Skip to content

McKesson Cardiology HemoRecalls & reported problems across all models

Change Healthcare Canada Company · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one McKesson Cardiology Hemo model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
2
Injuries reported
0
Recalls
2

We found 2 models of McKesson Cardiology Hemo by Change Healthcare Canada Company in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between May 4, 2018 and Oct 31, 2018 — classified as 2 deaths.

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

McKesson Cardiology Hemo recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Dec 2, 2024Open, Classified

    Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).

    Z-0957-2025 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  2. Aug 21, 2024Open, Classified

    A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.

    Z-0128-2025 · Root cause: Use error · names 2 of these models · Recall details · FDA record ↗

Most reported McKesson Cardiology Hemo problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Sudden Cardiac Death2

McKesson Cardiology Hemo models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent McKesson Cardiology Hemo adverse event reports

The 2 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Change Healthcare Canada Company products

All Change Healthcare Canada Company devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page