Skip to content

LifeGlobal Media recallZ-1421-2022 · CooperSurgical, Inc. · initiated Jun 13, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.

FDA root cause category: Error in labeling

What the company is doing

On June 13, 2022, the firm distributed Urgent Media Safety Notice letters to affected customers. Customers were informed that setting out a thawing blastocyte for 5 minutes may be detrimental to the survival of the blastocyte. Customers were asked to complete and return an acknowledgement form. Once the firm has received the completed form, they will provide the corrected IFU.

Recall details

Recall number
Z-1421-2022
Status
Open, Classified
Initiated by the company
Jun 13, 2022
Recalling firm
CooperSurgical, Inc.
Quantity
403 units
Product code
MQL
FDA submissions
K092963
Distribution
Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI, International: Belgium, Canada, Croatia, Netherlands, Portugal, Spain

Product description

LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

Code information (models, lots, serial numbers)

UDI: (01)00815965020914 (17)220819 (10)GFT5-200218U (01)00815965020914 (17)220819 (10)GFT5-201026U (01)00815965020914 (17)220819 (10)GFT5-21335060  Lots: 21335060, GFT5-200218U, GFT5-201026U

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page