MONARCHRecalls & reported problems across all models
Medos International Sàrl · 13 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 230
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls
- 0
We found 13 models of MONARCH by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 230 adverse event reports in the FDA's MAUDE database, received between Aug 15, 2017 and Sep 17, 2026 — classified as 8 injuries, 222 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (125 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported MONARCH problems
Device problems
- Calibration Problem125
- Mechanical Problem49
- Break22
- Material Twisted/Bent12
- Device-Device Incompatibility8
- Mechanical Jam7
- Contamination /Decontamination Problem4
- Corroded4
Patient problems
- No Clinical Signs, Symptoms or Conditions142
- No Consequences Or Impact To Patient45
- No Patient Involvement16
- No Known Impact Or Consequence To Patient14
- Not Applicable5
- Insufficient Information3
- No Code Available3
- Nonunion/Delayed-union Bone Fracture2
MONARCH models
13 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MONARCH 277040510WRENCH199 reports
- MONARCH 277030110ORTHOPEDIC MANUAL SURGICAL INSTRUMENT8 reports
- MONARCH 277020030DRIVER, PROSTHESIS6 reports
- MONARCH 277030545TEMPLATE3 reports
- MONARCH 277050110EXTRACTOR3 reports
- MONARCH 177492060ORTHOSIS, SPINAL PEDICLE FIXATION2 reports
- MONARCH 277030500WRENCH2 reports
- MONARCH 277040525DRIVER, PROSTHESIS2 reports
- MONARCH 177440040ORTHOSIS, SPINAL PEDICLE FIXATION1 report
- MONARCH 177440060ORTHOSIS, SPINAL PEDICLE FIXATION1 report
- MONARCH 177440062ORTHOSIS, SPINAL PEDICLE FIXATION1 report
- MONARCH 177470600ORTHOSIS, SPINAL PEDICLE FIXATION1 report
- MONARCH 277030105ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
Recent MONARCH adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Calibration Problem
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 50811925 ↗MalfunctionApr 29, 2026Calibration Problem
IT WAS REPORTED A DEVICE MALFUNCTION OCCURRED DUE TO FAILURE TO PASS THE TORQUE MEASUREMENT TEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25034705 ↗MalfunctionApr 24, 2026Calibration Problem
IT WAS REPORTED DR. STATED THAT THE TORQUES ARE APPARENTLY NOT EXERTING THE FORCE NECESSARY TO BLOCK THE LOCKING NUT ON THE BAR. WHEN THE BAR WAS TORQUED, IT CONTINUED TO MOVE ALONG THE CONSTRUCT, WHICH OBVIOUSLY SHOULD NOT HAPPEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24979864 ↗MalfunctionApr 22, 2026Calibration Problem
IT WAS REPORTED THE TORQUE WRENCH DID NOT EXERT THE FORCE NECESSARY TO LOCK THE LOCKING NUT ON THE BAR. WHEN THE BAR WAS TORQUED, IT CONTINUED TO MOVE ALONG THE CONSTRUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24955380 ↗MalfunctionApr 20, 2026Mechanical Jam
IT WAS REPORTED THAT HANDLE WAS NOT RATCHETING. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931946 ↗MalfunctionApr 20, 2026Mechanical Jam
IT WAS REPORTED THAT HANDLES WERE NOT RATCHETING. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931857 ↗MalfunctionApr 20, 2026Mechanical Jam
IT WAS REPORTED THAT HANDLES WERE NOT RATCHETING. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931851 ↗MalfunctionApr 20, 2026Mechanical Jam
IT WAS REPORTED THAT POLY SHAFT DRIVERS ARE STRIPPED. THE HANDLES ARE NOT RATCHING. THE TAPS ARE DULL AND THE BENDERS ARE DAMAGED, METAL BENT. ROD HOLDERS WORN. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24925468 ↗MalfunctionFeb 13, 2026Calibration Problem
IT WAS REPORTED THAT DURING ANNUAL TORQUE TESTING BY THE DISTRIBUTOR, SEVERAL DEVICES FAILED. THERE WERE NO RELATED PROCEDURES, AND NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335474 ↗MalfunctionFeb 13, 2026Calibration Problem
IT WAS REPORTED THAT DURING ANNUAL TORQUE TESTING BY THE DISTRIBUTOR, SEVERAL DEVICES FAILED. THERE WERE NO RELATED PROCEDURES, AND NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335463 ↗MalfunctionFeb 13, 2026Calibration Problem
IT WAS REPORTED THAT DURING ANNUAL TORQUE TESTING BY THE DISTRIBUTOR, SEVERAL DEVICES FAILED. THERE WERE NO RELATED PROCEDURES, AND NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24335452 ↗MalfunctionFeb 12, 2026Calibration Problem
IT WAS REPORTED THAT THE DEVICE FAILED TO PASS THE TORQUE MEASUREMENT TEST, WHICH LED TO A REQUEST FOR REPLACEMENT OF COMPONENTS IN THE KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24324753 ↗MalfunctionFeb 3, 2026Calibration Problem
ADDITIONAL INFORMATION RECEIVED: WHAT IS THE TIMING OF THE REPORTED EVENT? (PRE-OP, INTRA-OP, POST-OP): DURING INSPECTION AT LOCAL WAREHOUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24249256 ↗MalfunctionJan 13, 2026Calibration Problem
IT WAS REPORTED THAT THE TORQUE FAILED THE TORQUE TEST AT LOCAL WAREHOUSE DURING INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24057988 ↗MalfunctionDec 10, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THE TORQUE TEST FAILED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23767496 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page