PowerPICC Sherlock 3CGRecalls & reported problems across all models
Bard Access Systems, Inc. · 7 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one PowerPICC Sherlock 3CG model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 2
We found 7 models of PowerPICC Sherlock 3CG by Bard Access Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 6 adverse event reports in the FDA's MAUDE database, received between Apr 22, 2025 and May 11, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (3 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
PowerPICC Sherlock 3CG recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 21, 2026Open, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · names 3 of these models · Recall details · FDA record ↗
Most reported PowerPICC Sherlock 3CG problems
Device problems
- Difficult or Delayed Separation3
- Fluid/Blood Leak1
- Material Deformation1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
PowerPICC Sherlock 3CG models
7 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PowerPICC Sherlock 3CG CK000668BCatheter, intravascular, therapeutic, long-term greater than 30 days1 recall3 reports
- PowerPICC Sherlock 3CG CK000983Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Sherlock 3CG CK000746ACatheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Sherlock 3CG CK001067Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Sherlock 3CG CK000887ACatheter, intravascular, therapeutic, long-term greater than 30 days1 report
- PowerPICC Sherlock 3CG CK001062Catheter, intravascular, therapeutic, long-term greater than 30 days1 report
- PowerPICC Sherlock 3CG CK000631CCatheter, intravascular, therapeutic, long-term greater than 30 days1 report
Recent PowerPICC Sherlock 3CG adverse event reports
The 6 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT DURING THE INITIAL ATTEMPT TO DILATE THE VEIN USING THE INTRODUCER, THE DILATOR FAILED TO FUNCTION AS EXPECTED. THE PATIENT DID NOT HAVE ANY ADVERSE EFFECTS AND DID NOT HAVE SKIN THAT WOULD LEAD ME TO THINK THE DILATOR WOULD FAIL. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25137723 ↗MalfunctionApr 23, 2026Difficult or Delayed Separation
IT WAS REPORTED THE SHEATH THAT BREAK AWAY DID NOT BREAK APART PROPERLY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966339 ↗MalfunctionApr 23, 2026Difficult or Delayed Separation
IT WAS REPORTED THE SHEATH THAT BREAK AWAY DID NOT BREAK APART PROPERLY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966309 ↗MalfunctionFeb 6, 2026Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT THE CATHETER THAT WAS IMPLANTED IN (B)(6). RECENTLY, THE PATIENT NOTICED A PROBLEM, AND UPON EXAMINATION, THE DOCTOR DISCOVERED A LEAK IN THE PICC LINE. THE PICC WAS SUBSEQUENTLY REPLACED. NO FURTHER DETAILS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24276980 ↗MalfunctionFeb 3, 2026Difficult or Delayed Separation
IT WAS REPORTED ONE OF OUR PICC RNS ATTEMPTED TO PLACE A PICC LINE ON (B)(6) 2026. THEY ADVISED THAT THE SHEATH WAS NOT SCORED AND COULD NOT BE BROKEN, THUS MALFUNCTIONING LEADING TO A SECOND PROCEDURE FOR AN ALREADY ANXIOUS PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24251881 ↗MalfunctionApr 22, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED BY CUSTOMER THAT INSIDE OF KIT SMELLS MOLDY- HORRIBLE SMELL COMING FROM INSIDE KIT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21882912 ↗
Other Bard Access Systems, Inc. products
- Port Access Needle122 models · 3,744 reports · 3 recalls
- Groshong NXT ClearVue Catheter8 models · 3,266 reports
- PowerPort ClearVUE27 models · 1,993 reports · 15 recalls
- Site~Rite 8 Ultrasound System11 models · 1,965 reports
- PowerPort M.R.I., ChronoFlex22 models · 1,642 reports · 1 recall
- PowerPICC Provena116 models · 1,337 reports · 2 recalls
- PowerPort M.R.I., Groshong10 models · 1,285 reports
- Navarre15 models · 1,228 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page