G7 LongevityRecalls & reported problems across all models
Zimmer, Inc. · 28 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 85
- Deaths reported
- 0
- Injuries reported
- 34
- Recalls
- 0
We found 28 models of G7 Longevity by Zimmer, Inc. in the FDA's device registry (GUDID). Together they are named in 85 adverse event reports in the FDA's MAUDE database, received between Mar 27, 2020 and Sep 8, 2026 — classified as 34 injuries, 51 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported G7 Longevity problems
Device problems
- Difficult to Insert34
- Adverse Event Without Identified Device or Use Problem17
- Device Dislodged or Dislocated10
- Manufacturing, Packaging or Shipping Problem5
- Packaging Problem5
- Detachment of Device or Device Component4
- Positioning Failure3
- Device Contaminated During Manufacture or Shipping2
Patient problems
- No Clinical Signs, Symptoms or Conditions50
- Pain16
- Joint Dislocation12
- Bursitis6
- Ambulation or Postural Difficulties3
- Hematoma3
- Failure of Implant2
- Insufficient Information2
G7 Longevity models
28 models, most reported first. Each model has its own page with its reports, recalls and device details.
- G7® Longevity® 20103606Prosthesis, hip, semi-constrained, metal/polymer, cemented15 reports
- G7® Longevity® 20123605Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented7 reports
- G7® Longevity® 20123607Prosthesis, hip, semi-constrained, metal/polymer, cemented6 reports
- G7® Longevity® 20123203Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented5 reports
- G7® Longevity® 20123606Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented5 reports
- G7® Longevity® 20123204Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented4 reports
- G7® Longevity® 20124006Prosthesis, hip, constrained, cemented or uncemented, metal/polymer4 reports
- G7® Longevity® 20103604Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented3 reports
- G7® Longevity® 20103605Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented3 reports
- G7® Longevity® 20103607PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- G7® Longevity® 20104007Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- G7® Longevity® 20104008Hip, semi-constrained, cemented, metal/polymer + additive, cemented3 reports
- G7® Longevity® 20113606Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- G7® Longevity® 20123205Prosthesis, hip, constrained, cemented or uncemented, metal/polymer3 reports
- G7® Longevity® 20123604PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- G7® Longevity® 20123202Prosthesis, hip, semi-constrained, metal/polymer, cemented2 reports
- G7® Longevity® 20153606Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- G7® Longevity® 20103203Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented1 report
- G7® Longevity® 20103608Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- G7® Longevity® 20113203Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- G7® Longevity® 20113204Hip, semi-constrained, cemented, metal/polymer + additive, cemented1 report
- G7® Longevity® 20113607Hip, semi-constrained, cemented, metal/polymer + additive, cemented1 report
- G7® Longevity® 20114006Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented1 report
- G7® Longevity® 20123608Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- G7® Longevity® 20124007PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- G7® Longevity® 20152801Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented1 report
- G7® Longevity® 20153604PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- G7® Longevity® 20153605Hip, semi-constrained, cemented, metal/ceramic/polymer + additive, porous uncemented1 report
Recent G7 Longevity adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 8, 2026Difficult to Insert; Separation Failure
It was reported that during an initial procedure, the liner could not be implanted in a satisfactory position or extracted, requiring it to be cut into pieces and removed. Another liner was used to complete the surgery, causing an approximately 20-minute delay.Attempts have been made and no further information has been provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50668362 ↗MalfunctionAug 10, 2026Difficult to Insert; Difficult or Delayed Separation
It was reported that during an initial hip procedure, the liner could not be implanted properly after multiple attempts. When implanted, the liner was not aligned with the edge of the cup and it appeared crooked. Repeated attempts to implant the liner failed to resolve the issue, resulting in loosening of the acetabular cup. It was decided to remove the liner. An attempt to remove the liner with instruments was unsuccessful. A threaded wrench was used to remove the liner. Two screws were implanted to stabilize the cup and a new liner was successfully implanted. There was a 20 minute delay.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50191961 ↗InjuryJun 14, 2026Fracture
IT WAS REPORTED THAT APPROXIMATELY FOUR AND A HALF YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO A BROKEN LINER. THE PATIENT FIRST NOTICED AN ISSUE WHILE DOING LEG PRESSES IN THE GYM. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 25497697 ↗MalfunctionApr 27, 2026Difficult to Insert
IT WAS REPORTED THAT A G LINER DID NOT FULLY SEAT AND LOCK INTO A CUP. THERE WAS NO PATIENT IMPACTS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25001168 ↗MalfunctionMar 11, 2026Difficult to Insert
IT WAS REPORTED THAT THE LINER COULD NOT BE ASSEMBLED WITH THE MATING CUP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24568381 ↗MalfunctionFeb 18, 2026Difficult to Insert
IT WAS REPORTED THAT DURING INSERTION OF A HIGH WALL LINER INTO AN ACETABULAR CUP, THE LINER WOULD NOT SEAT WITH THE CUP. THE LOCKING MECHANISM AND SOME TABS WERE DAMAGED. THE REPORTER NOTED THAT NO UNUSUAL FORCE WAS APPLIED AND THERE WAS NO OBSTRUCTION WITHIN THE CUP. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24376978 ↗InjuryFeb 16, 2026Difficult to Insert
NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.
Patient problems: Joint Dislocation
MAUDE report 24345329 ↗InjuryJan 27, 2026Unintended Movement; Noise, Audible
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SNAPPING IN THE HIP APPROXIMATELY TWO YEARS POST-IMPLANTATION. IT WAS CONFIRMED THAT THE PATIENT HAD ILIOPSOAS TENDINITIS CAUSED BY IMPINGEMENT. PATIENT WAS REFERRED FOR AN INJECTION UNDER FLUOROSCOPY. NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.
Patient problems: Joint Impingement; Pain; Tendinopathy
MAUDE report 24188713 ↗MalfunctionDec 29, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING A RIGHT HIP PROCEDURE, THE LINER WOULD NOT MATE WITH THE CUP. IT WAS ENSURED THE CUP WAS FREE OF DEBRIS BEFORE ATTEMPTING TO IMPLANT A SECOND LINER. THE SECOND LINER NEVER SAT FLUSH WITH IMPACTION. A DIFFERENT SIZE LINER WAS USED TO COMPLETE THE PROCEDURE AND IT FIT WITH NO ISSUES. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23918602 ↗MalfunctionDec 29, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING A RIGHT HIP PROCEDURE, THE LINER WOULD NOT MATE WITH THE CUP. IT WAS ENSURED THE CUP WAS FREE OF DEBRIS BEFORE ATTEMPTING TO IMPLANT A SECOND LINER. THE SECOND LINER NEVER SAT FLUSH WITH IMPACTION. A DIFFERENT SIZE LINER WAS USED TO COMPLETE THE PROCEDURE AND IT FIT WITH NO ISSUES. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23918601 ↗MalfunctionDec 17, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SURGEON WAS UNABLE TO SEAT THE G7 LINER INTO THE G7 SHELL. A SECOND G7 LINER WAS OPENED AND ATTEMPTED WITH THE SAME RESULT. A THIRD G7 LINER WAS THEN ASSEMBLED, AND THE CASE WAS COMPLETED SUCCESSFULLY. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS WERE MADE; HOWEVER, NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23831453 ↗MalfunctionDec 17, 2025Difficult to Insert; Difficult to Insert
IT WAS CONFIRMED THAT NO FURTHER INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23831438 ↗MalfunctionDec 13, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SURGEON COULD NOT IMPACT THE LINER. A SECOND LINER WAS USED AND ALSO COULD NOT BE IMPLANTED. A THIRD LINER WAS ABLE TO BE IMPLANTED INTO THE SAME CUP. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23797878 ↗MalfunctionOct 22, 2025Difficult to Insert; Difficult to Insert
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23363671 ↗InjuryOct 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
Patient problems: Hematoma
MAUDE report 23323630 ↗
Other Zimmer, Inc. products
- PERSONA364 models · 2,000 reports · 3 recalls
- Persona The Personalized Knee77 models · 1,396 reports
- PERSONA Vivacit-E285 models · 1,152 reports
- NexGen198 models · 842 reports · 4 recalls
- Affixus18 models · 1 reports · 17 recalls
- PERSONA Natural Tibia25 models · 417 reports · 2 recalls
- NexGen Legacy119 models · 319 reports · 3 recalls
- Alliance32 models · 275 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.