LifeGlobal MediaRecalls & reported problems across all models
Lifeglobal Group, LLC · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one LifeGlobal Media model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 3
We found 5 models of LifeGlobal Media by Lifeglobal Group, LLC in the FDA's device registry (GUDID). None of them is named in an adverse event report in our data.
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
LifeGlobal Media recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 20, 2023Open, Classified
The firm has become aware of a sudden increase in complaints for three associated lots. Performance issues may lead to impaired embryo development prior to the blastocyst stage.
Z-1089-2024 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗
- Feb 13, 2023Open, Classified
It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
Z-1348-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
- Jun 13, 2022Open, Classified
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
Z-1421-2022 · Root cause: Error in labeling · Recall details · FDA record ↗
LifeGlobal Media models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- LifeGlobal® Media LGGG-020MEDIA, REPRODUCTIVE1 recall0 reports
- LifeGlobal® Media LGGG-050MEDIA, REPRODUCTIVE1 recall0 reports
- LifeGlobal® Media LGGG-100MEDIA, REPRODUCTIVE1 recall0 reports
- LifeGlobal® Media H5GT-060MEDIA, REPRODUCTIVE1 recall0 reports
- LifeGlobal® Media GFT5-055MEDIA, REPRODUCTIVE1 recall0 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page