GETINGE 88-SERIES 88-5 (REF 88-303)Recalls & reported problems

Getinge Disinfection AB · GETINGE WD88, WASHER DISINFECTOR · Class 2

All models of GETINGE 88-SERIES (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07340153700109), received between Sep 19, 2017 and Sep 23, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fire” (3 reports).

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 29, 2025Open, Classified

    Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

    Z-0016-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 10, 2020Terminated

    Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water

    Z-1475-2020 · Root cause: Device Design · Recall details · FDA record ↗

  3. Mar 2, 2018Terminated

    There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

    Z-1380-2018 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fire3
  • Overheating of Device2
  • Device Handling Problem1
  • Improper or Incorrect Procedure or Method1
  • Inadequate User Interface1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • Hematoma1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2024Fire; Fire

    ON SEPTEMBER 19TH, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH THE 88-SERIES WASHER DISINFECTOR WITH THE MODEL NAME: 88-5. AS IT WAS STATED, THE WASHER DISINFECTOR CAUGHT ON FIRE AFTER ALL EMPLOYEES HAD GONE HOME. SO FAR, WE HAVE NOT BEEN INFORMED ABOUT ANY SERIOUS INJURY AS A CONSEQUENCE OF REPORTED ISSUE, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20276739 ↗
  • MalfunctionOct 31, 2023Fire; Fire

    ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH THE 88-SERIES WASHER DISINFECTOR WITH THE MODEL NAME:88-5. AS IT WAS STATED, THE WASHER CAUGHT FIRE DURING CYCLE. IT WAS CONFIRMED THAT A FIRE ALARM STARTED AND THE FLOOR LEVEL WAS EVACUATED. SO FAR, WE HAVE NOT BEEN INFORMED ABOUT ANY SERIOUS INJURY AS A CONSEQUENCE OF REPORTED ISSUE, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AND BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATION WAS TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18037297 ↗
  • MalfunctionMar 13, 2020Inadequate User Interface; Device Handling Problem; Inadequate User Interface; Device Handling Problem

    MANUFACTURER REFERENCE NUMBER (B)(4).

    Patient problems: Hematoma

    MAUDE report 9829114 ↗
  • MalfunctionOct 16, 2019Fire; Overheating of Device; Fire; Overheating of Device

    MANUFACTURER REFERENCE NUMBER (B)(4).

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9198503 ↗
  • MalfunctionAug 27, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON 2ND OF AUGUST 2018 GETINGE BECAME AWARE OF AN SITUATION WHERE AN ARM OF OPERATOR WAS CAUGHT BETWEEN WASHER DISINFECTOR AND THE CART TRANSFERRED INTO THE WASHER BY THE FREE STANDING LOADER. AFTER THE INCIDENT OCCURRENCE, THE PERSON WENT TO THE EMERGENCY DEPARTMENT WHERE, AFTER THE ASSESSMENT, IT HAS BEEN CONFIRMED THAT THE ARM WAS NOT BROKEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7817431 ↗
  • MalfunctionSep 19, 2017Overheating of Device; Overheating of Device

    MANUFACTURER'S REFERENCE NUMBER TW: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 6875167 ↗

Other Disinfector, Medical Devices devices

All Disinfector, Medical Devices devices →