ENTERRA IIRecalls & reported problems across all models
Medtronic, Inc. · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1,224
- Deaths reported
- 18
- Injuries reported
- 796
- Recalls
- 0
We found 4 models of ENTERRA II by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 1,224 adverse event reports in the FDA's MAUDE database, received between Nov 8, 2016 and Aug 5, 2025 — classified as 18 deaths, 796 injuries, 410 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (244 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported ENTERRA II problems
Device problems
- Adverse Event Without Identified Device or Use Problem313
- Insufficient Device Problem Information244
- Inappropriate/Inadequate Shock/Stimulation183
- Migration or Expulsion of Device104
- Device Operates Differently Than Expected95
- High impedance77
- Battery Problem69
- Electromagnetic Compatibility Problem59
Patient problems
- Electric Shock223
- Pain213
- Vomiting178
- Nausea162
- No Known Impact Or Consequence To Patient131
- No Clinical Signs, Symptoms or Conditions122
- Complaint, Ill-Defined119
- Insufficient Information107
ENTERRA II models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ENTERRA® II 37800intestinal Stimulator1 death544 reports
- Enterra® II 37800intestinal Stimulator16 deaths359 reports
- Enterra® II 37800intestinal Stimulator1 death248 reports
- ENTERRA® II 37800intestinal Stimulator73 reports
Recent ENTERRA II adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryAug 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT STATES SHE HAS AN ENTERRA DEVICE IN HER STOMACH, AND IT'S GIVING HER EXTREME PAIN AND SHE WOULD LIKE IT REMOVED TODAY. PATIENT REPORTING SIDE EFFECTS POST IMPLANT. PATIENT REPORTS THAT HER STOMACH IS GIVING HER GRIEF AND SHE HAS BEEN HAVING ACHES, PAINS, JOLTS SINCE HER DEVICE WAS PLACED. SHE WANTS IT OUT OF HER BODY ASAP. SHE REPORTS THAT SHE HAS NOTIFIED HER HCP AND THAT PROVIDER ADJUSTED IT FIRST, THEN LATER TURNED IT OFF. HOWEVER, SHE REPORTS THAT SHE CONTINUES TO EXPERIENCE THESE SYMPTOMS. SHE HAS CALLED FOR A FOLLOW UP APPT AND STATES THAT SHE IS SCHEDULED FOR (B)(6) WHICH IS THE EARLIEST AVAILABLE. ENCOURAGED PATIENT TO CONTACT OFFICE TO EXPLAIN HER SYMPTOMS AND ASK ABOUT SOONER APPT AVAILABILITY.
Patient problems: Electric Shock
MAUDE report 22718951 ↗InjuryJul 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT CALLED IN AND REPORTED THAT THEY HAD A RETURN OF N/V SYMPTOMS, THEY ARE HOME SICK AND NOT EATING. IF THEY EAT, THEY ARE NAUSEOUS AND VOMITING. PATIENT SAID THEY THINK THE BATTERY IS DEAD AND THE DEVICE HAD LASTED FOR 2 YEARS AND THEY THOUGHT IT WOULD LAST 5 YEARS.
Patient problems: Failure of Implant
MAUDE report 22666425 ↗InjuryJul 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT REPORTS THAT SHE WAS IMPLANTED ON (B)(6) 2022 BY DR (B)(6). STATES SHE HAS BEEN HAVING RECURRENCE OF SYMPTOMS AND BURNING NEAR DEVICE SITE. SHE DENIES HAVING A DEVICE CARD AND NO IMPLANT RECORD FOUND IN CRM. SHE PLANS TO F/U WITH SURGEON. PATIENT EXPLANTED DUE TO DISCOMFORT AND RETURN OF N/V SYMPTOMS.
Patient problems: Failure of Implant
MAUDE report 22650370 ↗MalfunctionJul 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT/PHYSICIAN CALLED TO REQUEST THE DEVICE BE INTERROGATED. WHILE THERE THE PATIENT REPORTED AN INCREASE IN NAUSEA AND THE UPON INTERROGATION THE DEVICE SHOWED HIGH IMPEDANCE. PATIENT IS BEING REFERRED TO DR JOHNSON MANN, SURGEON. PATIENT REPORTS NO TRAUMA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22649954 ↗MalfunctionJul 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
SHOCKING AT LEADS SITE. THE PATIENT REPORTED EXPERIENCING SHOCK AT THE SITE OF LEAD INSERTION. SURGEON AND PATIENT OPTED FOR A FULL EXPLANT AND DECIDED NOT TO REIMPLANT DUE TO THE PATIENT BEING ON LOW SETTINGS AND NO LONGER NEEDING THE DEVICE, WITH NO SYMPTOMS OF GP.
Patient problems: Electric Shock
MAUDE report 22648884 ↗InjuryJul 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE SURGEON REPORTED THE ISSUE: COMPLETE SYSTEM REMOVAL DUE TO LEAD EROSION AND CHRONIC SEROMA OF THE GENERATOR POCKET. NO NEW SYSTEM WAS IMPLANTED AT THAT TIME, THE PROVIDER AND PATIENT WILL DECIDE AFTER HEALING IF A NEW SYSTEM WILL BE IMPLANTED.
Patient problems: Erosion
MAUDE report 22645672 ↗InjuryJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
LEADS INADVERTENTLY CUT DURING UNRELATED GP SURGERY (KIDNEY TRANSPLANT). PATIENT SYMPTOMS HAVE BEEN WORSENING EXPONENTIALLY IN THE LAST 6 MONTHS WITHOUT THERAPY (POST LEADS DAMAGE), WITH THE PATIENT BEING ADMITTED TO THE HOSPITAL AND REQUIRING IV FLUIDS INFUSION AT HOME. DURING A KIDNEY TRANSPLANT SURGERY, BOTH LEADS WERE INADVERTENTLY CUT. THE TRANSPLANT TEAM INFORMED THE IMAGING PHYSICIAN. REPLACEMENT WAS DELAYED FOR THE PATIENT TO RECOVER FROM THE TRANSPLANT. A NEW ENTERRA SYSTEM WAS IMPLANTED ON (B)(6) 2025. THE PATIENT TOLERATED THE PROCEDURE WELL, WHICH WAS PERFORMED OPEN DUE TO THE PRESENCE OF A LARGE AMOUNT OF ADHESIONS. PATIENT HAD LEADS CUT ACCIDENTALLY DURING A NON-RELATED PROCEDURE; LATER REVISED TO REPLACE ENTIRE SYSTEM.
Patient problems: Failure of Implant
MAUDE report 22644347 ↗InjuryJul 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE DEVICE IS SHOCKING THE PATIENT. PATIENT LEFT ER WITHOUT TREATMENT; NO FURTHER ACTION TO BE TAKEN UNLESS PATIENT RETURNS TO ER.
Patient problems: Electric Shock
MAUDE report 22510235 ↗InjuryJul 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PRODUCT RETURNED TO MEDTRONIC FOR ANALYSIS - REPORTED IN MEDTRONIC DATABASE. PATIENT EXPLANTED (B)(6) 2023.
Patient problems: Failure of Implant
MAUDE report 22509805 ↗InjuryJul 10, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
FROM THE CALL CENTER: MY HUSBAND NEEDS HIS STIMULATOR INTERROGATED. IT'S BEEN 10 YEARS SINCE IT WAS PLACED IN SEATTLE. HE'S BEEN HAVING ISSUES SINCE (B)(6) & BEEN IN & OUT OF HOSPITAL. HE'S IN SEVERE PAIN, NAUSEOUS, VOMITING & DISORIENTED. PATIENT NEEDED NEW BATTERY; EXPLANTED BATTERY WAS DISPOSED THEREFORE NO ANALYSIS POSSIBLE.
Patient problems: Failure of Implant
MAUDE report 22469009 ↗InjuryJul 7, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT CALLED THE PATIENT LIAISON LINE AND LEFT A VOICEMAIL STATING, "I HAD AN ENTERRA THERAPY PLACED ON (B)(6) AND WAS DOING FINE BUT STILL FEELING THE STIMULATION. SO WENT DOWN THE VOLTAGE A WEEK AND A HALF AGO, FEW DAYS LATER I SORT OF FEELING THE FIRING, THE STIMULATION AGAIN. PUT ME ON MUSCLE RELAXERS A LOT AND I CAN'T TAKE THAT MUCH. IT'S NOT HELPING, STILL FEELING THE STIMULATION AND NOW I HAVE A HARD, LIKE KNOT FEELING, ALL AROUND IN MY ABDOMEN, AROUND THE PACEMAKER. IT'S PRETTY UNCOMFORTABLE. I DID PLACE A CALL INTO THE NURSE AT MY SURGEON'S OFFICE WAITING ON A RETURN CALL. I DID HAVE TO CALL AN ON-CALL HOSPITAL DOCTOR THIS WEEKEND AND THEY TOLD ME TO CALL DR. (B)(6)." PATIENT UNDERWENT AN EXPLORATORY SURGERY ON (B)(6) 2025 WITH DR (B)(6). SHE HAD A SMALL SEROMA. THE DEVICE WAS IN THE FLUID. DR (B)(6) TESTED THE SECURITY OF THE LEADS, AND THEY WERE STILL SECURE. HE DISCONNECTED THE LEADS AND SAID THERE APPEARED TO BE FLUID IN THE FLUID IN THE HEADER BY THE SET SCREWS. DR (B)(6) CLEANED OUT THE POCKET AND THEN TOOK GAUZE TO CLEAN OUT THE HEADER PORTS. HE WAS UNABLE TO GUARANTEE THAT ALL FLUIDS HAD BEEN REMOVED, SO HE DECIDED TO REPLACE THE GENERATOR. THE REGISTRATION HAS B
Patient problems: Failure of Implant
MAUDE report 22417387 ↗InjuryJul 7, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT STATES HER ENTERRA THERAPY BATTERY DIED. SHE STATES SYMPTOMS GOT WORSE THE 3RD WEEK OF MARCH WHEN SHE STARTED VOMITING CONSTANTLY. SHE STATES SHE SPENT THE LAST WEEK IN THE HOSPITAL. SHE STATES HER BATTERY IS ONLY LASTING ABOUT 1 YEAR DUE TO HIGH SETTINGS PER PATIENT. SHE STATES SHE IS A PATIENT OF DR. KURIAN WHO IS RETIRING, BUT STATES SHE IS WORKING WITH ANOTHER PROVIDER IN THE OFFICE WHO IS WORKING ON A PREAUTHORIZATION FOR REPLACEMENT SURGERY. PATIENT WAS TRANSFERRED TO TECH SERVICES. PATIENT EXPERIENCED RETURN OF N/V SYMPTOMS DUE TO BATTERY DEPLETION; BATTERY REPLACED; EXPLANTED IPG WAS DISPOSED OF. NO FURTHER ACTION TO BE TAKEN.
Patient problems: Failure of Implant
MAUDE report 22413622 ↗InjuryJul 7, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
FROM THE CALL CENTER: I DON'T KNOW IF MY BATTERY IN MY STIMULATOR IS LOW OR NOT, I NEED TO MAKE ARRANGEMENTS TO SEE IF THE BATTERY SHOULD BE REPLACED. I TAKE NAUSEA MEDICINE, ZOFRAN BUT IT'S NOT WORKING THIS WEEK. ID # (B)(6). PATIENT HAD BATTERY EXCHANGED. EXPLANTED BATTERY WAS DISPOSED OF THEREFORE NO ANALYSIS POSSIBLE. NO FURTHER ACTION TO BE TAKEN.
Patient problems: Abdominal Pain; Failure of Implant
MAUDE report 22413124 ↗InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
FROM THE CALL CENTER: I AM CALLING ABOUT A BATTERY REPLACEMENT FOR A NEUROSTIMULATOR. PATIENT FELT DEVICE WASN'T HELPING AND HAD IT REPLACED; ENDED UP WITH INFECTION OF NEW DEVICE WHICH WAS THEN EXPLANTED TO ALLOW INFECTION TO RESOLVE.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 22372709 ↗DeathJul 1, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT DEATH. PRODUCT RETURNED TO MEDTRONIC FOR ANALYSIS - REPORTED IN MEDTRONIC DATABASE. CHECKED IN AT MDT (B)(6) 2025. MDT RECORD ID: (B)(4). AS REPORTED EVENT DESCRIPTION: PATIENT DEATH. CAUSE OF DEATH: PENDING FURTHER STUDIES. (LEADS RECEIVED CONNECTED TO DEVICE AND SEGMENTED AT PROXIMAL END). REPORTED IN DATABASE ONLY; WAITING FOR DEVICE ANALYSIS FROM MDT. DEVICE EXPLANTED AS A RESULT OF PATIENT DEATH. NO INDICATION THAT PATIENT DEATH WAS DUE TO DEVICE. NO FURTHER ACTION TO BE TAKEN.
Patient problems: Insufficient Information
MAUDE report 22372442 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page