DLP Aortic Root Cannula with Flow-GuardRecalls & reported problems across all models
Medtronic, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one DLP Aortic Root Cannula with Flow-Guard model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 2
We found 2 models of DLP Aortic Root Cannula with Flow-Guard by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Oct 9, 2024 and Feb 9, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (3 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
DLP Aortic Root Cannula with Flow-Guard recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 5, 2025Open, Classified
Unexpected loose material in the male luer used in the aortic root cannula has been identified. Potential harms when identified prior to use - procedure delay. Potential harm if not identified and used - stroke (reversible and irreversible).
Z-1306-2025 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
- Oct 3, 2024Open, Classified
There is the potential for a potential sterility breach for three lots of the DLP Aortic Root Cannula due to a misaligned seal
Z-0289-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported DLP Aortic Root Cannula with Flow-Guard problems
Device problems
- Mechanical Problem3
- Manufacturing, Packaging or Shipping Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
DLP Aortic Root Cannula with Flow-Guard models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- DLP® Aortic Root Cannula with Flow-Guard™ 11012CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11014CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 recall0 reports
Recent DLP Aortic Root Cannula with Flow-Guard adverse event reports
The 5 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionFeb 9, 2026Material Deformation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP AORTIC ROOT CANNULA, IT WAS REPORTED THAT UPON INSPECTION BEFORE USE, THE NEEDLE WAS FOUND TO BE BENT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS KINKED. THE PUMP WAS LOCATED ON THE RIGHT SIDE OF THE PATIENT. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292866 ↗MalfunctionJan 23, 2025Mechanical Problem; Mechanical Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP AORTIC ROOT CANNULA, IT WAS REPORTED THAT THE DEVICE HAD DIFFICULTY LOCKING PROPERLY. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21219285 ↗MalfunctionJan 13, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT PRIOR TO USE OF A MINIMALLY INVASIVE AORTIC ROOT CANNULA THERE WAS AN ALLEGED PRODUCT ISSUE, AS THE STYLET COULD NOT LOCK. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21136478 ↗MalfunctionDec 6, 2024Mechanical Problem; Mechanical Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP AORTIC ROOT CANNULA, IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO LOCK. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20864618 ↗MalfunctionOct 9, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP AORTIC ROOT CANNULA, IT WAS REPORTED THAT THE PACKAGE WAS OPEN. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WERE NO MISSING OR WRONG DEVICES OR COMPONENTS IN THE PACKAGE. THE STERILE SEAL WAS NOT INTACT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER HAS MORE DEVICES FROM THE COMPLAINT LOT AT THEIR FACILITY; HOWEVER, THEY DID NOT OBSERVE THIS ISSUE WITH ANY OF THE OTHER DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20405726 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page