Protecta XT CRT-DRecalls & reported problems across all models
Medtronic, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 286
- Deaths reported
- 7
- Injuries reported
- 233
- Recalls
- 0
We found 3 models of Protecta XT CRT-D by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 286 adverse event reports in the FDA's MAUDE database, received between Feb 25, 2013 and Feb 7, 2024 — classified as 7 deaths, 233 injuries, 46 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (128 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Protecta XT CRT-D problems
Device problems
- Premature Discharge of Battery128
- Adverse Event Without Identified Device or Use Problem37
- Battery Problem16
- Failure to Charge15
- Inappropriate/Inadequate Shock/Stimulation15
- Delayed Charge Time13
- Invalid Sensing11
- Appropriate Device Problem Term/Code Not Available10
Patient problems
- No Known Impact Or Consequence To Patient193
- Unspecified Infection28
- No Clinical Signs, Symptoms or Conditions18
- Shock from Patient Lead(s)14
- Death7
- No Consequences Or Impact To Patient4
- Bradycardia3
- Dyspnea3
Protecta XT CRT-D models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Protecta™ XT CRT-D D314TRGDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)6 deaths185 reports
- Protecta™ XT CRT-D D314TRMDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 death99 reports
- Protecta™ XT CRT-D D314TRGDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 reports
Recent Protecta XT CRT-D adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2024Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) COULD NOT BE INTERROGATED. A DONUT MAGNET WAS PLACED OVER THE DEVICE AND THERE WERE NO TONES. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660728 ↗MalfunctionJan 11, 2024Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS UNABLE TO BE INTERROGATED AND IT WAS SUSPECTED THAT THE DEVICE WAS AT END OF SERVICE (EOS). THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18494267 ↗InjuryAug 18, 2023Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED BY THE PATIENT THAT THEIR PREVIOUS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) ONLY LASTED A FEW YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17573918 ↗InjuryJun 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO AN INFECTION AND WAS REPLACED WITH A LEADLESS PACEMAKER. IT WAS NOTED THAT THE PATIENT HAD LOW BLOOD PRESSURE, AND A BLOOD TRANSFUSION WAS REQUIRED DUE TO UNKNOWN COMPLICATIONS. IT WAS ADDITIONALLY REPORTED THAT DURING THE IMPLANT OF THE LEADLESS IMPLANTABLE PULSE GENERATOR (IPG) WHEN THE PHYSICIAN ATTEMPTED TO CROSS THE TRICUSPID VALVE, THE PATIENT WENT INTO VENTRICULAR FIBRILLATION (VF) AND REQUIRED EXTERNAL DEFIBRILLATION TO BREAK THE ARRHYTHMIA. FOLLOWING THE EXTERNAL DEFIBRILLATION WHEN THE LEADLESS IPG WAS POSITIONED IN THE PATIENT, THE MOBILE PROGRAMMER LOST TELEMETRY WITH THE DEVICE AND 'FROZE' DURING IMPEDANCE AND R-WAVE TESTING. THE LEADLESS IPG WAS SUCCESSFULLY CUT FROM THE TETHER AND REMAINS IMPLANTED, AND THE MOBILE PROGRAMMER REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Low Blood Pressure/ Hypotension; Unspecified Infection
MAUDE report 17142397 ↗MalfunctionMay 11, 2023Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE CLINICAL WAS TRYING TO DO A MANUAL TRANSMISSION, HOWEVER, THE REMOTE MONITOR HAD NO TELEMETRY WITH THE DEVICE. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE REMOTE MONITOR AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16911233 ↗InjuryMar 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT'S RELATIVE THAT THE PATIENT WENT TO THE HOSPITAL DUE TO A BRADYCARDIA EPISODE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 16506112 ↗InjuryAug 15, 2022Battery Problem; Insufficient Device Problem Information; Battery Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT STATED THAT THEIR PREVIOUS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE WAS EXPLANTED BECAUSE IT HAD A "PROBLEM." THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15227698 ↗InjuryJul 7, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO PREMATURE BATTERY DEPLETION. THE CRT-D WAS REPLACED WITH A COMPETITOR DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14944896 ↗MalfunctionFeb 2, 2022Failure to Charge; Failure to Interrogate; Failure to Charge; Failure to Interrogate
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS AT END OF SERVICE (EOS) AND EXPERIENCED A CHARGE CIRCUIT TIMEOUT ALERT. THE CRT-D REMAINS IN THE PATIENT FOLLOWING A HEART TRANSPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13425396 ↗MalfunctionOct 6, 2021Pacing Problem; Communication or Transmission Problem; Pacing Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THERE WAS A PROBLEM ESTABLISHING TELEMETRY WITH THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). WHEN TELEMETRY WAS FINALLY ABLE TO BE ESTABLISHED, IT SHOWED NO SIGNS OF PACING. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12584174 ↗InjuryAug 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED DUE TO CELLULITIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cellulitis
MAUDE report 12374433 ↗InjuryAug 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN OPEN AND INFECTED DEVICE IMPLANT SITE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection; Wound Dehiscence
MAUDE report 12371034 ↗MalfunctionAug 12, 2021Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time; Failure to Charge; Electrical /Electronic Property Problem; Delayed Charge Time
IT WAS FURTHER REPORTED THAT THE DECISION WAS MADE TO NOT REPLACE THE DEVICE AS THE PATIENT WENT TO HOSPICE CARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12313712 ↗InjuryJun 2, 2021Failure to Charge; Electrical /Electronic Property Problem; Battery Problem; Failure to Charge; Electrical /Electronic Property Problem; Battery Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAD AN ALERT FOR A CHARGE CIRCUIT TIMEOUT. IT WAS NOTED THAT THE DEVICE TRIGGERED RECOMMENDED REPLACEMENT TIME (RRT) BACK IN 2018. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921606 ↗InjuryJun 2, 2021Failure to Charge; Failure to Charge
IT WAS REPORTED THAT AN ALERT OCCURRED DUE TO A CHARGE CIRCUIT TIMEOUT ON A DEVICE THAT WAS PREVIOUSLY PROGRAMMED OFF PER PATIENT'S REQUEST. THE DEVICE WAS REPROGRAMMED AND THE CHARGE CIRCUIT WAS INACTIVATED. THE DEVICE REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921009 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page