Precice CTL4.5-040-SRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · 4.5mm x 40mm Screw, Locking, PL · Class 2 · Implant
All models of Precice (38)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887517029874), received between Apr 22, 2021 and Oct 13, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 13, 2021Break; Break
INFORMATION WAS RECEIVED THAT A BROKEN SCREW WAS EXTRACTED. NO PATIENT ADVERSE EVENT WAS REPORTED. REPORT 2 OF 2.
Patient problems: Failure of Implant
MAUDE report 12628480 ↗InjuryApr 22, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED IN (B)(6) 2021 TO ADDRESS LOOSE SCREWS. AS PER THE REPORTER, APPROXIMATELY TWO AND A HALF WEEKS POST-IMPLANTATION, IT WAS DISCOVERED THAT THE SCREWS IN THE PROXIMAL BONE SEGMENT HAD LOOSENED AND DISENGAGED THE PLATE. NO PATIENT ADVERSE EVENT WAS REPORTED. DURING THE REVISION, THE SCREWS WERE RE-TIGHTENED AND THE SCREW HOLES WERE BACK-FILLED WITH CEMENT TO PREVENT FUTURE SCREW BACK-OUT. POST REVISION, THE PATIENT HAS SUCCESSFULLY RESUMED LENGTHENING. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11712092 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page