EKOS +Recalls & reported problems across all models
Boston Scientific Corporation · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one EKOS + model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 1
We found 4 models of EKOS + by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 6 adverse event reports in the FDA's MAUDE database, received between Oct 18, 2023 and May 2, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
EKOS + recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 13, 2022Open, Classified
Consoles were shipped without an instruction for use (IFU) or with the incorrect IFU.
Z-0149-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported EKOS + problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Leak/Splash3
- Crack1
- Fluid/Blood Leak1
- Material Deformation1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
EKOS + models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EKOS™ EKOS Control Unit 4.0Catheter, Continuous Flush1 recall1 report
- EKOS™ + H74939605106120Catheter, continuous flush3 reports
- EKOS™ + H74939605135160Catheter, continuous flush1 report
- EKOS™ + H74939605135200Catheter, continuous flush1 report
Recent EKOS + adverse event reports
The 6 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2025Crack; Fluid/Blood Leak
ONE OF THE CORDS FOR THE EKOS [EKOSONIC ENDOVASCULAR SYSTEM] HAD A CRACK IN IT AND WAS LEAKING FLUID AT ONE OF THE CONNECTIONS. WE HAD TO OPEN A NEW PRODUCT, TAKE OUT THE BROKEN ONE AND RE-INSERT THE NEW ONE. NO HARM TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21950048 ↗MalfunctionAug 28, 2024Output Problem; Output Problem
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. DURING A BILATERAL EKOS PROCEDURE, THE CONTROL UNIT 4.0 STARTED ALARMING WITH AN E311 ERROR ON BOTH CATHETERS. TROUBLESHOOTING WAS PERFORMED BUT WAS UNABLE TO RESOLVE THE ALARM. THE EKOS THERAPY WAS DISCONTINUED APPROXIMATELY THREE HOURS AFTER THE TREATMENT STARTED. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20089455 ↗InjuryJul 22, 2024Leak/Splash; Adverse Event Without Identified Device or Use Problem; Leak/Splash; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DRUG PORT LEAKED, REQUIRING AN ADDITIONAL DEVICE. ONE 106CM 12 CM EKOS PLUS CATHETER AND ONE 135CM 16 CM EKOS PLUS CATHETER WERE SELECTED FOR USE. DURING THE PROCEDURE, BOTH EKOS CATHETERS WERE PLACED IN THE PATIENT AND HOOKED UP TO THE APPROPRIATE PUMPS. AFTER EVERYTHING WAS CONNECTED, IT WAS DETERMINED THAT ONE OF THE CATHETERS HAD MOVED INSIDE THE PATIENT AND NEEDED TO BE REPOSITIONED. ALL LINES WERE THEN DISCONNECTED, THE CATHETER WAS REPOSITIONED AND EVERYTHING WAS HOOKED BACK UP AGAIN. THE PATIENT WAS MOVED TO THE INTENSIVE CARE UNIT (ICU) AND IT WAS NOTICED THAT BOTH DRUG LUMEN PORTS WERE LEAKING FLUID AT THE CONNECTOR. THE PATIENT WAS BROUGHT BACK TO THE LAB AND THE ORIGINAL CATHETERS WERE REMOVED AND TWO NEW EKOS CATHETERS WERE PLACED WITHOUT ISSUE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19801112 ↗InjuryJul 22, 2024Leak/Splash; Adverse Event Without Identified Device or Use Problem; Leak/Splash; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DRUG PORT LEAKED, REQUIRING AN ADDITIONAL DEVICE. ONE 106CM 12 CM EKOS PLUS CATHETER AND ONE 135CM 16 CM EKOS PLUS CATHETER WERE SELECTED FOR USE. DURING THE PROCEDURE, BOTH EKOS CATHETERS WERE PLACED IN THE PATIENT AND HOOKED UP TO THE APPROPRIATE PUMPS. AFTER EVERYTHING WAS CONNECTED, IT WAS DETERMINED THAT ONE OF THE CATHETERS HAD MOVED INSIDE THE PATIENT AND NEEDED TO BE REPOSITIONED. ALL LINES WERE THEN DISCONNECTED, THE CATHETER WAS REPOSITIONED AND EVERYTHING WAS HOOKED BACK UP AGAIN. THE PATIENT WAS MOVED TO THE INTENSIVE CARE UNIT (ICU) AND IT WAS NOTICED THAT BOTH DRUG LUMEN PORTS WERE LEAKING FLUID AT THE CONNECTOR. THE PATIENT WAS BROUGHT BACK TO THE LAB AND THE ORIGINAL CATHETERS WERE REMOVED AND TWO NEW EKOS CATHETERS WERE PLACED WITHOUT ISSUE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19793615 ↗MalfunctionJan 9, 2024Leak/Splash; Material Deformation; Leak/Splash; Material Deformation
IT WAS REPORTED THAT THE DRUG PORT WAS CRACKED AND LEAKING. AN EKOS+ DEVICE WAS SELECTED FOR USE TO TREAT A HEAVY ARTERIAL THROMBUS BURDEN. THE DEVICE WAS PLACED WITHOUT ISSUE AND WAS INTENDED TO DELIVERY 0.5MG OF LYTIC THERAPY FOR 24 HOURS. THE PATIENT WAS TRANSFERRED TO THE ICU. AT 9:00PM, A POOL OF SEROSANGUINOUS FLUID WAS OBSERVED AROUND THE CATHETER. UPON FURTHER INSPECTION, IT WAS NOTED THAT THERE WAS A LEAK AT THE DRUG CONNECTION TO THE STOPCOCK. A CRACK ON THE CATHETER WAS OBSERVED AT THIS LOCATION. THE CRACK WAS UNSUCCESSFULLY ATTEMPTED TO BE SEALED. TO BYPASS THE ISSUE, THE TPA LINE WAS CONNECTED TO THE COOLANT LINE TO CONTINUE EKOS THERAPY. NO FURTHER ISSUES WERE REPORTED WITH THE DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18477100 ↗MalfunctionOct 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 28SEP2023. AN EKOS+ ENDOVASCULAR DEVICE WAS RETURNED TO THE COMPLAINT INVESTIGATION SITE WITHOUT AN ASSOCIATED COMPLAINT. DEVICE ANALYSIS REVEALED THAT THE COOLANT AND DRUG LUERS WERE CRACKED AND LEAKING. BASED ON THE CONDITION OF THE RETURNED DEVICE, THE DEVICE WAS USED DURING A PROCEDURE. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17962443 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page