BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155… recallZ-1507-2025 · Bard Access Systems, Inc. · initiated Mar 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Access Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

FDA root cause category: Process control

What the company is doing

On 3/11/2025, recall notices were distributed to customers and distributors who were asked to do the following: Actions if catheter damage is suspected: 1. Stop any infusion if catheter damage is suspected. 2. Follow your institution's guidelines for catheters with suspected damage. 3. If catheter fracture is confirmed, catheter should be removed and an alternative route for access should be obtained. 4. Report any catheter replacements due to leakage to firm's Complaint Center. Please Take the Following Actions: 1. Destroy all unused product following your institution's process for destruction. 2. If affected device was previously used on a patient without incident, no further action is necessary. 3. Recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 4. Complete and return Customer Response Form via email to Email: [email protected] Distributors asked to identify all customers within distribution network that purchased any affected product. Provide a copy of the attached customer letter to all customers to advise them of this recall Actions Taken by Firm: 2. Firm is notifying users that adhesive-backed securement systems are recommended. 3. Firm is notifying users that if they do use compression-style securement systems, they must be sized appropriately to accommodate the taper region. 4. Firm has been updating the product IFU with cautions related to the use of compression-style securement systems around the taper region of the PICC. 5. Firm is notifying customers to insert the catheter as close as possible to the zero mark per the IFU. For further assistance, contact: North American Regional Complaint Center: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT or [email protected]. On 5/15/2025 additional recall notices were distributed to customers informing them of additional affected devices.

Recall details

Recall number
Z-1507-2025
Status
Open, Classified
Initiated by the company
Mar 11, 2025
Posted by the FDA
Apr 16, 2025
Recalling firm
Bard Access Systems, Inc.
Quantity
718,461
Product code
LJS
FDA submissions
K210264
Distribution
Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Product description

BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784

Code information (models, lots, serial numbers)

Show the full code information (12k characters)

REF/UDI-DI/Lot(Expiration): 3174155/00801741027611/REGQ0844(6/30/2025), REGZ1439(11/30/2025), REHP0860(11/30/2025), REHQ3097(6/30/2026), REHR0022(6/30/2026), REHT1151(7/31/2026), REHU1525(11/30/2026), REHU2455(6/30/2026), REHU2614(11/30/2026), REHV2765 (7/31/2026), REJN3472(2/28/2027), REJP4216(5/31/2027), REJQ2272(7/31/2027), REJR1411(8/31/2027), REJR2565(4/30/2027), REJR2614(8/31/2027), REJT0755(9/30/2027); CK000540/00801741105142/REGZ1425(10/31/2025), REHP0808(11/30/2025), REHU1513(1/31/2026), REHV2221(1/31/2026), REHY2842(10/31/2026), REJN4145 (12/31/2026), REJR1322(3/31/2027), REJS2110(4/30/2027), REJS3773(2/28/2027); 3174118/00801741027598/REGZ2512(11/30/2025), REHP1760(1/31/2026), REHP2603(1/31/2026), REJN1059(12/31/2026); CK000483/00801741099397/REHN3187(10/31/2025), REHQ1804(6/30/2026), REJN1065(5/31/2027), REJP2018(6/30/2027), REJR1321(8/31/2027); 3174335/00801741027628/REHN3908(10/31/2025), REHP0800(10/31/2025), REHQ3070(11/30/2025), REHT1129(5/31/2026), REHU1519(6/30/2026), REHV1391(7/31/2026), REHV3050(7/31/2026), REHZ1011(11/30/2026), REJN4109(12/31/2026), REJQ0405(2/28/2027), REJR2090(3/31/2027), REJT0833(5/31/2027); 3174355/00801741027635/REHQ1781(11/30/2025), REHS1925(4/30/2026), REHT1120(5/31/2026), REHU0975(6/30/2026), REHU2488(6/30/2026), REHV1400(7/31/2026), REHV3020(7/31/2026), REHW1329(8/31/2026), REHY1684(10/31/2026), REHZ1903(11/30/2026), REJN3007(12/31/2026), REJP0855(1/31/2027), REJP2021(1/31/2027), REJQ2284(2/28/2027), REJR2157(3/31/2027), REJS2171(4/30/2027), REJS4490(4/30/2027); CK000196/00801741028847/REHQ3064(6/30/2026), REHT2954(1/31/2026), REHX2926(12/31/2026), REHZ1873(3/31/2027), REJQ2279(8/31/2027); 9174118/00801741035562/REHQ3091(6/30/2026), REHR0017(6/30/2026), REHT1132(6/30/2026), REHU2496(6/30/2026), REHY2811(2/28/2026), REHZ1882(3/31/2027); CK000955/00801741200922/REHQ4029(11/30/2025), REHV2232(2/28/2026), REHX2916(12/31/2025), REHY3954(7/31/2026), REJN2995(9/30/2025); 3174135/00801741027604/REHR0003(11/30/2025), REHS4184(7/31/2026), REHU2356(11/30/2026), REHU2612(6/30/2026), REJS2089(9/30/2027); CK000617/00801741122286/REHS1001(9/30/2025), REJQ2280(2/28/2027); CK000839/00801741154393/REHS4241(4/30/2026), REHV3026(7/31/2026), REHX2908(9/30/2026), REHY2826(10/31/2026), REJN3000(12/31/2026), REJS3479(4/30/2027); 3194355/00801741027789/REHT2157(5/31/2025), REHT2967(5/31/2025), REHU3317(6/30/2025), REHV3031(7/31/2025), REHY3957(10/31/2025), REHZ1878(11/30/2025), REJN3008(12/31/2025), REJR1276(12/31/2025), REJR2062(3/31/2026), REJS2065(12/31/2025); 6174118/00801741139017/REHT2984(1/31/2026); 3194135/00801741027758/REHU1029(5/31/2025), REHY1636(10/31/2025), REHY2820(10/31/2025), REJN2980(12/31/2025); 3194335/00801741027772/REHU1430(6/30/2025), REHU2339(6/30/2025), REHW1354(8/31/2025), REHY2700(10/31/2025), REHZ1026(10/31/2025), REJN1463(12/31/2025), REJN3016(12/31/2025), REJR2013(3/31/2026), REJS3853(12/31/2025); 2174108/00801741138966/REHU3327(9/30/2025); 9194118/00801741028618/REHV0263(6/30/2025), REJQ0411(1/31/2026); 3194118/00801741027741REHY1654(10/31/2025), REJP0843(5/31/2025), REJS0206(3/31/2026); 1174118/00801741034466/REHY2815(5/31/2025), REJN1443(8/31/2025), REJQ0415(5/31/2025), REJQ2243(8/31/2025), REJS4488(10/31/2025); CK000285C/00801741161117/REHZ1477(5/31/2025), REJN2365(5/31/2025), REJQ2812(8/31/2025), REJR2208(9/30/2025), REJS1287(9/30/2025), REJS4740(10/31/2025); CK000335A/00801741101281/REHZ1620(5/31/2025), REJN2425(6/30/2025), REJQ0819(8/31/2025), REJQ1491(8/31/2025); CK000796/00801741144783/REHZ1625(5/31/2025), REJQ1521(8/31/2025), REJR1458(9/30/2025), REJS0258(9/30/2025), REJS3583(10/31/2025), REJT0877(11/30/2025), REJX0310(2/28/2026); 1194108D/00801741034510/REHZ1633(5/31/2025), REJN3882(5/31/2025), REJN3972(6/30/2025), REJP4057(7/31/2025), REJP4060(7/31/2025), REJP4066(7/31/2025), REJP4068(7/31/2025), REJR2209(9/30/2025), REJR2227(9/30/2025), REJS1269(9/30/2025), REJS2211(9/30/2025), REJS3526(10/31/2025), REJU1142(10/31/2025), REJV3948(1/31/2026); CK000757/00801741141133/REHZ1652(5/31/2025), REJQ0793(6/30/2025), REJR1443(8/31/2025), REJS1253(7/31/2025), REJS3543(8/31/2025), REJV0288(10/31/2025); CK001042/00801741224324/REHZ1657(5/31/2025), REJN2389(6/30/2025), REJR1448(9/30/2025), REJT0890(11/30/2025); 9174108/00801741035548/REJN0035(6/30/2025), REJQ0431(8/31/2025), REJR0582(9/30/2025); 1174108D1/00801741154737/REJN0722(5/31/2025), REJP1731(7/31/2025), REJR0469(8/31/2025), REJR2556(9/30/2025), REJS2083(10/31/2025); CK000631B/00801741143748/REJN0748(5/31/2025), REJP2723(7/31/2025), REJP2747(7/31/2025), REJS1310(9/30/2025), REJS4709(10/31/2025); 3194155/00801741027765/REJN0964(6/30/2025), REJN1080(6/30/2025), REJN4126(6/30/2025), REJQ0418(8/31/2025), REJS3757(10/31/2025), REJT1225(11/30/2025); 3174108/00801741027574/REJN1098(6/30/2025), REJR2206(9/30/2025), REJS2212(10/31/2025); CK000567/00801741108167/REJN1450(5/31/2025), REJQ0400(8/31/2025); 1194108D4/00801741155093/REJN2245(5/31/2025), REJP1828(7/31/2025), REJQ1321(6/30/2025), REJR0440(6/30/2025), REJR2590(7/31/2025), REJS2165(7/31/2025), REJS4634(7/31/2025); 1174108D2/00801741154836/REJN2302(5/31/2025), REJP1701(5/31/2025), REJQ1440(8/31/2025), REJR1289(9/30/2025), REJR2589(9/30/2025), REJS2453(10/31/2025), REJS4511(9/30/2025), REJU1046(11/30/2025); CK000677/00801741131790/REJN2329(5/31/2025), REJR1433(9/30/2025), REJS3919(10/31/2025); CK000864/00801741163302/REJN2340(5/31/2025), REJR1423(7/31/2025), REJR2685(7/31/2025), REJS2226(6/30/2025); CK000937A/00801741224003/REJN2347(5/31/2025), REJQ0824(6/30/2025), REJR2631(7/31/2025), REJS3571(8/31/2025); CK000930/00801741188091/REJN2351(5/31/2025), REJR2649(9/30/2025); CK001027/00801741223921/REJN2359(5/31/2025), REJR2690(9/30/2025), REJT1434(9/30/2025); 1174108D/00801741034459/REJN2362(5/31/2025), REJN2363(5/31/2025), REJP4065(7/31/2025), REJQ0816(8/31/2025), REJR0595(9/30/2025), REJR0600(9/30/2025), REJR2677(9/30/2025), REJS1284(9/30/2025), REJS2210(10/31/2025), REJS3915(10/31/2025), REJS4729(10/31/2025); CK000867/00801741163272/REJN2377(5/31/2025); CK000900/00801741184956/REJN2381(5/31/2025), REJR0589(6/30/2025), REJR1461(6/30/2025), REJS1259(6/30/2025), REJS2603(6/30/2025); CK000315A/00801741182723/REJN2391(6/30/2025), REJN3978(5/31/2025), REJQ2835(8/31/2025), REJR2697(9/30/2025), REJT1407(11/30/2025); CK000589A/00801741137624/REJN2395(6/30/2025), REJR2198(7/31/2025); CK000854B/00801741225802/REJN2400(6/30/2025), REJN3856(6/30/2025), REJQ2816(8/31/2025), REJS1317(9/30/2025), REJT0876(11/30/2025); CK000897/00801741184925/REJN2404(6/30/2025), REJN3860(6/30/2025), REJQ2807(6/30/2025), REJR2686(7/31/2025), REJS0277(7/31/2025), REJS1295(6/30/2025), REJT0874(7/31/2025); CK000953A/00801741223303/REJN2409(6/30/2025), REJS2228(8/31/2025), REJV1726(1/31/2026); CK000586B/00801741212789/REJN2420(6/30/2025), REJR1455(6/30/2025), REJR2211(7/31/2025); CK000549/00801741107313/REJN2427(6/30/2025), REJU2672(12/31/2025); CK000844/00801741154690/REJN3737(5/31/2025), REJS0060(9/30/2025); 1194108D1/00801741154782/REJN3778(5/31/2025), REJQ1238(8/31/2025), REJR2004(8/31/2025), REJS1173(7/31/2025); 1174108D4/00801741155048/REJN3820(5/31/2025), REJP1880(7/31/2025), REJR0536(6/30/2025), REJR1407(8/31/2025), REJR2563(6/30/2025), REJS2541(8/31/2025), REJS3729(7/31/2025); 1194108D3/00801741154997/REJN3821(5/31/2025), REJP3949(5/31/2025), REJR0537(9/30/2025), REJR2557(7/31/2025), REJS2084(10/31/2025), REJS4512(10/31/2025), REJU0239(11/30/2025); CK000934/00801741188381/REJN3864(6/30/2025), REJQ1483(8/31/2025), REJS4722(10/31/2025); CK000986/00801741216008/REJN3867(6/30/2025), REJN3923(5/31/2025), REJQ0825(7/31/2025), REJQ1499(7/31/2025), REJR2670(9/30/2025), REJS1324(10/31/2025), REJS4724(10/31/2025); CK000411/00801741096303/REJN3872(5/31/2025); CK001018/00801741222993/REJN3880(6/30/2025), REJQ1526(8/31/2025), REJT1397(9/30/2025); 1174108/00801741034442/REJN3881(5/31/2025), REJQ0432(8/31/2025), REJR2638(9/30/2025); CK000060/00801741028731/REJN3891(6/30/2025), REJQ0799(8/31/2025), REJQ1517(8/31/2025), REJR1432(9/30/2025), REJS1329(10/31/2025), REJS3916(10/31/2025); CK000513A/00801741110030/REJN3898(5/31/2025), REJS3539(8/31/2025); CK001021/00801741222863/REJN3927(6/30/2025) CK001030/00801741223976/REJN3970(6/30/2025), REJS0262(8/31/2025); CK000501A/00801741140815/REJN3980(6/30/2025), REJR2220(9/30/2025), REJT1408(11/30/2025), REJU2651(12/31/2025); CK000619A/00801741132001/REJN3985(6/30/2025); CK000903/00801741183089/REJN3999(6/30/2025), REJS0283(9/30/2025), REJT0909(11/30/2025); 3194108/00801741027710/REJN4166(6/30/2025), REJS3535(10/31/2025); 1194108D2/00801741154881/REJP0630(7/31/2025), REJR0470(8/31/2025), REJR0502(9/30/2025), REJS2556(10/31/2025); 9194108D/00801741035609/REJP1767(7/31/2025), REJR2695(9/30/2025); CK000959/00801741200960/REJP1942(7/31/2025), REJQ1524(8/31/2025), REJS0284(9/30/2025); CK000613/00801741121395/REJP1943(7/31/2025), REJQ1493(8/31/2025), REJR1670(9/30/2025), REJS0293(9/30/2025), REJS2216(10/31/2025), REJS4779(10/31/2025), REJV1721(1/31/2026); CK000991/00801741213298/REJP1950(5/31/2025), REJQ1512(8/31/2025), REJT1415(9/30/2025); CK000983/00801741225734/REJP2027(7/31/2025), REJP4183(7/31/2025), REJS1105(10/31/2025), REJS3409(10/31/2025); CK000881/00801741188787/REJP2717(7/31/2025), REJR1424(8/31/2025); CK000721/00801741137655/REJP2725(7/31/2025), REJR2213(9/30/2025), REJS1291(9/30/2025), REJT1365(11/30/2025); CK000642/00801741127908/REJP2727(7/31/2025), REJT0895(10/31/2025); CK000289C/00801741161148/REJP2728(7/31/2025), REJQ0813(8/31/2025), REJR2219(9/30/2025), REJS2203(9/30/2025), REJT1406(11/30/2025), REJV0222(9/30/2025), REJV1719(9/30/2025), REJX3527(1/31/2026); CK000894A/00801741188732/REJP2737(5/31/2025), REJS2218(10/31/2025); CK000914/00801741202599/REJP2741(7/31/2025), REJR1425(8/31/2025), REJS1319(9/30/2025), REJT0899(11/30/2025); CK000445/00801741101557/REJP2744(7/31/2025), REJR2662(9/30/2025); CK000687/00801741136535/REJP2746(6/30/2025), REJV3817(9/30/2025); CK000524A/00801741124068/REJP2748(6/30/2025), REJR1418(7/31/2025); CK000819/00801741152566/REJP2750(6/30/2025), REJS0294(6/30/2025); CK000266A/00801741185014/REJP2755(7/31/2025); 1174108D3/00801741154935/REJP4030(7/31/2025), REJR1272(9/30/2025), REJS1196(10/31/2025), REJS4633(7/31/2025); 3194108D/00801741027727/REJP4058(7/31/2025), REJS1304(5/31/2025); CK000545/00801741107399/REJP4222(1/31/2027), REJQ2267(2/28/2027); 1174108D5/00801741155147/REJQ0781(5/31/2025), REJQ1423(6/30/2025), REJS2164(6/30/2025), REJT0678(7/31/2025); 3174108D/00801741027581/REJQ0800(8/31/2025), REJQ1341(8/31/2025), REJR2656(9/30/2025); CK000775A/00801741188060/REJQ0807(8/31/2025), REJR1435(9/30/2025), REJS4754(10/31/2025), REJT0896(9/30/2025); 1194108D5/00801741155192/REJQ1441(6/30/2025), REJR1268(8/31/2025), REJS2058(7/31/2025), REJS3889(7/31/2025), REJT0862(7/31/2025); CK000736/00801741139499/REJQ1480(6/30/2025); CK000739/00801741139659/REJQ1601(6/30/2025), REJR2683(6/30/2025), REJS3894(8/31/2025), REJU1681(8/31/2025), REJV0873(8/31/2025), REJV3827(1/31/2026); 9194108/00801741028595/REJR0580(9/30/2025); 1194108/)00801741034503/REJR0596(9/30/2025), REJR2204(9/30/2025), REJS3533(10/31/2025); CK000660/00801741129650/REJR1419(9/30/2025), REJS1312(9/30/2025); CK000609A/00801741222504/REJR1654(6/30/2025); 9174108D/00801741035555/REJR1668(9/30/2025), REJR2230(9/30/2025); CK001067/00801741228711/REJR2202(8/31/2025), REJS3899(10/31/2025); CK000077D/00801741139598/REJR2231(9/30/2025); CK000882/00801741188794/REJS0259(8/31/2025), REJT0908(11/30/2025); CK000874/00801741178696/REJS0282(7/31/2025), REJS2188(6/30/2025), REJT1411(11/30/2025); CK000994/00801741215698/REJS0285(7/31/2025), REJS3572(8/31/2025); CK000870A/00801741184246/REJS3569(10/31/2025); CK000443C/00801741226557/REJT0880(10/31/2025); CK000668B/00801741228674/REJT1820(10/31/2025); CK000784/00801741189012/REJU0379(9/30/2025)