PLEUR-EVAC IPN028446Recalls & reported problems
Teleflex Incorporated · PLEUR-EVAC ADULT-PEDIATRIC WET LATEX FREE · Class 2
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 34026704631279), received between Mar 29, 2021 and Jun 6, 2023 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Leak/Splash11
- Fluid/Blood Leak6
- Inability to Auto-Fill2
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- Insufficient Information3
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2023Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT "THE WATER SEAL CHAMBER BECOMES OVERFILLED WITH WATER AFTER THE PLEURA-VAC HAS BEEN SET UP. DURING SET UP IN THESE INSTANCES, THE WATER SEAL CHAMBER HAS BEEN SET TO THE RECOMMENDED 2 CM AND THE SUCTION CONTROL CHAMBER HAS BEEN SET TO THE RECOMMENDED 20 CM OF WATER. ABOUT AN HOUR AND A HALF AFTER SET UP, IT HAS BEEN NOTICED THAT THE WATER SEAL CHAMBER IS WAY OVER THE 2 CM MARK". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION, SPECIFICALLY REGARDING PATIENT DETAILS. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17070241 ↗MalfunctionJun 6, 2023Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED THAT "THE WATER SEAL CHAMBER BECOMES OVERFILLED WITH WATER AFTER THE PLEURA-VAC HAS BEEN SET UP. DURING SET UP IN THESE INSTANCES, THE WATER SEAL CHAMBER HAS BEEN SET TO THE RECOMMENDED 2 CM AND THE SUCTION CONTROL CHAMBER HAS BEEN SET TO THE RECOMMENDED 20 CM. OF WATER. ABOUT AN HOUR AND A HALF AFTER SET UP, IT HAS BEEN NOTICED THAT THE WATER SEAL CHAMBER IS WAY OVER THE 2 CM MARK". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION, SPECIFICALLY REGARDING PATIENT DETAILS. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 17070169 ↗MalfunctionApr 28, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT "THE CHEST TUBE CONTAINER WAS BUBBLING A LARGE AMOUNT WHEN APPLIED TO SUCTION EVEN WHEN THE TUBING TO THE PATIENT WAS CLAMPED SUGGESTING AIR ENTRY INTO THE CONTAINER. THERE WAS NO VISIBLE DAMAGE TO THE CONTAINER. THE CONTAINER WAS CHANGED FOR A NEW ONE AND THE BUBBLING SLOWED SIGNIFICANTLY. THE NEW CONTAINER ALSO DID NOT BUBBLE WHEN CLAMPED". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: Insufficient Information
MAUDE report 16833529 ↗MalfunctionNov 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LEAKAGE WAS FOUND BETWEEN THE RED AND BLUE FAST CLICK CONNECTOR WHILE USING IN PATIENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15759092 ↗MalfunctionNov 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LEAKAGE WAS FOUND BETWEEN THE RED AND BLUE FAST CLICK CONNECTORS WHILE USING IN THE PATIENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15759044 ↗MalfunctionNov 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LEAKAGE WAS FOUND BETWEEN THE RED AND BLUE FAST CLICK CONNECTOR WHILE USING IN PATIENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15759012 ↗MalfunctionOct 26, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AIR LEAKS ON CONNECTION TUBING. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15677012 ↗MalfunctionOct 26, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT AIR LEAKS ON CONNECTION TUBING. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15676985 ↗MalfunctionSep 23, 2022Leak/Splash; Leak/Splash
THE CUSTOMER MENTIONED THAT THE CLINICAL STAFF HEARD A HISSING NOISE (POTENTIAL AIR LEAK) AT THE SITE OF THE QUICK DISCONNECT. ADDITIONAL INFORMATION INDICATES THAT THE NOISE WAS COMING FROM THE RED/BLUE CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15469631 ↗MalfunctionAug 12, 2022Leak/Splash; Leak/Splash
AIR LEAK AT THE QUICK DISCONNECT FEATURE ON THE PATIENT TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15218681 ↗MalfunctionAug 1, 2022Leak/Splash; Leak/Splash
WE HAVE HAD 3 CASES IN THE LAST WEEK IN WHICH WE HAVE IDENTIFIED A DEFECT IN THE PLEUR-EVAC, NO LOT NUMBERS OR DEFECTIVE PRODUCT A VAILABLE. THE IS A DEFECT AT THE JUNCTION OF THE RUBBER TUBING TO THE CORRUGATED TUBING (RED TO BLUE CONNECTION). WE HAVE HAD TO CHANGE OUT THESE 3 VACS FOR THESE PATIENTS WITHIN 24 HOURS POST OP. WE HAVE ALSO COMPLETED FDA FORM 3500.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15138469 ↗MalfunctionAug 1, 2022Leak/Splash; Leak/Splash
WE HAVE HAD 3 CASES IN THE LAST WEEK IN WHICH WE HAVE IDENTIFIED A DEFECT IN THE PLEUR-EVAC, NO LOT NUMBERS OR DEFECTIVE PRODUCT A VAILABLE. THE IS A DEFECT AT THE JUNCTION OF THE RUBBER TUBING TO THE CORRUGATED TUBING (RED TO BLUE CONNECTION). WE HAVE HAD TO CHANGE OUT THESE 3 VACS FOR THESE PATIENTS WITHIN 24 HOURS POST OP. WE HAVE ALSO COMPLETED FDA FORM 3500.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15138460 ↗MalfunctionAug 1, 2022Leak/Splash; Leak/Splash
WE HAVE HAD 3 CASES IN THE LAST WEEK IN WHICH WE HAVE IDENTIFIED A DEFECT IN THE PLEUR-EVAC, NO LOT NUMBERS OR DEFECTIVE PRODUCT AV AILABLE. THE IS A DEFECT AT THE JUNCTION OF THE RUBBER TUBING TO THE CORRUGATED TUBING (RED TO BLUE CONNECTION). WE HAVE HAD TO CHANGE OUT THESE 3 VACS FOR THESE PATIENTS WITHIN 24 HOURS POST OP. WE HAVE ALSO COMPLETED FDA FORM 3500.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15138407 ↗MalfunctionJul 20, 2022Leak/Splash; Leak/Splash
AN AIR LEAK AT THE QUICK DISCONNECT FEATURE ON THE PATIENT TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15069836 ↗MalfunctionApr 14, 2022Leak/Splash; Leak/Splash
AIR HEARD HISSING FROM PLEUREVAC. AIR LEAK NOTED. ENTIRE PLEUREVAC CHANGED AND THEN NO ISSUE. NO VISIBLE CRACK BUT AIR LEAK FROM RED/BLUE CONNECTOR. THE SAME ISSUE HAS HAPPENED WITH ALL OF THESE CHEST DRAINS. 2 OF THE 6 WERE USED IN A CLINICAL SETTING. THE REMAINING DRAINS WERE TESTED OUTSIDE OF USING ON A PATIENT AND THE CUSTOMER FOUND SIMILAR ISSUES ON THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14116024 ↗MalfunctionApr 14, 2022Leak/Splash; Leak/Splash
AIR HEARD HISSING FROM PLEUREVAC. AIR LEAK NOTED. ENTIRE PLEUREVAC CHANGED AND THEN NO ISSUE. NO VISIBLE CRACK BUT AIR LEAK FROM RED/BLUE CONNECTOR. THE SAME ISSUE HAS HAPPENED WITH ALL OF THESE CHEST DRAINS. 2 OF THE 6 WERE USED IN A CLINICAL SETTING. THE REMAINING DRAINS WERE TESTED OUTSIDE OF USING ON A PATIENT AND THE CUSTOMER FOUND SIMILAR ISSUES ON THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14116007 ↗MalfunctionApr 19, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE RED CONNECTION HAD A LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11687832 ↗MalfunctionApr 19, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE RED CONNECTION HAD A LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11687350 ↗MalfunctionMar 29, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE RED CONNECTION HAD A LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11581074 ↗
Other BOTTLE, COLLECTION, VACUUM devices
- OASIS 3600-100Atrium Medical Corporation · Bottle, Collection, Vacuum3 recalls1 death175 reports
- OCEAN 2002-000Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM3 recalls44 reports
- EXPRESS 16400Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall110 reports
- EXPRESS 4000-100NAtrium Medical Corporation · Bottle, Collection, Vacuum2 recalls30 reports
- OCEAN 2002-400Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM2 recalls18 reports
- OASIS 3612-100Atrium Medical Corporation · Bottle, Collection, Vacuum1 recall18 reports
- Medline DYNDSCL1500Medline Industries, Inc. · BOTTLE, COLLECTION, VACUUM1 recall13 reports
- PNEUMOSTAT 16100Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall3 deaths12 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.