Skip to content

LINQ IIRecalls & reported problems across all models

Medtronic, Inc. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one LINQ II model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
7,610
Deaths reported
0
Injuries reported
341
Recalls
2

We found 3 models of LINQ II by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 7,610 adverse event reports in the FDA's MAUDE database, received between Aug 27, 2020 and Sep 30, 2026 — classified as 341 injuries, 7,269 malfunctions.

The most frequently reported specific device problem is “Reset Problem” (1,846 reports).

2 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 124 reportsDec 2023: 81 reportsJan 2024: 142 reportsFeb 2024: 127 reportsMar 2024: 136 reportsApr 2024: 125 reportsMay 2024: 303 reportsJun 2024: 111 reportsJul 2024: 151 reportsAug 2024: 149 reportsSep 2024: 160 reportsOct 2024: 235 reportsNov 2024: 158 reportsDec 2024: 41 reportsJan 2025: 184 reportsFeb 2025: 157 reportsMar 2025: 140 reportsApr 2025: 95 reportsMay 2025: 188 reportsJun 2025: 190 reportsJul 2025: 194 reportsAug 2025: 347 reportsSep 2025: 100 reportsOct 2025: 173 reportsNov 2025: 173 reportsDec 2025: 134 reportsJan 2026: 153 reportsFeb 2026: 131 reportsMar 2026: 143 reportsApr 2026: 268 reportsMay 2026: 110 reportsJun 2026: 124 reportsJul 2026: 104 reportsAug 2026: 107 reportsSep 2026: 66 reportsOct 2026: 0 reports202420252026347
Last 36 months. The most recent month may be incomplete.

LINQ II recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Nov 3, 2023Open, Classified

    It may create the potential for amplified noise and/or overall signal reduction of the ICM, which may interfere with intended recordings of heart rhythms. This noise pattern is different from occasional noise due to device position/migration, patient activity, or external electromagnetic interference.

    Z-0421-2024 · Root cause: Process change control · names 3 of these models · Recall details · FDA record ↗

  2. Jan 18, 2022Terminated

    Devices may lose functionality due to susceptibility to moisture ingress.

    Z-0782-2022 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗

Most reported LINQ II problems

Device problems

  • Reset Problem1,846
  • Premature Discharge of Battery1,140
  • Inappropriate or Unexpected Reset1,069
  • Output Problem948
  • Signal Artifact/Noise695
  • Under-Sensing624
  • Over-Sensing449
  • Data Problem407

Patient problems

  • No Clinical Signs, Symptoms or Conditions7,225
  • Unspecified Infection150
  • Pain93
  • Erosion90
  • Discomfort68
  • Erythema52
  • Purulent Discharge30
  • Swelling/ Edema22

LINQ II models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent LINQ II adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic, Inc. products

All Medtronic, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page