Monarch Bronchscope 2.0/Monarch Platform (P2)Recalls & reported problems across all models

Auris Health, Inc. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one Monarch Bronchscope 2.0/Monarch Platform (P2) model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
109
Deaths reported
2
Injuries reported
92
Recalls
1

We found 2 models of Monarch Bronchscope 2.0/Monarch Platform (P2) by Auris Health, Inc. in the FDA's device registry (GUDID). Together they are named in 109 adverse event reports in the FDA's MAUDE database, received between Mar 8, 2021 and Jul 27, 2026 — classified as 2 deaths, 92 injuries, 15 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 2 reportsDec 2023: 5 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 4 reportsMay 2025: 2 reportsJun 2025: 5 reportsJul 2025: 0 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 4 reportsJan 2026: 2 reportsFeb 2026: 8 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Monarch Bronchscope 2.0/Monarch Platform (P2) recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jan 21, 2026Open, Classified

    Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

    Z-1383-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported Monarch Bronchscope 2.0/Monarch Platform (P2) problems

Device problems

  • Adverse Event Without Identified Device or Use Problem80
  • Appropriate Device Problem Term/Code Not Available9
  • Insufficient Device Problem Information6
  • Computer Software Problem4
  • Unintended Movement4
  • Application Program Problem3
  • Loose or Intermittent Connection2
  • Mechanical Problem2

Patient problems

  • Pneumothorax52
  • No Clinical Signs, Symptoms or Conditions42
  • Hemorrhage/Blood Loss/Bleeding9
  • Cardiac Arrest4
  • Decreased Respiratory Rate2
  • Dyspnea2
  • Hemoptysis2
  • Aortic dissection1

Monarch Bronchscope 2.0/Monarch Platform (P2) models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Monarch Bronchscope 2.0/Monarch Platform (P2) adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page