OptifluxRecalls & reported problems across all models
Fresenius Medical Care Holdings, Inc. · 9 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Optiflux model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5,700
- Deaths reported
- 17
- Injuries reported
- 266
- Recalls
- 3
We found 9 models of Optiflux by Fresenius Medical Care Holdings, Inc. in the FDA's device registry (GUDID). Together they are named in 5,700 adverse event reports in the FDA's MAUDE database, received between Dec 30, 2015 and Sep 30, 2026 — classified as 17 deaths, 266 injuries, 5,417 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5,279 reports).
3 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Optiflux recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 29, 2025Open, Classified
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
Z-0041-2026 · Root cause: Device Design · names 4 of these models · Recall details · FDA record ↗
- Dec 5, 2024Open, Classified
Potential for internal blood leaks due to cracked polyurethane
Z-0801-2025 · Root cause: Process control · Recall details · FDA record ↗
- May 4, 2022Terminated
Dialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.
Z-1303-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Optiflux problems
Device problems
- Fluid/Blood Leak5,279
- Adverse Event Without Identified Device or Use Problem225
- Labelling, Instructions for Use or Training Problem134
- Obstruction of Flow34
- Leak/Splash15
- Device Operates Differently Than Expected14
- Air Leak7
- Crack7
Patient problems
- Hemorrhage/Blood Loss/Bleeding2,455
- No Clinical Signs, Symptoms or Conditions1,577
- No Known Impact Or Consequence To Patient1,434
- Blood Loss106
- Hypersensitivity/Allergic reaction98
- Itching Sensation88
- Reaction61
- Dyspnea40
Optiflux models
9 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Optiflux® 0500318EDialyzer, high permeability with or without sealed dialysate system1 recall6 deaths2,849 reports
- Optiflux® 0500316EDIALYZER, CAPILLARY, HOLLOW FIBER2 recalls7 deaths2,282 reports
- Optiflux® 0500320EDialyzer, high permeability with or without sealed dialysate system2 recalls242 reports
- Optiflux® 0500325EDialyzer, high permeability with or without sealed dialysate system1 recall1 death198 reports
- Optiflux® 0500318NDialyzer, high permeability with or without sealed dialysate system2 deaths55 reports
- Optiflux® 0500308EDialyzer, high permeability with or without sealed dialysate system28 reports
- Optiflux® 0500320NDialyzer, high permeability with or without sealed dialysate system24 reports
- Optiflux® 0500306EDialyzer, high permeability with or without sealed dialysate system13 reports
- Optiflux® 0500316NDialyzer, high permeability with or without sealed dialysate system1 death9 reports
Recent Optiflux adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50952208 ↗MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50952096 ↗MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50951907 ↗MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50951708 ↗MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50951569 ↗InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Hypersensitivity/Allergic reaction; Hypoxia; Respiratory Failure
MAUDE report 50951299 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
A nurse reported that during a hemodialysis (HD) treatment there was a dialyzer blood leak. In a follow up, the nurse said the leak occurred at the start of the hemodialysis (HD) treatment. The leak was seen as being external at Hansen hoses. Blood tests strips were not used.The leak was external. The B-Braun machine Braun bloodlines were being used.There was no damage noted on the dialyzer. There was patient blood loss of 250mL. There was no patient injury, adverse event or medical intervention reported. The patient went to emergency department for lab draws. The device is not available to be returned to the manufacturer for evaluation.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50936382 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50934857 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50933825 ↗MalfunctionSep 29, 2026Fluid/Blood Leak
A user facility¿s Clinic Manager (CM) reported that a visible internal dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (HD) treatment. The machine, a FME 2008T machine, alarmed appropriately. Blood leak test strips were used and tested positive for the presence of blood. A FME bloodline was in use. There was no obvious defect or damage seen on the dialyzer. The patient¿s estimated blood loss (EBL) was approximately 200mL. The patient was restarted on a new machine and treatment completed successfully with new supplies. There was no patient injury or medical intervention required due to the reported event. The complaint device is not available to be returned to the manufacturer for evaluation. Three photos were provided.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50927064 ↗MalfunctionSep 28, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50926292 ↗MalfunctionSep 28, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50923327 ↗MalfunctionSep 25, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50904141 ↗MalfunctionSep 25, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50899303 ↗MalfunctionSep 24, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50889998 ↗
Other Fresenius Medical Care Holdings, Inc. products
- Liberty5 models · 11,598 reports · 1 recall
- 2008T BlueStar Hemodialysis Machine4 models · 767 reports · 11 recalls
- CombiSet10 models · 819 reports · 1 recall
- stay.safe8 models · 187 reports · 6 recalls
- NaturaLyte2 models · 25 reports · 1 recall
- Bibag2 models · 0 reports · 1 recall
- Crit-Line2 models · 30 reports
- Hemoflow4 models · 29 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page