Skip to content

Powerheart G5 Semi Automatic AEDRecalls & reported problems across all models

Zoll Medical Corporation · 5 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name at least one Powerheart G5 Semi Automatic AED model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
31
Deaths reported
1
Injuries reported
0
Recalls
3

We found 5 models of Powerheart G5 Semi Automatic AED by Zoll Medical Corporation in the FDA's device registry (GUDID). Together they are named in 31 adverse event reports in the FDA's MAUDE database, received between Dec 17, 2018 and Aug 12, 2026 — classified as 1 death, 30 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (15 reports).

3 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Powerheart G5 Semi Automatic AED recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 13, 2025Class IIOpen, Classified

    The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

    Z-1322-2025 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

  2. Feb 13, 2025Class IIOpen, Classified

    The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

    Z-1320-2025 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

  3. May 31, 2024Class IIOpen, Classified

    G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

    Z-2324-2024 · Root cause: Use error · names 2 of these models · Recall details · FDA record ↗

Most reported Powerheart G5 Semi Automatic AED problems

Device problems

  • Unable to Obtain Readings15
  • Failure of Device to Self-Test6
  • Failure to Deliver Shock/Stimulation2
  • Failure to Power Up2
  • Output Problem2
  • Use of Device Problem2
  • Charging Problem1
  • Communication or Transmission Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Cardiac Arrest1
  • Insufficient Information1
  • Necrosis1

Powerheart G5 Semi Automatic AED models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Powerheart G5 Semi Automatic AED adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Zoll Medical Corporation products

All Zoll Medical Corporation devices →

Sources and updates