PLEUR-EVACRecalls & reported problems across all models
Teleflex Incorporated · 15 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 146
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 1
We found 15 models of PLEUR-EVAC by Teleflex Incorporated in the FDA's device registry (GUDID). Together they are named in 146 adverse event reports in the FDA's MAUDE database, received between Feb 4, 2020 and Mar 25, 2026 — classified as 3 injuries, 143 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (73 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
PLEUR-EVAC recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 12, 2021Terminated
Teleflex is recalling this product due to a lack of assurance of sterility.
Z-1298-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported PLEUR-EVAC problems
Device problems
- Fluid/Blood Leak73
- Leak/Splash34
- Unsealed Device Packaging8
- Gas/Air Leak7
- Defective Component3
- Disconnection3
- Appropriate Device Problem Term/Code Not Available2
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions129
- Insufficient Information12
- Distress3
- No Consequences Or Impact To Patient3
- Pneumothorax2
- Pericardial Effusion1
- Pulmonary Emphysema1
PLEUR-EVAC models
15 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PLEUR-EVAC IPN028442BOTTLE, COLLECTION, VACUUM54 reports
- PLEUR-EVAC IPN028459CATHETER, IRRIGATION1 recall0 reports
- PLEUR-EVAC IPN028484BOTTLE, COLLECTION, VACUUM29 reports
- PLEUR-EVAC IPN028446BOTTLE, COLLECTION, VACUUM19 reports
- PLEUR-EVAC IPN028485APPARATUS, AUTOTRANSFUSION17 reports
- PLEUR-EVAC IPN028441APPARATUS, AUTOTRANSFUSION8 reports
- PLEUR-EVAC IPN025424BOTTLE, COLLECTION, VACUUM5 reports
- PLEUR-EVAC IPN025425APPARATUS, AUTOTRANSFUSION3 reports
- PLEUR EVAC IPN010586BOTTLE, COLLECTION, VACUUM3 reports
- PLEUR-EVAC IPN028454APPARATUS, AUTOTRANSFUSION2 reports
- PLEUR-EVAC IPN028472CATHETER, IRRIGATION2 reports
- PLEUR-EVAC IPN011527CATHETER, IRRIGATION1 report
- PLEUR-EVAC A-6002-08LFBOTTLE, COLLECTION, VACUUM1 report
- PLEUR EVAC IPN010597BOTTLE, COLLECTION, VACUUM1 report
- PLEUR-EVAC IPN028449BOTTLE, COLLECTION, VACUUM1 report
Recent PLEUR-EVAC adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryMar 25, 2026Gas/Air Leak
IT WAS REPORTED THAT "CHEST DRAIN, BLUE-COLOURED FLUID, WHICH TYPICALLY CONCERNS THE SUCTION CONTROL CHAMBER, DID NOT CHANGE COLOUR. BUBBLING OBSERVED IN THE DRAIN, EMPHYSEMA OBSERVED IN THE PATIENT. MEASURES TAKEN: DRAIN CHANGED + PATIENT TREATED IN INTENSIVE CARE."
Patient problems: Pulmonary Emphysema
MAUDE report 24683458 ↗MalfunctionSep 8, 2025Material Puncture/Hole; Gas/Air Leak
AIR LEAK THROUGH PLEUR-EVAC WHEN ON SUCTION DID NOT RESOLVE WITH REINFORCING DRESSING AROUND CHEST TUBE. CHANGING PLEUR-EVAC RESOLVED THE AIR LEAK, SUGGESTING THE CHAMBER OR TUBING MAY HAVE A HOLE. POSTOPERATIVE EFFUSION, WHICH WAS DRAINING WELL PRIOR TO MORNING OF [DATE REDACTED], STOPPED DRAINING WELL DURING THIS PERIOD OF TIME.
Patient problems: Pericardial Effusion
MAUDE report 22991971 ↗MalfunctionJun 14, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED " FAIL TO FUNCTION / DEFECTIVE, LEAKAT RED/BLUE JUNCTION. DURING USE." TO CONTINUE THERAPY, THE CANISTER WAS CHANGED. NO PATIENT HARM OR INJURY REPORTED. THE PATIENT CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19542888 ↗MalfunctionMar 20, 2024Break; Break
IT WAS REPORTED THAT, "WE WERE PLACING ONE OF THEM YESTERDAY AND THE KNOB BROKE OFF OF THE STYLET AFTER PLACEMENT, CAUSING THE USE OF HEMOSTATS TO REMOVE STYLET, WHICH CAUSED DISPLACEMENT." ADDITIONAL INFORMATION RECEIVED STATES THAT THE INCIDENT "LED TO BABY HAVING TO HAVE ANOTHER CHEST TUBE TO BE PLACED. BABY IS CRITICAL BUT STABLE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18941940 ↗MalfunctionSep 28, 2023Disconnection; Disconnection
IT WAS REPORTED THAT DURING USE ON PATIENT, "CONNECTIONS ARE NOT SECURE BETWEEN CHEST DRAIN CATHETER AND THE DRAINAGE TUBE. THE DRAINAGE TUBE WILL SPONTANEOUSLY DISCONNECT FROM THE CATHETER". AS A RESULT, THE "STAFF HAVE TAPED THE CONNECTIONS OR USED CABLE TIES". THE ISSUE OCCURRED WHILE THE PATIENT WAS LYING IN BED. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "STABLE AND DISCHARGED HOME".
Patient problems: Distress; No Clinical Signs, Symptoms or Conditions
MAUDE report 17835902 ↗MalfunctionSep 28, 2023Disconnection; Disconnection
IT WAS REPORTED THAT DURING USE ON PATIENT, "CONNECTIONS ARE NOT SECURE BETWEEN CHEST DRAIN CATHETER AND THE DRAINAGE TUBE. THE DRAINAGE TUBE WILL SPONTANEOUSLY DISCONNECT FROM THE CATHETER". AS A RESULT, THE "STAFF HAVE TAPED THE CONNECTIONS OR USED CABLE TIES". THE ISSUE OCCURRED WHILE THE PATIENT WAS LYING IN BED. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "STABLE AND DISCHARGED HOME".
Patient problems: Distress; No Clinical Signs, Symptoms or Conditions
MAUDE report 17835715 ↗MalfunctionSep 28, 2023Disconnection; Disconnection
IT WAS REPORTED THAT DURING USE ON PATIENT, "CONNECTIONS ARE NOT SECURE BETWEEN CHEST DRAIN CATHETER AND THE DRAINAGE TUBE. THE DRAINAGE TUBE WILL SPONTANEOUSLY DISCONNECT FROM THE CATHETER". AS A RESULT, THE "STAFF HAVE TAPED THE CONNECTIONS OR USED CABLE TIES". THE ISSUE OCCURRED WHILE THE PATIENT WAS LYING IN BED. NO PATIENT HARM OR INJURY HOWEVER THE CHEST DRAIN HAD TO BE PLACED ON SUCTION AS THE PATIENT'S PNEUMOTHORAX HAD INCREASED. THE PATIENT STATUS IS REPORTED AS "STABLE AND DISCHARGED HOME".
Patient problems: Distress; No Clinical Signs, Symptoms or Conditions; Pneumothorax
MAUDE report 17835662 ↗MalfunctionJul 28, 2023
PATIENT WAS PLACED ON A PLEUR-EVAC FOLLOWING OPEN HEART SURGERY. IN THE SURGICAL ICU, IT WAS NOTED THAT THE PLEUR-EVAC FAILED AS IT WOULD NOT KEEP SUCTION. THE PLEUR-EVAC WAS CHANGED, WHICH RESOLVED THE AIR LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17412846 ↗MalfunctionJun 6, 2023Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT "THE WATER SEAL CHAMBER BECOMES OVERFILLED WITH WATER AFTER THE PLEURA-VAC HAS BEEN SET UP. DURING SET UP IN THESE INSTANCES, THE WATER SEAL CHAMBER HAS BEEN SET TO THE RECOMMENDED 2 CM AND THE SUCTION CONTROL CHAMBER HAS BEEN SET TO THE RECOMMENDED 20 CM OF WATER. ABOUT AN HOUR AND A HALF AFTER SET UP, IT HAS BEEN NOTICED THAT THE WATER SEAL CHAMBER IS WAY OVER THE 2 CM MARK". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION, SPECIFICALLY REGARDING PATIENT DETAILS. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17070241 ↗MalfunctionJun 6, 2023Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT "THE WATER SEAL CHAMBER BECOMES OVERFILLED WITH WATER AFTER THE PLEURA-VAC HAS BEEN SET UP. DURING SET UP IN THESE INSTANCES, THE WATER SEAL CHAMBER HAS BEEN SET TO THE RECOMMENDED 2 CM AND THE SUCTION CONTROL CHAMBER HAS BEEN SET TO THE RECOMMENDED 20 CM. OF WATER. ABOUT AN HOUR AND A HALF AFTER SET UP, IT HAS BEEN NOTICED THAT THE WATER SEAL CHAMBER IS WAY OVER THE 2 CM MARK". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION, SPECIFICALLY REGARDING PATIENT DETAILS. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 17070169 ↗MalfunctionMay 18, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT WHO WAS ADMITTED WITH A SPONTANEOUS PNEUMOTHORAX, THE PLEUREVAC "WAS UNABLE TO MAINTAIN WATER LEVEL AT 0 DESPITE USING PRESSURE RELEASE AND/OR ADDING SALINE". THE PLEUREVAC WAS FIRST STARTED WITH THE SUCTION ON AND THEN WAS TRANSITIONED TO GRAVITY. THE REPORT STATES "THE WATER LEVEL WAS VERY LOW, BELOW -2 SO WE ADDED FLUID THEN THE PRESSURE SHOT WAY UP THE CHAMBER AND WE COULDN'T RELIEVE IT EVEN WITH PRESSURE VALVE RELEASED". AS A RESULT, THE PLEUREVAC WAS REMOVED AND A 2ND PLEUREVAC WAS USED. FURTHER INFORMATION STATES "THE MALFUNCTION OF THE UNIT DID NOT AFFECT THE PATIENT AND HE HAS SINCE BEEN TRANSFERRED TO ANOTHER SITE FOR SPECIALIST CARE. NOT RELATED TO CHEST DRAINAGE SYSTEM".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16954426 ↗MalfunctionApr 28, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT "THE CHEST TUBE CONTAINER WAS BUBBLING A LARGE AMOUNT WHEN APPLIED TO SUCTION EVEN WHEN THE TUBING TO THE PATIENT WAS CLAMPED SUGGESTING AIR ENTRY INTO THE CONTAINER. THERE WAS NO VISIBLE DAMAGE TO THE CONTAINER. THE CONTAINER WAS CHANGED FOR A NEW ONE AND THE BUBBLING SLOWED SIGNIFICANTLY. THE NEW CONTAINER ALSO DID NOT BUBBLE WHEN CLAMPED". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 16833529 ↗MalfunctionApr 24, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE "INCIDENT OCCURRED IN THE RESUSCITATION DEPARTMENT. THERE WAS A PLEURAL LIQUID LEAK AT THE LEVEL OF THE RED CONNECTOR. THERE WAS NO CONSEQUENCE FOR THE PATIENT". NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16799790 ↗MalfunctionApr 14, 2023Gas/Air Leak; Gas/Air Leak
REPORTED ISSUE: DEFECTIVE DRAINAGE SYSTEM WITH BUBBLING WHEN IT SHOULD NOT. THE AIR LEAK WAS FOUND AT THE RE/BLUE CONNECTION. THE DEVICE WAS CHANGED. NO CLINICAL CONSEQUENCES. ADDITIONAL INFORMATION RECEIVED ON 14 APRIL 2023 STATES THAT THE PATIENT'S CONDITION DID NOT WORSEN AFTER THE EVENT, THERE WAS NO INJURY TO THE PATIENT, AND THAT THERE WAS NO MEDICAL INTERVENTION. THE LEAK OCCURRED WHILE TRANSPORTING THE PATIENT BACK FROM SURGERY AND IT WAS AN AIR LEAK THAT OCCURRED ON THE RED/BLUE CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16743849 ↗MalfunctionMar 27, 2023Leak/Splash; Leak/Splash
REPORTED ISSUE: LEAK WOULD RESOLVE WHEN USING PRESSURE TO HOLD THE RED CLAMP AND TUBE TOGETHER TIGHTLY. WATERPROOF TAPE ATTEMPTED AND DID NOT HELP. CHANGED DRAINAGE SYSTEMS WHICH FIXED THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16624137 ↗
Other Teleflex Incorporated products
- ARROW1,494 models · 9,043 reports · 62 recalls
- RUSCH220 models · 436 reports · 130 recalls
- WECK97 models · 3,790 reports · 1 recall
- HUDSON RCI152 models · 1,461 reports · 7 recalls
- LANGSTON4 models · 15 reports · 5 recalls
- LMA AIRWAY MANAGEMENT47 models · 139 reports
- UROLIFT13 models · 137 reports
- GIBECK7 models · 32 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page