Powerheart® G5 Semi Automatic AED G5S-01P1Recalls & reported problems

Zoll Medical Corporation · Powerheart® G5 Semi-Automatic AED no battery or pads · Class 3

All models of Powerheart G5 Semi Automatic AED (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00812394021260), received between Dec 17, 2018 and Apr 10, 2024 — classified as 1 death, 17 malfunctions.

The most frequently reported specific device problem is “Unable to Obtain Readings” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unable to Obtain Readings9
  • Failure to Deliver Shock/Stimulation2
  • Output Problem2
  • Use of Device Problem2
  • Charging Problem1
  • Failure of Device to Self-Test1
  • High impedance1
  • Incorrect Interpretation of Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Cardiac Arrest1
  • Necrosis1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A RED "X" INDICATOR. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19081460 ↗
  • MalfunctionApr 10, 2024Use of Device Problem; Use of Device Problem

    COMPLAINANT ALLEGED THAT DURING A FUNCTIONAL TESTING, THE DEVICE DISPLAYED A RED "X" INDICATOR. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19077934 ↗
  • MalfunctionJul 7, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17282703 ↗
  • MalfunctionJul 7, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17281985 ↗
  • MalfunctionJul 7, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17281877 ↗
  • MalfunctionJul 7, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17281659 ↗
  • MalfunctionJul 7, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED THE WATCHDOG TIMER SELFTEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17281462 ↗
  • MalfunctionJul 6, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT - 501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17270803 ↗
  • MalfunctionJun 26, 2023Charging Problem; Charging Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INTERNALLY DUMPED THE ENERGY AFTER CHARGING UP TO DEFIBRILLATE THE PATIENT. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17198293 ↗
  • MalfunctionMar 30, 2023Output Problem; Output Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16645200 ↗
  • MalfunctionMar 30, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16645047 ↗
  • MalfunctionMar 30, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16645008 ↗
  • MalfunctionMar 29, 2023Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16636558 ↗
  • MalfunctionMar 29, 2023Output Problem; Output Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DEFIB OUTPUT WAS OUT OF SPECIFICATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16636256 ↗
  • MalfunctionMar 2, 2023High impedance; High impedance

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED FOR HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16470373 ↗
  • MalfunctionOct 19, 2022Unable to Obtain Readings; Unable to Obtain Readings

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "ECG FAULT -501" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15632083 ↗
  • DeathMay 7, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE FAILED TO DISCHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED, HOWEVER IT WAS NOT A RESULT OF THE REPORTED MALFUNCTION.

    Patient problems: Necrosis

    MAUDE report 14319772 ↗
  • MalfunctionDec 17, 2018Failure to Deliver Shock/Stimulation; Incorrect Interpretation of Signal; Failure to Deliver Shock/Stimulation; Incorrect Interpretation of Signal

    THE FIRE FIGHTER CREW WAS THE FIRST ON THE SCENE OF A CARDIAC ARREST AND ATTACHED THEIR CARDIAC SCIENCE AED TO THE PATIENT. THE AED ANALYZED THE PATIENT'S RHYTHM AND DETERMINED NO SHOCK WAS REQUIRED. THE PARAMEDICS ARRIVED ON THE SCENE AT APPROXIMATELY THE SAME TIME AS WHEN CPR WAS STARTED. THE PARAMEDICS ATTACHED THEIR PADS TO THE PATIENT AND USING THEIR EQUIPMENT IMMEDIATELY DETERMINED THE PATIENT WAS IN VF AND SHOCKED THE PATIENT. THE CUSTOMER IS QUESTIONING WHY THE CARDIAC SCIENCE AED DID NOT ADVISE A SHOCK. PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Cardiac Arrest

    MAUDE report 8172039 ↗