GETINGE 88-SERIESRecalls & reported problems across all models
Getinge Disinfection AB · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one GETINGE 88-SERIES model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 3
We found 2 models of GETINGE 88-SERIES by Getinge Disinfection AB in the FDA's device registry (GUDID). Together they are named in 6 adverse event reports in the FDA's MAUDE database, received between Sep 19, 2017 and Sep 23, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fire” (3 reports).
3 FDA recall records name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
GETINGE 88-SERIES recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 29, 2025Open, Classified
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Z-0016-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
- Feb 10, 2020Terminated
Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water
Z-1475-2020 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
- Mar 2, 2018Terminated
There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
Z-1380-2018 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗
Most reported GETINGE 88-SERIES problems
Device problems
- Fire3
- Overheating of Device2
- Device Handling Problem1
- Improper or Incorrect Procedure or Method1
- Inadequate User Interface1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Hematoma1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
GETINGE 88-SERIES models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- GETINGE 88-SERIES 88-5Disinfector, Medical Devices3 recalls6 reports
- GETINGE 88-SERIES 88-5Disinfector, Medical Devices3 recalls0 reports
Recent GETINGE 88-SERIES adverse event reports
The 6 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2024Fire; Fire
ON SEPTEMBER 19TH, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH THE 88-SERIES WASHER DISINFECTOR WITH THE MODEL NAME: 88-5. AS IT WAS STATED, THE WASHER DISINFECTOR CAUGHT ON FIRE AFTER ALL EMPLOYEES HAD GONE HOME. SO FAR, WE HAVE NOT BEEN INFORMED ABOUT ANY SERIOUS INJURY AS A CONSEQUENCE OF REPORTED ISSUE, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20276739 ↗MalfunctionOct 31, 2023Fire; Fire
ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH THE 88-SERIES WASHER DISINFECTOR WITH THE MODEL NAME:88-5. AS IT WAS STATED, THE WASHER CAUGHT FIRE DURING CYCLE. IT WAS CONFIRMED THAT A FIRE ALARM STARTED AND THE FLOOR LEVEL WAS EVACUATED. SO FAR, WE HAVE NOT BEEN INFORMED ABOUT ANY SERIOUS INJURY AS A CONSEQUENCE OF REPORTED ISSUE, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION WAS TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18037297 ↗MalfunctionMar 13, 2020Inadequate User Interface; Device Handling Problem; Inadequate User Interface; Device Handling Problem
MalfunctionOct 16, 2019Fire; Overheating of Device; Fire; Overheating of Device
MANUFACTURER REFERENCE NUMBER (B)(4).
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9198503 ↗MalfunctionAug 27, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ON 2ND OF AUGUST 2018 GETINGE BECAME AWARE OF AN SITUATION WHERE AN ARM OF OPERATOR WAS CAUGHT BETWEEN WASHER DISINFECTOR AND THE CART TRANSFERRED INTO THE WASHER BY THE FREE STANDING LOADER. AFTER THE INCIDENT OCCURRENCE, THE PERSON WENT TO THE EMERGENCY DEPARTMENT WHERE, AFTER THE ASSESSMENT, IT HAS BEEN CONFIRMED THAT THE ARM WAS NOT BROKEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7817431 ↗MalfunctionSep 19, 2017Overheating of Device; Overheating of Device
MANUFACTURER'S REFERENCE NUMBER TW: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 6875167 ↗
Other Getinge Disinfection AB products
All Getinge Disinfection AB devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page