Lyra Coronavirus (REF M120)Recalls & reported problems
Diagnostic Hybrids, Inc. · Lyra SARS-CoV-2 Assay · Class N
All models of Lyra (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Diagnostic Hybrids Lyra SARS-CoV-2 assay has 86 malfunction reports, with no deaths or injuries. The reports mostly describe false positive results, where the test indicated the virus in samples that were not positive, often without confirmatory testing. Some reports note an unapproved extraction system or transport media was in use. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 86
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 30014613339090), received between Apr 20, 2021 and Mar 20, 2025 — classified as 86 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (73 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event86
- False Positive Result73
- False Negative Result8
- Imprecise or Erratic Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions86
Recent adverse event reports
The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2025False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
REPORTED ONE FALSE POSITIVE SARS RESULTS FOR ONE SYMPTOMATIC PATIENT. UNKNOWN IF THE RESULT WAS CONFIRMED. CONFIRMATION METHOD(S) NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21654465 ↗MalfunctionMar 13, 2025False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
FALSE POSITIVE RATE OF 36% WITH NO DETAILS REGARDING THE EXACT NUMBER OF INDIVIDUAL PATIENT EVENTS. NO CONFIRMATORY TESTING COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21596853 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18104127 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103933 ↗MalfunctionNov 9, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SARS RESULTS REPORTING AS POSITIVE DUE TO HIGH BACKGROUND NOISE (POTENTIAL FALSE POSITIVE RESULTS). NO CONFIRMATION TESTING WAS PERFORMED. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103923 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840733 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840732 ↗MalfunctionSep 28, 2023False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING 3 SUSPECTED FALSE POSITIVE SARS RESULTS. NO CONFIRMATION TESTING WAS PERFORMED ON THE SAMPLES AND DURING INTERVIEW WITH THE CUSTOMER IT WAS FOUND THEY WERE USING AN UNAPPROVED EXTRACTION SYSTEM AS WELL AS AN OFF-LABEL TRANSPORT MEDIA. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17840731 ↗MalfunctionJul 20, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING SUSPECTED FALSE POSITIVE SARS RESULT WITH AN ATYPICAL CURVE. SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN OR IF ADDITIONAL METHODS WERE RUN. IT IS NOTED THE SAMPLE IS COLLECTED OFFSITE AND PLACED INTO VIRAL TRANSPORT MEDIA (VTM) SOLUTION PRIOR TO TESTING (OFF LABEL USAGE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15070732 ↗MalfunctionJul 11, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
CUSTOMER REPORTING FALSE POSITIVE SARS RESULT WITH A CT VALUE OF 28. CUSTOMER STATED THE SAMPLE WAS NEGATIVE BY ANOTHER LABORATORY TEST. SYMPTOMATIC STATUS OF PATIENT IS UNKNOWN. ATTEMPTS WERE MADE TO GATHER FURTHER INFORMATION FROM THE CUSTOMER WITHOUT SUCCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14978817 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887254 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887245 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887223 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887196 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887179 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887156 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887099 ↗MalfunctionJul 1, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
ONE FALSE POSITIVE SARS RESULT. THE CUSTOMER RULED OUT ENVIRONMENTAL CONTAMINATION SOURCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14887093 ↗MalfunctionJun 30, 2022False Positive Result; No Apparent Adverse Event; False Positive Result; No Apparent Adverse Event
THE CUSTOMER REPORTED A HIGH NUMBER OF FALSE POSITIVE RESULTS DUE TO SUSPECTED CONTAMINATION. THE EXACT NUMBER IS UNKNOWN AND MOST OF THE RESULTS WERE NOT LIKELY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880908 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880490 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880487 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880477 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880476 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880422 ↗MalfunctionJun 30, 2022False Negative Result; No Apparent Adverse Event; False Negative Result; No Apparent Adverse Event
CUSTOMER REPORTING 1 FALSE NEGATIVE SARS RESULT. CUSTOMER COMMUNICATED THE SAMPLE RESULTED POSITIVE BY 2 OTHER MOLECULAR METHODS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14880295 ↗
Common questions about the Lyra Coronavirus
Has the Lyra Coronavirus been recalled?
No FDA recall record in our data names this exact device identifier (DI 30014613339090), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Lyra Coronavirus?
The FDA's MAUDE database lists 86 adverse event reports for this device, received between Apr 20, 2021 and Mar 20, 2025. In those reports, the most common specific device problems are “False Positive Result” (73), “False Negative Result” (8) and “Imprecise or Erratic Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Lyra Coronavirus?
None of the reports are classified as deaths, 0 as injuries and 86 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Lyra Coronavirus, and is it still sold?
The FDA's device database lists Diagnostic Hybrids, Inc. as the company responsible for it (brand Lyra, model Coronavirus). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Reagents, 2019-novel coronavirus nucleic acid devices
- ID NOW™ COVID-19 190-000Alere Scarborough, Inc. · Reagents, 2019-Novel Coronavirus Nucleic Acid4 deaths1,932 reports
- Alinity m 09N78-095Abbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid3 recalls357 reports
- Lyra CoronavirusDiagnostic Hybrids, Inc. · Reagents, 2019-novel coronavirus nucleic acid403 reports
- Solana SARS-CoV-2 Assay CoronavirusDiagnostic Hybrids, Inc. · Reagents, 2019-novel coronavirus nucleic acid328 reports
- Alinity m 09N79-096Abbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid2 recalls179 reports
- Alinity m 09N78-03EAbbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid3 recalls0 reports
- BioFire® Respiratory Panel 2.1 EUA 423738Biofire Diagnostics, LLC · Reagents, 2019-Novel Coronavirus Nucleic Acid3 recalls0 reports
- Alinity m 09N79-05DAbbott Molecular Inc. · Reagents, 2019-novel coronavirus nucleic acid2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page