LCB 00-1200-2001Recalls & reported problems
Orthopediatrics Corp. · 90 Deg Cannulated Infant Blade Plate 30mm x 5mm x 2 Hole · Class 2 · Implant
All models of LCB (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132109731), received between Jul 30, 2025 and Jul 30, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K110959, K260323); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
Patient problems
- Failure of Implant1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2025Fracture; Fracture
DR. (B)(6) FROM THE(B)(6) REPORTED A COMPLAINT INVOLVING A CONSIGNED, NON-STERILE "INFANT 2-HOLE" PLATE (PART NUMBER 00-1200-2001) FROM THE TRAUMA & DEFORMITY - LCB SYSTEM. THE ISSUE WAS DISCOVERED POST-OPERATIVELY VIA X-RAY, REVEALING A FRACTURE OF THE PLATE OVER THE CUT SITE. THE PATIENT, A MALE, HAD BEEN LYING ON HIS SIDE WITH HIS LEG ELEVATED AT THE TIME. ALTHOUGH THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT DELAY OR INJURY, THE DEVICE FAILURE NECESSITATED A REVISION PROCEDURE. THE IMPLANT WAS PLACED ON (B)(6) 2025 AND REMOVED ON (B)(6) 2025. FRACTURED COMPONENTS ARE AVAILABLE AND WILL BE PROVIDED TO ENGINEERING FOR FURTHER ANALYSIS. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: Failure of Implant
MAUDE report 22664173 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110959, K260323. One submission often covers many models, so check the product and lot details in each record.
- Z-1535-2013 · Apr 19, 2013 · Terminated90 Degree Cannulated Infant Blade Plate 30mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractur
- Z-1534-2013 · Apr 19, 2013 · Terminated90 Degree Cannulated Infant Blade Plate 25mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractur
- Z-1536-2013 · Apr 19, 2013 · Terminated90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractur
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.