IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab)Recalls & reported problems across all models
Siemens Healthcare Diagnostics Products Ltd · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name at least one IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 2
We found 2 models of IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) by Siemens Healthcare Diagnostics Products Ltd in the FDA's device registry (GUDID). Together they are named in 17 adverse event reports in the FDA's MAUDE database, received between Jan 29, 2018 and Jan 29, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (12 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 23, 2024Class IIOpen, Classified
The potential for falsely elevated patient sample results when using impacted lots (785, 790, 791, and 793) of IMMULITE 2000 Anti-TG Ab.
Z-1887-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 23, 2024Class IIOpen, Classified
The potential for falsely elevated patient sample results when using impacted lots (782, 783, 785, 786, 787, 788, 789, 790, 791, 792 and 793), of IMMULITE 2000 Anti-TG Ab.
Z-1886-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) problems
Device problems
- Low Test Results12
- Adverse Event Without Identified Device or Use Problem5
Patient problems
- No Known Impact Or Consequence To Patient16
- No Clinical Signs, Symptoms or Conditions1
IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- IMMULITE® 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) 10381659SYSTEM, TEST, THYROID AUTOANTIBODY1 recall17 reports
- IMMULITE® 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) 10381655SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
Recent IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2026Low Test Results
THE CUSTOMER CONTACTED SIEMENS AND INDICATED THAT A FALSE LOW ANTI-THYROGLOBULIN (ATG) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN IMMULITE 2000 XPI SYSTEM USING IMMULITE 2000 ATG REAGENT. THE FALSE LOW RESULT WAS REPORTED AND QUESTIONED. THE FOLLOWING DAY, THE SAMPLE WAS RUN FOR ATG IN ANOTHER LABORATORY USING A NON-SIEMENS METHOD AND THE RESULT RECOVERED HIGHER. THE HIGHER RESULT WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO LOW ATG RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24206836 ↗MalfunctionFeb 4, 2020Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI ON A SINGLE PATIENT. THREE INITIAL TESTING USING SIEMENS ADVIA CENTAUR XP WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). A LATER TEST WITH IMMULITE 2000 XPI WAS CONSIDERED DISCORDANT BUT WAS REPORTED TO THE PHYSICIAN(S) THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9667670 ↗MalfunctionFeb 4, 2020Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI ON A SINGLE PATIENT. THREE INITIAL TESTING USING SIEMENS ADVIA CENTAUR XP WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). A LATER TEST WITH IMMULITE 2000 XPI WAS CONSIDERED DISCORDANT BUT WAS REPORTED TO THE PHYSICIAN(S) THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9667669 ↗MalfunctionFeb 4, 2020Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI ON A SINGLE PATIENT. THREE INITIAL TESTING USING SIEMENS ADVIA CENTAUR XP WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). TWO ADDITIONAL TESTING USING AN ALTERNATE NON-SIEMENS METHOD WERE ALSO CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). A LATER TEST USING IMMULITE 2000 XPI WAS CONSIDERED DISCORDANT BUT WAS REPORTED TO THE PHYSICIAN(S). ANOTHER TEST USING SIEMENS ATELLICA IM WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9667659 ↗MalfunctionFeb 4, 2020Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI ON A SINGLE PATIENT. THREE INITIAL TESTING USING SIEMENS ADVIA CENTAUR XP WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). A LATER TEST WITH IMMULITE 2000 XPI WAS CONSIDERED DISCORDANT BUT WAS REPORTED TO THE PHYSICIAN(S) THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9667648 ↗MalfunctionFeb 4, 2020Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI ON A SINGLE PATIENT. THREE INITIAL TESTING USING SIEMENS ADVIA CENTAUR XP WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). A LATER TEST WITH IMMULITE 2000 XPI WAS CONSIDERED DISCORDANT BUT WAS REPORTED TO THE PHYSICIAN(S) THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9667484 ↗MalfunctionOct 30, 2019Low Test Results; Low Test Results
THREE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULTS WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE REPEAT RESULTS USING SIEMENS ATELLICA IM WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9257612 ↗MalfunctionSep 27, 2019Low Test Results; Low Test Results
THREE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULTS WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE REPEAT RESULTS USING SIEMENS ATELLICA IM WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131712 ↗MalfunctionSep 27, 2019Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE REPEAT RESULT USING SIEMENS ATELLICA IM WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131711 ↗MalfunctionSep 27, 2019Low Test Results; Low Test Results
TWO DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS USING SIEMENS ATELLICA IM WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). THE REPEAT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S) AND WERE CONSIDERED DISCORDANT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131708 ↗MalfunctionSep 27, 2019Low Test Results; Low Test Results
TWO DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULT WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS USING SIEMENS ATELLICA IM WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). THE REPEAT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S) AND WERE CONSIDERED DISCORDANT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131704 ↗MalfunctionJul 5, 2019Low Test Results; Low Test Results
FOUR DISCORDANT, FALSELY DEPRESSED ANTI-THYROGLOBULIN ANTIBODY (ATG) RESULTS ON A SINGLE PATIENT WERE OBTAINED USING IMMULITE 2000 XPI ON THE SAME DAY. THE INITIAL RESULT WAS REPORTED AND QUESTIONED BY THE PHYSICIAN(S). THREE REPEAT RESULTS USING THE SAME IMMULITE 2000 XPI WERE CONSIDERED DISCORDANT. THREE REPEAT RESULTS USING AN ALTERNATE NON-SIEMENS METHOD WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). TWO ADDITIONAL REPEAT RESULTS USING SIEMENS ATELLICA SOLUTIONS WERE ALSO CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ATG RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8765362 ↗MalfunctionMar 26, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DISCORDANT, FALSELY LOW AUTOANTIBODIES TO THYROGLOBULIN (ANTI-TG AB) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON AN IMMULITE 2000 XPI INSTRUMENT, WHILE USING KIT LOT 673. THE RESULTS FROM IMMULITE 2000 XPI INSTRUMENT WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ADVIA CENTAUR XPT INSTRUMENT, RESULTING HIGHER. THE RESULTS FROM THE ADVIA CENTAUR XPT INSTRUMENT WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW ANTI-TG AB RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371322 ↗MalfunctionMar 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A DISCORDANT, FALSELY LOW RESULT WAS OBTAINED FOR AUTOANTIBODIES TO THYROGLOBULIN (ANTI-TG AB) ASSAY ON ONE PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT, WHILE USING KIT LOT 675. THE RESULT FROM THE IMMULITE 2000 XPI INSTRUMENT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED IT. THE SAMPLE WAS REPEATED ON AN ALTERNATE PLATFORM, RESULTING HIGHER. IT IS UNKNOWN IF THE RESULT FROM THE ALTERNATE PLATFORM WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCE DUE TO THE DISCORDANT, FALSELY LOW ANTI-TG AB RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7310279 ↗MalfunctionMar 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A DISCORDANT, FALSELY LOW RESULT WAS OBTAINED FOR AUTOANTIBODIES TO THYROGLOBULIN (ANTI-TG AB) ASSAY ON ONE PATIENT SAMPLE ON AN IMMULITE 2000 XPI INSTRUMENT, WHILE USING KIT LOT 675. THE RESULT FROM THE IMMULITE 2000 XPI INSTRUMENT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED IT. THE SAMPLE WAS REPEATED ON AN ALTERNATE PLATFORM, RESULTING HIGHER. IT IS UNKNOWN IF THE RESULT FROM THE ALTERNATE PLATFORM WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCE DUE TO THE DISCORDANT, FALSELY LOW ANTI-TG AB RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7310277 ↗
Other Siemens Healthcare Diagnostics Products Ltd products
- IMMULITE 2000 Systems HCG2 models · 189 reports · 1 recall
- IMMULITE 2000 Systems E2 Estradiol2 models · 40 reports · 1 recall
- IMMULITE 2000 Systems IPT Intact PTH2 models · 18 reports · 1 recall
- IMMULITE 2000 Systems PRG Progesterone2 models · 7 reports · 1 recall
- IMMULITE 2000 Systems PTS PSA2 models · 1 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page