Power-LOAD 6390 (REF 6390-000-000)Recalls & reported problems
Stryker Corporation · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Power-LOAD wheeled stretcher is used to load and unload patients. Many of the 125 reports describe the stretcher falling, dislodging, or having sharp exposed edges, and some describe crew members hurt during loading. The reports include 3 deaths and 19 injuries. Reports fell to 2 in the last 12 months from 10 in the previous 12. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 124
- Deaths reported
- 3
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 124 adverse event reports that list this exact device (DI 07613327261523), received between Feb 16, 2018 and Apr 23, 2026 — classified as 3 deaths, 18 injuries, 103 malfunctions.
The most frequently reported specific device problem is “Device Fell” (47 reports).
3 other recalls cover devices cleared under the same FDA submission (K113598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Fell47
- Sharp Edges27
- Device Dislodged or Dislocated17
- Break10
- Patient Device Interaction Problem5
- Difficult to Remove4
- Installation-Related Problem3
- Mechanical Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Patient Involvement20
- Insufficient Information16
- Pain16
- No Consequences Or Impact To Patient9
- No Known Impact Or Consequence To Patient9
- Injury5
- Muscle/Tendon Damage5
Recent adverse event reports
The 25 most recent of 124 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24967216 ↗MalfunctionOct 27, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1- SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393823 ↗MalfunctionOct 27, 2025Device Fell; Device Fell
NO NEW INFORMATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions; Pain; Unspecified Musculoskeletal problem
MAUDE report 23393421 ↗InjuryOct 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE RAISED THE COT AND THEN DROPPED IT. AS A RESULT, THE USER INJURED A BICEP AND HAD TO HAVE SURGERY.
Patient problems: Muscle/Tendon Damage
MAUDE report 23373011 ↗MalfunctionOct 10, 2025Fire; Fire
IT WAS REPORTED THAT THERE WAS A FIRE. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23267418 ↗MalfunctionApr 28, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS, WHERE IT WAS REPORTED, THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922045 ↗MalfunctionApr 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21921719 ↗MalfunctionApr 28, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT WITH A FALL; NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 21919821 ↗MalfunctionJan 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A PINCH/CRUSH POINT. THIS ALLEGEDLY LED TO A MINOR INJURY TO A STAFF MEMBER IN THE FORM OF A LACERATION.
Patient problems: Laceration(s)
MAUDE report 21276326 ↗MalfunctionJan 30, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A FAILURE OF THE FASTENER TO FACILITATE PROPER COT OPERATIONS DURING LOADING OR UNLOADING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275772 ↗MalfunctionJan 28, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21257414 ↗MalfunctionOct 29, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560518 ↗MalfunctionOct 29, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES/ FALSE ENGAGEMENT INTO FASTENER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556627 ↗MalfunctionOct 29, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556597 ↗MalfunctionJul 29, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19850338 ↗MalfunctionJul 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER UNEXPECTEDLY DISENGAGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19843397 ↗MalfunctionJul 26, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19843302 ↗MalfunctionJun 18, 2024Device Fell; Device Fell
IT WAS REPORTED THAT THE DEVICE USER EXPERIENCED A COMPLETE POWER LOAD HYDRAULIC FAILURE WHILE LOADING A PATIENT INTO THE BACK OF AN AMBULANCE. THIS FAILURE CAUSED THE POWER LOAD AND STRETCHER (XT) TO GIVE WAY. FORTUNATELY, THE CREW MANAGED TO SUPPORT THE STRETCHER, PREVENTING THE PATIENT FROM FALLING TO THE FLOOR. ALTHOUGH THE PATIENT WAS UNINJURED, THE CREW EXPERIENCED SOME STRAIN ON THEIR BACKS.
Patient problems: Muscle/Tendon Damage
MAUDE report 19560619 ↗MalfunctionJun 3, 2024Device Fell; Device Fell
IT WAS REPORTED THAT THE DEVICE DROPPED THE COT DURING LOADING, WHICH RESULTED IN THE CAREGIVER HURTING THEIR HAND. THE CUSTOMER REPORTED THAT THERE WAS NOT A SIGNIFICANT INJURY FROM THIS EVENT.
Patient problems: Insufficient Information; Pain
MAUDE report 19454679 ↗MalfunctionApr 29, 2024Device Fell; Device Fell
IT WAS REPORTED THAT THE SCREWS ON THE DEVICE'S WHEELS SNAPPED AND CAUSED THE COT TO FALL DOWN TO THE GROUND. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT FELL OFF OF THE COT AND WAS INJURED.
Patient problems: Abrasion; Hematoma; Insufficient Information
MAUDE report 19209313 ↗MalfunctionApr 23, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19162975 ↗MalfunctionApr 23, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THE PATIENT EXPERIENCED ANXIETY AND PAIN.
Patient problems: Anxiety; No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 19162969 ↗InjuryApr 12, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE USER WILL HAVE TO HAVE A PARTIAL AMPUTATION OF A FINGER. ATTEMPTS ARE BEING MADE TO GATHER ADDITIONAL DETAILS FROM THE USER FACILITY.
Patient problems: Deformity/ Disfigurement
MAUDE report 19101275 ↗MalfunctionJan 24, 2024Patient Device Interaction Problem; Device Fell; Patient Device Interaction Problem; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FASTENER NO LONGER FACILITATES PROPER COT OPERATIONS DURING LOADING/UNLOADING AND PINCH/CRUSH POINT. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18574533 ↗MalfunctionJan 24, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18573187 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113598. One submission often covers many models, so check the product and lot details in each record.
- Z-1350-2019 · Nov 20, 2018 · TerminatedPower-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000
- Z-1389-2018 · Mar 12, 2018 · TerminatedPower-LOAD Cot Fastening System, Model #6390, Catalog #6390-000-000, UDI: 07613327261523 Power-LOAD is intended to assist with loading and unloading of a compatible wheeled stretc
- Z-0566-2013 · Oct 11, 2012 · TerminatedThe Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of the ambulance.
Other STRETCHER, WHEELED devices
- Power-PRO 6506Stryker Corporation · STRETCHER, WHEELED5 deaths472 reports
- Procedural stretcher P8000HHill-Rom de México, S. de R.L. de C.V. · STRETCHER, WHEELED361 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 death251 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 recall199 reports
- Power-PRO 2, High Configuration 650705550001Stryker Corporation · STRETCHER, WHEELED2 recalls2 deaths99 reports
- NA 1089Stryker Corporation · STRETCHER, WHEELED175 reports
- NA 5050Stryker Corporation · STRETCHER, WHEELED1 recall99 reports
- Power-PRO 2 650700000000Stryker Corporation · STRETCHER, WHEELED2 recalls1 death42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.