Skip to content

NexdriveRecalls & reported problems across all models

Medtronic, Inc. · 4 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls
1

We found 4 models of Nexdrive by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Apr 28, 2017 and Jan 7, 2020 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

1 FDA recall record name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Nexdrive recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 9, 2019Terminated

    Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).

    Z-1995-2020 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported Nexdrive problems

Device problems

  • Insufficient Device Problem Information3
  • Device Operates Differently Than Expected2
  • Break1
  • Device Damaged Prior to Receipt by User1
  • Fitting Problem1

Patient problems

  • No Known Impact Or Consequence To Patient5

Nexdrive models

4 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Nexdrive adverse event reports

The 5 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic, Inc. products

All Medtronic, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page