Skip to content

PreciceRecalls & reported problems across all models

Nuvasive Specialized Orthopedics, Inc. · 38 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one Precice model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
4
Recalls
1

We found 38 models of Precice by Nuvasive Specialized Orthopedics, Inc. in the FDA's device registry (GUDID). Together they are named in 9 adverse event reports in the FDA's MAUDE database, received between Feb 22, 2021 and Mar 11, 2026 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Precice recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · names 30 of these models · Recall details · FDA record ↗

Most reported Precice problems

Device problems

  • Break4
  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Corroded1
  • Defective Device1
  • Fracture1

Patient problems

  • Failure of Implant3
  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Deformity/ Disfigurement1
  • Insufficient Information1

Precice models

38 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Precice adverse event reports

The 9 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Nuvasive Specialized Orthopedics, Inc. products

All Nuvasive Specialized Orthopedics, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page