Power-LOADRecalls & reported problems across all models
Stryker Corporation · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 196
- Deaths reported
- 3
- Injuries reported
- 22
- Recalls
- 0
We found 3 models of Power-LOAD by Stryker Corporation in the FDA's device registry (GUDID). Together they are named in 196 adverse event reports in the FDA's MAUDE database, received between Feb 16, 2018 and Apr 23, 2026 — classified as 3 deaths, 22 injuries, 171 malfunctions.
The most frequently reported specific device problem is “Device Fell” (77 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Power-LOAD problems
Device problems
- Device Fell77
- Sharp Edges54
- Device Dislodged or Dislocated27
- Break10
- Patient Device Interaction Problem6
- Difficult or Delayed Positioning4
- Difficult to Remove4
- Fire3
Patient problems
- No Clinical Signs, Symptoms or Conditions117
- Pain23
- No Patient Involvement22
- Insufficient Information21
- No Consequences Or Impact To Patient10
- Muscle/Tendon Damage9
- No Known Impact Or Consequence To Patient9
- Injury5
Power-LOAD models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Power-LOAD 6390STRETCHER, WHEELED3 deaths124 reports
- Power-LOAD 6390STRETCHER, WHEELED48 reports
- Power-LOAD 639005550003STRETCHER, WHEELED24 reports
Recent Power-LOAD adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24967393 ↗MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24967218 ↗MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24967216 ↗MalfunctionJan 26, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24177663 ↗MalfunctionJan 26, 2026Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24177458 ↗MalfunctionJan 26, 2026Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24177231 ↗MalfunctionOct 27, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT THAT EXPERIENCED PAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 23393833 ↗MalfunctionOct 27, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; IT WAS NOT KNOWN WHAT THE PATIENT EXPERIENCED.
Patient problems: Insufficient Information
MAUDE report 23393831 ↗MalfunctionOct 27, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1- SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393823 ↗MalfunctionOct 27, 2025Device Fell; Device Fell
NO NEW INFORMATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions; Pain; Unspecified Musculoskeletal problem
MAUDE report 23393421 ↗InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE RAISED THE COT AND THEN DROPPED IT. AS A RESULT, THE USER INJURED A BICEP AND HAD TO HAVE SURGERY.
Patient problems: Muscle/Tendon Damage
MAUDE report 23373011 ↗MalfunctionOct 10, 2025Fire; Smoking; Fire; Smoking
IT WAS REPORTED THAT THE DEVICE'S BATTERY COMPARTMENT MAY HAVE CAUGHT ON FIRE. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23267440 ↗MalfunctionOct 10, 2025Fire; Fire
IT WAS REPORTED THAT THERE WAS A FIRE. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23267418 ↗MalfunctionJul 24, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 13 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592715 ↗MalfunctionJul 24, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592713 ↗
Other Stryker Corporation products
- RemB7 models · 3,073 reports
- Prime Series21 models · 1,548 reports · 2 recalls
- TARGET114 models · 1,077 reports
- ProCuity17 models · 368 reports · 13 recalls
- Spetzler-Malis20 models · 7 reports · 19 recalls
- SYNCHRO 28 models · 313 reports · 9 recalls
- Neptune E-SEP19 models · 412 reports · 7 recalls
- StrykeFlow4 models · 552 reports · 4 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.