GETINGE 88-SERIESRecalls & reported problems across all models

Getinge Disinfection AB · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one GETINGE 88-SERIES model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls
3

We found 2 models of GETINGE 88-SERIES by Getinge Disinfection AB in the FDA's device registry (GUDID). Together they are named in 6 adverse event reports in the FDA's MAUDE database, received between Sep 19, 2017 and Sep 23, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fire” (3 reports).

3 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

GETINGE 88-SERIES recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 29, 2025Open, Classified

    Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

    Z-0016-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  2. Feb 10, 2020Terminated

    Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water

    Z-1475-2020 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  3. Mar 2, 2018Terminated

    There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

    Z-1380-2018 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported GETINGE 88-SERIES problems

Device problems

  • Fire3
  • Overheating of Device2
  • Device Handling Problem1
  • Improper or Incorrect Procedure or Method1
  • Inadequate User Interface1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • Hematoma1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

GETINGE 88-SERIES models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent GETINGE 88-SERIES adverse event reports

The 6 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Getinge Disinfection AB products

All Getinge Disinfection AB devices →