EVIS EXERA DUODENOVIDEOSCOPERecalls & reported problems across all models
Olympus Medical Systems Corp. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one EVIS EXERA DUODENOVIDEOSCOPE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 60
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls
- 1
We found 2 models of EVIS EXERA DUODENOVIDEOSCOPE by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 60 adverse event reports in the FDA's MAUDE database, received between Oct 3, 2019 and Mar 5, 2026 — classified as 7 injuries, 53 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (27 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
EVIS EXERA DUODENOVIDEOSCOPE recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 25, 2023Open, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0346-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported EVIS EXERA DUODENOVIDEOSCOPE problems
Device problems
- Failure to Clean Adequately27
- Microbial Contamination of Device13
- Device Reprocessing Problem6
- Contamination4
- Crack4
- Break3
- Corroded3
- Peeled/Delaminated3
Patient problems
- No Clinical Signs, Symptoms or Conditions49
- No Consequences Or Impact To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Injury2
- Perforation2
- Cholangitis1
- Fever1
- Pain1
EVIS EXERA DUODENOVIDEOSCOPE models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EVIS EXERA DUODENOVIDEOSCOPE OLYMPUS TJF TYPE 145Endoscopic video imaging system/component, gastroenterology-urology1 recall52 reports
- EVIS EXERA DUODENOVIDEOSCOPE OLYMPUS PJF TYPE 160Endoscopic video imaging system/component, gastroenterology-urology8 reports
Recent EVIS EXERA DUODENOVIDEOSCOPE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2026Crack
IT WAS REPORTED THAT THE DUODENOVIDEOSCOPE HAD LIGHT GUIDE AND CHARGED COUPLED DEVICE (CCD) CHIP GLUE IS CRACKED. THE ISSUE WAS IDENTIFIED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24515262 ↗MalfunctionFeb 5, 2026Microbial Contamination of Device
IT WAS REPORTED THAT, DURING REPROCESSING, THE DUODENOVIDEOSCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFUS) OF ENTEROCCUS HIRA. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24274163 ↗MalfunctionDec 3, 2025Corroded; Corroded
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE DISTAL END OF THE DUODENOVIDEOSCOPE WAS CORRODED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704396 ↗MalfunctionJul 11, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE DISTAL END WAS PRESENT WITH FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22484583 ↗MalfunctionJun 25, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE DUODENOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22329273 ↗MalfunctionMar 19, 2025Crack; Crack
IT WAS REPORTED THAT THERE IS DAMAGE TO THE DISTAL PART OF THE DUODENOVIDEOSCOPE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE ARE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21646060 ↗MalfunctionMar 18, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE DISTAL END HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21635595 ↗MalfunctionOct 24, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE DISTAL END OF THE DUODENOVIDEOSCOPE HAD FOREIGN MATERIAL AND RESIDUAL LIQUID. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20523214 ↗MalfunctionOct 23, 2024Positioning Problem; Positioning Problem
IT WAS REPORTED THAT, THE DUODENOVIDEOSCOPE¿S FORCEPS ELEVATOR DOES NOT MOVE DOWN AT ALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20514144 ↗MalfunctionOct 9, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE DUODENOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE LIGHT GUIDE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20408012 ↗MalfunctionSep 16, 2024Material Separation; Defective Component; Material Separation; Defective Component
IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE DUODENOVIDEOSCOPE HAD A GAP ON DISTAL END ADHESION AREA. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20223224 ↗MalfunctionSep 5, 2024Material Separation; Failure to Clean Adequately; Material Separation; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE DUODENOVIDEOSCOPE EXHIBITED DISTAL END HAD FOREIGN MATERIAL AND A GAP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20147955 ↗MalfunctionAug 28, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE DUODENOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL UNDER THE LIGHT GUIDE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20084984 ↗MalfunctionAug 5, 2024Device Reprocessing Problem; Device Reprocessing Problem
DURING THE DEVICE EVALUATION, THE DUODENOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL AT THE DISTAL END ADHESIVE BETWEEN THE SERVER PLUG-IN AND THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19913434 ↗MalfunctionAug 5, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE DUODENOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL FOUND INSIDE THE BIOPSY CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19913420 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.