Monarch Bronchscope 2.0/Monarch Platform (P2) recallZ-1383-2026 · Auris Health, Inc · initiated Jan 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Auris Health, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
FDA root cause category: Software design
What the company is doing
On 1/21/2026, correction notices were emailed to customers informing them of the following: Should a MONARCH application restart occur during procedure setup, please clear the fault message and then perform a full system restart. A full system restart will prevent this issue from occurring as it will require the user to perform patient-side selection and alignment again. Once the software update to address the issue is available, the firm will contact customers to schedule installation. Complete and return the acknowledgement form via email to [email protected] This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potential affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective actions. If you have questions contact the firm at +1 (800) 434-0032 (toll-free within the United States), +1 (650) 264-6000 (Worldwide) or via email at [email protected] For Medical Information requests, visit firm's website at https://www.jnjmedtech.com/mir.
Recall details
- Recall number
- Z-1383-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 21, 2026
- Posted by the FDA
- Feb 17, 2026
- Recalling firm
- Auris Health, Inc
- Quantity
- 173
- Product code
- EOQ
- FDA submissions
- K193534, K211493, K243219
- Distribution
- Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Product description
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Monarch Platform, Bronchoscopy 4.1 MON-000008Auris Health, Inc. · Bronchoscope (Flexible Or Rigid)1 recall2 reports
- Monarch Bronchscope 2.0/Monarch Platform (P2) MON-000006Auris Health, Inc. · Bronchoscope (Flexible Or Rigid)1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K193534, K211493, K243219 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Monarch Bronchoscope MBR-000211-AAuris Health, Inc. · Bronchoscope (Flexible Or Rigid)12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.