Skip to content

IMMULITE 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) recallZ-1886-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Apr 23, 2024

Class II Open, Classified All recalls posted in May 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The potential for falsely elevated patient sample results when using impacted lots (782, 783, 785, 786, 787, 788, 789, 790, 791, 792 and 793), of IMMULITE 2000 Anti-TG Ab.

FDA root cause category: Under Investigation by firm

What the company is doing

On April23, 2024, Urgent Medical Device Correction letters were sent to customers. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Perform the following: Discontinue use of and discard the kit lots listed in the Products Table. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product.

Recall details

Recall number
Z-1886-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 23, 2024
Posted by the FDA
May 23, 2024
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
2,956 units (463 US, 2,493 OUS)
Product code
JZO
FDA submissions
K991094
Distribution
Worldwide - US Nationwide distribution including in the states of AL, CA, DC, FL, HI, ID, IN, KS, KY MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, PR, TX, WA WI and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Finland, France, Germany, Greece Guatemala, Honduras, Hong Kong, Hungary, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Norway, P.R. China, Pakistan, Paraguay, Peru, Poland, Portugal, Republic South Korea, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, Ukraine, United Kingdom, Uruguay, Vietnam.

Product description

IMMULITE 2000 Anti-TG Ab Catalog # L2KTG2 (200 Tests), SMN 10381659, System, Test, Thyroid Autoantibody.

Code information (models, lots, serial numbers)

Catalog # L2KTG2, SMN 10381659, UDI/DI: 00630414962108, Lot #s: 782, 783, 786, 787, 788, 789, 791, 792, 793.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K991094 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94495).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page