Evera XT VRRecalls & reported problems across all models
Medtronic, Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Evera XT VR model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 867
- Deaths reported
- 29
- Injuries reported
- 641
- Recalls
- 4
We found 6 models of Evera XT VR by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 867 adverse event reports in the FDA's MAUDE database, received between Dec 9, 2015 and Sep 29, 2026 — classified as 29 deaths, 641 injuries, 197 malfunctions.
The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (212 reports).
4 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Evera XT VR recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 10, 2023Open, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1796-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 10, 2023Open, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1795-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 3, 2021Open, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1305-2021 · Root cause: Device Design · names 4 of these models · Recall details · FDA record ↗
- Dec 19, 2017Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0587-2018 · Root cause: Process design · names 2 of these models · Recall details · FDA record ↗
Most reported Evera XT VR problems
Device problems
- Inappropriate/Inadequate Shock/Stimulation212
- Adverse Event Without Identified Device or Use Problem182
- Device Sensing Problem178
- Premature Discharge of Battery164
- Pacing Problem81
- Over-Sensing62
- Electromagnetic Interference49
- Appropriate Device Problem Term/Code Not Available44
Patient problems
- No Clinical Signs, Symptoms or Conditions215
- Shock from Patient Lead(s)211
- No Known Impact Or Consequence To Patient150
- Unspecified Infection124
- No Consequences Or Impact To Patient30
- Tachycardia26
- Bradycardia19
- Sepsis17
Evera XT VR models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Evera™ XT VR DVBB1D4Implantable cardioverter defibrillator (non-CRT)2 recalls20 deaths489 reports
- Evera™ XT VR DVBB1D1Implantable cardioverter defibrillator (non-CRT)2 recalls9 deaths359 reports
- Evera™ XT VR DVBB1D1Implantable cardioverter defibrillator (non-CRT)1 recall18 reports
- Evera™ XT VR DVBB1D4Implantable cardioverter defibrillator (non-CRT)1 recall1 report
- Evera™ XT VR DVBB1D4Implantable cardioverter defibrillator (non-CRT)1 recall0 reports
- Evera™ XT VR DVBB1D1Implantable cardioverter defibrillator (non-CRT)1 recall0 reports
Recent Evera XT VR adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Failure to Deliver Shock/Stimulation; Device Sensing Problem
Patient problems: Syncope/Fainting; Ventricular Fibrillation
MAUDE report 50931672 ↗InjurySep 17, 2026Electrical /Electronic Property Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
It was reported that the Implantable Cardioverter Defibrillator (ICD) possibly delivered inappropriate shocks and Anti-Tachycardia Pacing (ATP) for Atrial Fibrillation (AF) with Rapid Ventricular Response (RVR). The ICD detected the episodes as Fast Ventricular Tachycardia (FVT). The ICD also alerted due to the charge circuit being inactive. The ICD remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50803436 ↗InjurySep 11, 2026Delayed Charge Time
It was reported that the Implantable Cardioverter Defibrillator (ICD) exhibited long charge times and a charge circuit timeout, resulting in a delay of therapy for the patient's Ventricular Fibrillation (VF) episode. The patient remained in Ventricular Fibrillation (VF). The device was explanted and replaced.
Patient problems: Ventricular Fibrillation
MAUDE report 50732678 ↗InjurySep 9, 2026Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation
Patient problems: Shock from Patient Lead(s)
MAUDE report 50697992 ↗MalfunctionJul 8, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25750634 ↗InjuryJun 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO SENSING INTEGRITY COUNTER (SIC) AND PACING IMPEDANCE VARIATION. OVERSENSING OF NOISE WAS OBSERVED. TACHYCARDIA THERAPY WAS PROGRAMMED OFF AND THE RV LEAD REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE PATIENT WAS REFERRED TO A DIFFERENT FACILITY FOR RV LEAD EXTRACTION/IMPLANT. THE PHYSICIAN DECLINED TO PERFORM THE PROCEDURE DUE TO INFECTION ON THE RV LEAD. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 25642144 ↗InjuryJun 9, 2026Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCKS FOR A DETECTED VENTRICULAR FIBRILLATION (VF) THAT WAS SUSPECTED TO BE SUPRA VENTRICULAR TACHYCARDIA (SVT). IN ADDITION, IT WAS NOTED THAT THE ICD DETECTED RUNS OF VENTRICULAR TACHYCARDIA (VT) THAT WAS SUSPECTED TO BE ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE.
Patient problems: Shock from Patient Lead(s)
MAUDE report 25447007 ↗MalfunctionJun 5, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT NOTED THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) IS "NOT FIRING CORRECTLY" AND IS EXPERIENCING A "BURNING" SENSATION AT DEVICE SITE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 25409425 ↗InjuryMay 29, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT HAD A "MAJOR" STROKE. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Stroke/CVA
MAUDE report 25336687 ↗MalfunctionMay 22, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) MISDETECTED AS VENTRICULAR T ACHYCARDIA (VT). THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Shock from Patient Lead(s)
MAUDE report 25257375 ↗InjuryMay 21, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO A POCKET INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Unspecified Infection
MAUDE report 25247272 ↗InjuryMay 19, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED ANTI-TACHYCARDIA PACING (ATP) AND HIGH VOLTAGE THERAPY FROM THE SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FOR POSSIBLE SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE DEVICE CLASSIFIED AS VENTRICULAR TACHYCARDIA (VT). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 25218443 ↗MalfunctionMay 18, 2026Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) FOR DETECTED VENTRICULAR TACHYCARDIA (VT) EPISODES THAT WAS SUSPECTED TO BE SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25207561 ↗InjuryMay 5, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEY WERE HOSPITALIZED FOR A SMALL STROKE FOR TWO DAYS. THE PATIENT NOTED THAT AT NIGHT, THEIR HEART MONITOR WOULD "ALARM" THAT THEIR HEART RATE WAS LOW AND IN THE THIRTIES. THE PATIENT NOTED THAT TWO WEEKS PRIOR TO THEIR HOSPITAL STAY, THEY NOTICED THAT THEIR HEART RATE WAS LOW AS WELL. ADDITIONALLY, THE PATIENT MENTIONED THAT APPROXIMATELY TWO MONTHS PRIOR TO THE CALL, THEY HAD BRADYCARDIA AGAIN AFTER ABOUT TEN YEARS WITHOUT IT. THE PATIENT WAS CONCERNED WITH THEIR HEART RATE DROPPING LOW AT NIGHT AND WHEN THEY REST, STATING "IT COULD BE PROBLEMATIC". THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Transient Ischemic Attack
MAUDE report 25086435 ↗DeathMay 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED A SUSPECTED INFECTION AND RECEIVED EXTENSIVE ANTIBIOTIC TREATMENT. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 25084528 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.