We found 15 models of LCB by Orthopediatrics Corp. in the FDA's device registry (GUDID). Together they are named in 20 adverse event reports in the FDA's MAUDE database, received between Mar 23, 2018 and Sep 30, 2026 — classified as 9 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Fracture” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported LCB problems
Device problems
- Fracture19
- Bent1
Patient problems
- Failure of Implant11
- No Information4
- Pain2
- Discomfort1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
LCB models
15 models, most reported first. Each model has its own page with its reports, recalls and device details.
- LCB 00-1200-1002PLATE, FIXATION, BONE6 reports
- LCB 00-1200-2001PLATE, FIXATION, BONE1 report
- LCB 00-1200-2002PLATE, FIXATION, BONE1 report
- LCB 00-1200-4503PLATE, FIXATION, BONE1 report
- LCB 00-1200-4518PLATE, FIXATION, BONE1 report
- LCB 00-1200-4519PLATE, FIXATION, BONE1 report
- LCB 00-1200-5000PLATE, FIXATION, BONE1 report
- LCB 00-1200-5002PLATE, FIXATION, BONE1 report
- LCB 00-1200-5003PLATE, FIXATION, BONE1 report
- LCB 00-1200-7001PLATE, FIXATION, BONE1 report
- LCB 00-1200-7002PLATE, FIXATION, BONE1 report
- LCB 00-1200-7003PLATE, FIXATION, BONE1 report
- LCB 00-1200-7005PLATE, FIXATION, BONE1 report
- LCB 00-1200-2503Plate, fixation, bone1 report
- LCB 00-1200-8048Plate, fixation, bone1 report
Recent LCB adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Fracture
Patient problems: Failure of Implant
MAUDE report 50952150 ↗InjurySep 15, 2026Fracture
The patient underwent surgery on (b)(6) 2026, during which bilateral 3.5 mm Blade Plates were used. Subsequently, one of the plates was reported to have fractured and was replaced with an LPF.
Patient problems: Failure of Implant
MAUDE report 50778280 ↗MalfunctionFeb 9, 2026Fracture
PATIENT CAME TO THE HOSPITAL WITH SEVERE PAIN, DOCTOR SENT FOR XRAY AND AT THIS POINT DISCOVERED THAT THE PLATE WAS BROKEN INSIDE.
Patient problems: Pain
MAUDE report 24292456 ↗MalfunctionFeb 9, 2026Fracture
THE PATIENT HAS BEEN EXPERIENCING SEVERAL CRACKING SOUNDS FOR SEVERAL WEEKS. THERE IS NO DATE FOR THE NEW SURGERY AT THE MOMENT.
Patient problems: Failure of Implant
MAUDE report 24292454 ↗MalfunctionOct 24, 2025Fracture; Fracture
DR. (B)(6) FROM (B)(6) HOSPITAL REPORTED A POST-OPERATIVE ISSUE INVOLVING A TRAUMA & DEFORMITY SYSTEM, SPECIFICALLY AN LCB BLADE PLATE (PART NUMBER: 00-1200-1002, LOT NUMBER: 241069-B), WHICH HAD BEEN IN USE FOR LESS THAN SIX MONTHS. THE DEVICE WAS IMPLANTED IN AN 8-YEAR-OLD FEMALE PATIENT (INITIALS (B)(6)) ON (B)(6) 2025 AND WAS REMOVED ON (B)(6) 2025, AFTER THE PLATE BROKE. THE ISSUE WAS DISCOVERED DURING A CLINIC VISIT POST-OPERATION. THE BROKEN DEVICE WAS SCRAPPED AND NOT RETURNED FOR EVALUATION.
Patient problems: Failure of Implant
MAUDE report 23380666 ↗MalfunctionJul 30, 2025Fracture; Fracture
DR. (B)(6) FROM THE(B)(6) REPORTED A COMPLAINT INVOLVING A CONSIGNED, NON-STERILE "INFANT 2-HOLE" PLATE (PART NUMBER 00-1200-2001) FROM THE TRAUMA & DEFORMITY - LCB SYSTEM. THE ISSUE WAS DISCOVERED POST-OPERATIVELY VIA X-RAY, REVEALING A FRACTURE OF THE PLATE OVER THE CUT SITE. THE PATIENT, A MALE, HAD BEEN LYING ON HIS SIDE WITH HIS LEG ELEVATED AT THE TIME. ALTHOUGH THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT DELAY OR INJURY, THE DEVICE FAILURE NECESSITATED A REVISION PROCEDURE. THE IMPLANT WAS PLACED ON (B)(6) 2025 AND REMOVED ON (B)(6) 2025. FRACTURED COMPONENTS ARE AVAILABLE AND WILL BE PROVIDED TO ENGINEERING FOR FURTHER ANALYSIS. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: Failure of Implant
MAUDE report 22664173 ↗MalfunctionJul 9, 2025Fracture; Fracture
130 DEGREE BLADE PLATE BROKE UNDER THE BLADE.
Patient problems: Failure of Implant
MAUDE report 22465593 ↗MalfunctionMay 27, 2025Fracture; Fracture
AFTER POST OP IMAGING AND PATIENT COMPLAINT ON PAIN THE SURGEON DISCOVER PLATE BREAKAGE.
Patient problems: Failure of Implant
MAUDE report 22088060 ↗MalfunctionMay 13, 2025Fracture; Fracture
MalfunctionDec 26, 2024Fracture; Fracture
IMPLANT FAILED WHILE IN SITU AFTER APPROX. 6 MONTHS POST IMPLANTATION. IMPLANT FRACTURED
Patient problems: Insufficient Information
MAUDE report 21023912 ↗MalfunctionOct 27, 2024Fracture; Fracture
BLADE PLATE BROKEN SEEN BY XRAY AND DISCOMFORT BY PATIENT.
Patient problems: Failure of Implant
MAUDE report 20541654 ↗MalfunctionOct 27, 2024Fracture; Fracture
INFANT BLADE PLATE BROKEN AT OSTEOTOMY LEVEL.
Patient problems: Failure of Implant
MAUDE report 20541652 ↗MalfunctionOct 10, 2024Fracture; Fracture
LOCKING CANNULATED BLADE PLATE 50MM 3.5 130 DEGREE BROKE AT PROXIMAL COMPRESSION SCREW HOLE. THIS WAS ALSO THE LOCATION OF THE DEFORMITY CORRECTION OSTEOTOMY SITE. 10 Y/O FEMALE W CP AND MOVING DISORDER. BILATERAL LCB CASE ORIGINALLY COMPLETED (B)(6) 2024.
Patient problems: Failure of Implant
MAUDE report 20424895 ↗InjuryApr 23, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT FOLLOWING THE PLACEMENT OF A LOCKING CANNULATED BLADE PLATE, THE PATIENT PRESENTED WITH PAIN. THE PLATE WAS FOUND TO BE FRACTURED AND THE PATIENT UNDERWENT A REVISION PROCEDURE THE FOLLOWING DAY. NO ADDITIONAL PATIENT CONSEQUENCES HAVE BEEN REPORTED.
Patient problems: Pain
MAUDE report 9993786 ↗InjuryFeb 24, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT APPROXIMATELY FOUR MONTHS FOLLOWING THE PLACEMENT OF A LOCKING CANNULATED BLADE PLATE, THE PLATE WAS DISCOVERED TO BE FRACTURED. THE PATIENT WAS REVISED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Information
MAUDE report 9743842 ↗
Other Orthopediatrics Corp. products
All Orthopediatrics Corp. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page