LyraRecalls & reported problems across all models
Diagnostic Hybrids, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 489
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 0
We found 2 models of Lyra by Diagnostic Hybrids, Inc. in the FDA's device registry (GUDID). Together they are named in 489 adverse event reports in the FDA's MAUDE database, received between Apr 20, 2021 and Mar 20, 2025 — classified as 489 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (388 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Lyra problems
Device problems
- No Apparent Adverse Event489
- False Positive Result388
- False Negative Result33
- Incorrect, Inadequate or Imprecise Result or Readings26
- Imprecise or Erratic Results20
Patient problems
- No Clinical Signs, Symptoms or Conditions489
- Insufficient Information1
Lyra models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Lyra CoronavirusReagents, 2019-novel coronavirus nucleic acid403 reports
- Lyra CoronavirusReagents, 2019-novel coronavirus nucleic acid86 reports
Recent Lyra adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2025False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
REPORTED ONE FALSE POSITIVE SARS RESULTS FOR ONE SYMPTOMATIC PATIENT. UNKNOWN IF THE RESULT WAS CONFIRMED. CONFIRMATION METHOD(S) NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21654465 ↗MalfunctionMar 13, 2025False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
FALSE POSITIVE RATE OF 36% WITH NO DETAILS REGARDING THE EXACT NUMBER OF INDIVIDUAL PATIENT EVENTS. NO CONFIRMATORY TESTING COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596853 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18104127 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103933 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103923 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840733 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840732 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840731 ↗MalfunctionDec 16, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING AN ABNORMALLY HIGH SARS POSITIVITY RATE ON ONE INSTRUMENT VS THE OTHER. CUSTOMER STATES THEY TESTED THE SAME PATIENT PLATE ON BOTH INSTRUMENTS. THE INSTRUMENT IN QUESTION PRODUCED 25 POSITIVE SARS RESULTS WHILE THE OTHER PRODUCED 2 POSITIVE SARS RESULTS. PATIENTS ARE BOTH SYMPTOMATIC AND ASYMPTOMATIC. IT IS NOTED THAT SERVICING OF THE INSTRUMENT WAS PERFORMED RECENTLY. REPORT 23 OF 23 DISCORDANT SAMPLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16000881 ↗MalfunctionSep 29, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING FALSE POSITIVE SARS RESULT. CUSTOMER STATES THE RESULT WAS CONFIRMED NEGATIVE BY ANOTHER MOLECULAR METHOD. IT IS NOTED THE CUSTOMER STARTED EXPERIENCING ISSUES AFTER A 3RD PARTY PIECE OF EQUIPMENT WAS REPLACED AND IS ALSO HAVING THE NEGATIVE QC CONTROL RESULT POSITIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15507083 ↗MalfunctionSep 29, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 4 FALSE POSITIVE SARS RESULTS. CUSTOMER STATES THE RESULTS WERE CONFIRMED NEGATIVE BY ANOTHER MOLECULAR METHOD. IT IS NOTED THE CUSTOMER STARTED EXPERIENCING ISSUES AFTER A 3RD PARTY PIECE OF EQUIPMENT WAS REPLACED AND IS ALSO HAVING THE NEGATIVE QC CONTROL RESULT POSITIVE. REPORT 3 OF 4
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15506722 ↗MalfunctionSep 29, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 4 FALSE POSITIVE SARS RESULTS. CUSTOMER STATES THE RESULTS WERE CONFIRMED NEGATIVE BY ANOTHER MOLECULAR METHOD. IT IS NOTED THE CUSTOMER STARTED EXPERIENCING ISSUES AFTER A 3RD PARTY PIECE OF EQUIPMENT WAS REPLACED AND IS ALSO HAVING THE NEGATIVE QC CONTROL RESULT POSITIVE. REPORT 2 OF 4.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15506656 ↗MalfunctionSep 29, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 4 FALSE POSITIVE SARS RESULTS. CUSTOMER STATES THE RESULTS WERE CONFIRMED NEGATIVE BY ANOTHER MOLECULAR METHOD. IT IS NOTED THE CUSTOMER STARTED EXPERIENCING ISSUES AFTER A 3RD PARTY PIECE OF EQUIPMENT WAS REPLACED AND IS ALSO HAVING THE NEGATIVE QC CONTROL RESULT POSITIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15506645 ↗MalfunctionSep 29, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 4 FALSE POSITIVE SARS RESULTS. CUSTOMER STATES THE RESULTS WERE CONFIRMED NEGATIVE BY ANOTHER MOLECULAR METHOD. IT IS NOTED THE CUSTOMER STARTED EXPERIENCING ISSUES AFTER A 3RD PARTY PIECE OF EQUIPMENT WAS REPLACED AND IS ALSO HAVING THE NEGATIVE QC CONTROL RESULT POSITIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15506543 ↗MalfunctionJul 27, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING A FALSE POSITIVE SARS RESULT WITH A SAMPLE SHOWING AN ATYPICAL CURVE. THE CUSTOMER COMMUNICATED THE RETEST OF THE SAMPLE IS NEGATIVE AND THE NEGATIVE RESULT WAS SUBMITTED. MOST PATIENTS WERE ASYMPTOMATIC AS THEY WERE TESTED FOR PRE-OP SCREENING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15117376 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page