EVIS EXERA III COLONOVIDEOSCOPERecalls & reported problems across all models
Olympus Medical Systems Corp. · 6 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
5 open recalls name at least one EVIS EXERA III COLONOVIDEOSCOPE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6,117
- Deaths reported
- 0
- Injuries reported
- 62
- Recalls
- 5
We found 6 models of EVIS EXERA III COLONOVIDEOSCOPE by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 6,117 adverse event reports in the FDA's MAUDE database, received between Aug 14, 2018 and Sep 30, 2026 — classified as 62 injuries, 6,055 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (2,565 reports).
5 FDA recall records name at least one of these models by its device identifier, 5 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
EVIS EXERA III COLONOVIDEOSCOPE recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jan 11, 2024Class IIOpen, Classified
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Z-1213-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Jan 11, 2024Class IIOpen, Classified
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Z-1212-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Jan 11, 2024Class IIOpen, Classified
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Z-1210-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Oct 16, 2023Class IIOpen, Classified
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Z-0418-2024 · Root cause: Process control · Recall details · FDA record ↗
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0341-2024 · Root cause: Under Investigation by firm · names 6 of these models · Recall details · FDA record ↗
Most reported EVIS EXERA III COLONOVIDEOSCOPE problems
Device problems
- Failure to Clean Adequately2,565
- Communication or Transmission Problem959
- No Display/Image668
- Residue After Decontamination579
- Poor Quality Image438
- Device Reprocessing Problem406
- Microbial Contamination of Device341
- Contamination171
Patient problems
- No Clinical Signs, Symptoms or Conditions6,025
- Insufficient Information27
- No Consequences Or Impact To Patient22
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available15
- Diarrhea14
- Fever13
- Hemorrhage/Blood Loss/Bleeding7
- No Known Impact Or Consequence To Patient5
EVIS EXERA III COLONOVIDEOSCOPE models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLColonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LEndoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TLColonoscope and accessories, flexible/rigid2 recalls184 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DIColonoscope and accessories, flexible/rigid1 recall164 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TIEndoscope, accessories, narrow band spectrum1 recall77 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190IColonoscope and accessories, flexible/rigid1 recall34 reports
Recent EVIS EXERA III COLONOVIDEOSCOPE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954958 ↗MalfunctionSep 30, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954920 ↗MalfunctionSep 30, 2026Erratic or Intermittent Display; No Display/Image; Poor Quality Image; Communication or Transmission Problem; Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954835 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954708 ↗MalfunctionSep 30, 2026No Display/Image; Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949736 ↗MalfunctionSep 30, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949295 ↗MalfunctionSep 30, 2026Erratic or Intermittent Display
The customer reported that the camera image was intermittently cutting in and out during a diagnostic procedure involving an Olympus Colonovideoscope. The procedure was completed using the same device. There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946817 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
The customer returned the Colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue on the channel and cylinder was observed. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946810 ↗MalfunctionSep 30, 2026No Display/Image; Communication or Transmission Problem
The customer reported that the colonovideoscope displayed a connection error during inspection for use. During the initial evaluation of the device, no image was observed. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946029 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945015 ↗MalfunctionSep 30, 2026Failure to Clean Adequately
During the initial evaluation of the colonovideoscope, it was found that white foreign material was present within the air/water tube and air/water cylinder. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943255 ↗MalfunctionSep 30, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941468 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940880 ↗MalfunctionSep 29, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939151 ↗MalfunctionSep 29, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938094 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
- 4K CAMERA HEAD3 models · 4,926 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page