Skip to content

Optiflux recallZ-0801-2025 · Fresenius Medical Care Holdings, Inc. · initiated Dec 5, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for internal blood leaks due to cracked polyurethane

FDA root cause category: Process control

What the company is doing

Fresenius Medical Care notified consignees on about 12/05/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were instructed to stock immediately for either of the affected lot numbers, immediately discontinue use of Optiflux F160NRe Dialyzer(s) with lot codes, place any of the affected units in a secure, segregated area, arrange for return of affected product and complete and return the provided Reply Form. Additionally, consignees were instructed to forward the notification to any personnel within the organization or to customers where product was further distributed.

Recall details

Recall number
Z-0801-2025
Status
Open, Classified
Initiated by the company
Dec 5, 2024
Posted by the FDA
Jan 3, 2025
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
5,351 cases (64,212 dialyzers)
Product code
KDI
FDA submissions
K162488
Distribution
Nationwide Distribution.

Product description

Dialyzer Optiflux 160NRe

Code information (models, lots, serial numbers)

UDI NUMBERS: 24HU06011-00840861100149-270630, 24HU06010-00840861100149-270630. Expiration date: 06/30/2027.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page