THE PATIENTS (SAMPLE IDS (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21333461 ↗MalfunctionJan 16, 2025High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING. INITIAL ELEVATED RESULTS WERE OBTAINED FOR FIVE PATIENT SAMPLES. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN, WHO DID NOT QUESTION THE RESULTS. THE SAME SAMPLES WERE REPEATED ON AN ALTERNATE ATELLICA IM ANALYZER AND OBTAINED LOWER RESULTS WHICH WERE CONSIDERED CORRECT AND ISSUED CORRECTED REPORTS TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION, DELAY IN REPORTING THE RESULTS, OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21167808 ↗MalfunctionMay 22, 2024High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING. THE CUSTOMER OBSERVED EE2 QUALITY CONTROL (QC) RESULTS ABOVE ACCEPTANCE RANGES, AND REPEAT-TESTED SAMPLES WHICH HAD BEEN TESTED WHEN QC WAS OUT. THE INITIAL ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEAT-TESTED AND LOWER RESULTS WERE OBTAINED. ACCORDING TO THE CUSTOMER, NO CORRECTED REPORTS WERE ISSUED SINCE THE RESULTS WERE NOT QUESTIONED BY THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED EE2 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19375193 ↗MalfunctionMay 22, 2024High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF AN ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULT WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING. THE CUSTOMER OBSERVED EE2 QUALITY CONTROL (QC) RESULTS ABOVE ACCEPTANCE RANGES, AND REPEAT-TESTED SAMPLES WHICH HAD BEEN TESTED WHEN QC WAS OUT. THE INITIAL ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEAT-TESTED AND A LOWER RESULT WAS OBTAINED. ACCORDING TO THE CUSTOMER, NO CORRECTED REPORT WAS ISSUED SINCE THE INITIAL RESULT WAS NOT QUESTIONED BY THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED EE2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19375133 ↗MalfunctionApr 12, 2024High Test Results; High Test Results
THE CUSTOMER REPORTED OBSERVATION OF DISCORDANT ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS OBTAINED ON TWO PATIENT SAMPLES WHILE QUALITY CONTROL (QC) RESULTS WERE OUTSIDE OF RANGES. THE INITIAL ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S); THESE RESULTS WERE QUESTIONED BY THE PHYSICIAN(S). THE SAME SAMPLES WERE RETESTED ON AN ALTERNATE ATELLICA IM AND LOWER RESULTS WERE OBTAINED. THE REPEAT RESULTS WERE REPORTED AS CORRECT RESULTS TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED ENHANCED ESTRADIOL (EE2) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19095161 ↗MalfunctionMar 23, 2024High Test Results; High Test Results
CUSTOMER REPORTED OBSERVATION OF A DISCORDANT ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULT ON ONE PATIENT SAMPLE. THE INITIAL ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S); THIS RESULT WAS NOT QUESTIONED. THE SAME SAMPLE WAS RETESTED ON AN ALTERNATE ATELLICA IM ANALYZER AND OBTAINED A LOWER RESULT. THE REPEAT RESULT WAS REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED EE2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965534 ↗MalfunctionMar 23, 2024High Test Results; High Test Results
CUSTOMER REPORTED OBSERVATION OF DISCORDANT ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS ON TWO PATIENT SAMPLES. THE INITIAL ELEVATED RESULTS WERE REPORTED TO THE PHYSICIAN(S); THESE RESULTS WERE NOT QUESTIONED BY THE PHYSICIAN(S). THE SAME SAMPLES WERE RETESTED ON AN ALTERNATE ATELLICA IM AND LOWER RESULTS WERE OBTAINED. THE REPEAT RESULTS WERE REPORTED AS THE CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED EE2 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965501 ↗MalfunctionMar 23, 2024High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF DISCORDANT ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULTS FOR TWO PATIENT SAMPLES WHICH WERE CONSIDERED DISCORDANT RELATIVE TO REPEAT TESTING ON AN ALTERNATE ATELLICA IM ANALYZER WHICH PRODUCED LOWER RESULTS. THE INITIAL ELEVATED RESULTS WERE REPORTED AND NOT QUESTIONED BY THE PHYSICIAN(S). A CORRECTED REPORT WAS ISSUED TO THE PHYSICIAN(S) BASED ON REPEAT TESTING. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED ENHANCED ESTRADIOL (EE2) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965378 ↗MalfunctionDec 18, 2023High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF AN ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. EE2 REAGENT LOT 097 WAS IN USE AT THE TIME. AN INITIAL EE2 TEST WAS PERFORMED ON A 62-YEAR-OLD MENOPAUSAL FEMALE PATIENT IN NOVEMBER, AND AN ELEVATED RESULT WAS OBTAINED. THE PHYSICIAN QUESTIONED THE HIGH RESULT. THIS DISCORDANT OBSERVATION IS ADRESSED IN THIS REPORT. A NEW SAMPLE WAS DRAWN AND MEASURED ABOUT THREE WEEKS LATER, AND ANOTHER ELEVATED ATELLICA IM EE2 RESULT WAS OBTAINED. THIS SECOND DISCORDANT OBSERVATION IS REPORTED SEPARATELY. THIS SAMPLE WAS SENT TO AN EXTERNAL LAB FOR TESTING BY AN ALTERNATE METHOD, AND A MUCH LOWER RESULT WAS PRODUCED; THIS LOWER RESULT WAS ACCEPTED AS CORRECT. THE SECOND SAMPLE WAS LATER RE-TESTED USING THE SAME ATELLICA IM ANALYZER TO INVESTIGATE. A SIMILAR ELEVATED RESULT WAS OBTAINED, INDICATING THAT THE INITIAL ELEVATED OBSERVATION WAS REPRODUCIBLE. WHEN THIS SAMPLE WAS TESTED USING A SECOND ALTERNATE METHOD, A LOW RESULT ¿ CONCORDANT WITH THAT FROM THE FIRST ALTERNATE METHOD ¿ WAS PRODUCED. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIATION W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18346608 ↗MalfunctionDec 18, 2023High Test Results; High Test Results
THE CUSTOMER REPORTS OBSERVATION OF AN ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. EE2 REAGENT LOT 097 WAS IN USE AT THE TIME. AN INITIAL EE2 TEST WAS PERFORMED ON A 62-YEAR-OLD MENOPAUSAL FEMALE PATIENT IN NOVEMBER, AND AN ELEVATED RESULT WAS OBTAINED. THE PHYSICIAN QUESTIONED THE HIGH RESULT. [THIS DISCORDANT OBSERVATION IS REPORTED SEPARATELY.] A NEW SAMPLE WAS DRAWN AND MEASURED ABOUT THREE WEEKS LATER, AND ANOTHER ELEVATED ATELLICA IM EE2 RESULT WAS OBTAINED. [THIS SECOND DISCORDANT OBSERVATION IS ADDRESSED IN THIS REPORT.] THIS SAMPLE WAS SENT TO AN EXTERNAL LAB FOR TESTING BY AN ALTERNATE METHOD, AND A MUCH LOWER RESULT WAS PRODUCED; THIS LOWER RESULT WAS ACCEPTED AS CORRECT. THE SECOND SAMPLE WAS LATER RE-TESTED USING THE SAME ATELLICA IM ANALYZER TO INVESTIGATE. A SIMILAR ELEVATED RESULT WAS OBTAINED, INDICATING THAT THE INITIAL ELEVATED OBSERVATION WAS REPRODUCIBLE. WHEN THIS SAMPLE WAS TESTED USING A SECOND ALTERNATE METHOD, A LOW RESULT ¿ CONCORDANT WITH THAT FROM THE FIRST ALTERNATE METHOD ¿ WAS PRODUCED. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS IN ASSOCIAT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18346598 ↗
Other Siemens Healthcare Diagnostics Inc. products
- Atellica IM TNIH2 models · 629 reports
- Dimension EXL 200 Integrated Chemistry System2 models · 516 reports · 1 recall
- Atellica IM CA 19-93 models · 121 reports · 3 recalls
- ADVIA Centaur XP Analyzer2 models · 269 reports
- ADVIA Centaur CA 19-92 models · 124 reports · 2 recalls
- RAPIDPoint 500 Measurement Cartridge4 models · 3 reports · 4 recalls
- ADVIA Centaur eE23 models · 22 reports · 3 recalls
- Atellica IM Fol3 models · 10 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page