McKesson Cardiology Hemo recallZ-0128-2025 · Change Healthcare Canada Company · initiated Aug 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Change Healthcare Canada Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
FDA root cause category: Use error
What the company is doing
Change Healthcare notified consignees on about 08/21/2024 via emailed letter. Consignees were instructed to pay careful attention to the hemoglobin unit of measurement displayed on the top of the "results" screen of the Hemo application, communicate the issue to all relevant team members to ensure vigilance in data entry, ensure that the hemoglobin values are entered in the correct unit of measurement (mmol/l or g/dl) as configured in the Management Console and displayed at the top of the results screen, contact Change Healthcare support to schedule the deployment of the software update that addresses this issue. Additionally, consignees were requested to complete and return the response form. With regard to potentially affected data, Change Healthcare recommended consignees verify f the UOM setting for hemoglobin is configured to what users would expect, if the UOM settings are different than expected, contact Change Healthcare support for assistance with identifying when the UOM setting was changed, and which procedures were affected, and examine affected procedures and implement necessary corrective actions to rectify the data and address patient care concerns.
Recall details
- Recall number
- Z-0128-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 21, 2024
- Posted by the FDA
- Oct 17, 2024
- Recalling firm
- Change Healthcare Canada Company
- Quantity
- 35 systems
- Product code
- DQK
- FDA submissions
- K131497
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.
Product description
McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
Code information (models, lots, serial numbers)
UDI Numbers: (01)80010939050010(10)130201, (01)80010939050027(10)140000. Versions Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- McKesson Cardiology Hemo 13.2.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable2 recalls0 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under K131497 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
- Change Healthcare Cardiology Hemo 14.1.1Change Healthcare Israel Ltd · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3Change Healthcare Canada Company · Computer, Diagnostic, Programmable2 recalls0 reports
- Change Healthcare Cardiology Hemo 14.1Change Healthcare Israel Ltd · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 recall0 reports
- Change Healthcare Cardiology Hemo 15.0Change Healthcare Canada Company · Computer, Diagnostic, Programmable1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95284).
- Z-0126-2025 · Open, ClassifiedChange Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page